BIORAPTOR 72201542Recalls & reported problems
Smith & Nephew, Inc. · BIORAPTOR 2.3 PK SUTURE ANCHOR WITH ONE No.2 PRELOADED ULTRABRAID SUTURE, COBRAID BLACK/WHITE · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 03596010595089), received between Feb 20, 2023 and Sep 16, 2024 — classified as 1 injury, 2 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K071586, K152566, K251627); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Oct 29, 2024Open, Classified
Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
Z-0581-2025 · Root cause: Package design/selection · FDA record ↗
Most reported problems
Device problems
- Break2
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2024Packaging Problem; Packaging Problem
IT WAS REPORTED THAT DURING A SHOULDER SCOPE, THE ANCHOR TIP OF THE BIORAPTOR SUTURE ANCHOR WAS PENETRATING THROUGH THE PACKAGING, IT WAS NO LONGER STERILE. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20225387 ↗MalfunctionApr 17, 2024Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE BIORAPTOR SUTURE PK ANCHOR BROKE WHEN INSERTING INTO THE PREPARED CHANNEL. THE COMPONENTS WERE REMOVED FROM THE JOINT. THE ORIGINAL BONE HOLE WAS USED. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19126453 ↗InjuryFeb 20, 2023Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE BIORAPTOR ANCHOR GOT BROKEN WHILE INSERTING INSIDE THE BONE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY USING A BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16405008 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071586, K152566, K251627. One submission often covers many models, so check the product and lot details in each record.
- Z-1273-2020 · Dec 30, 2019 · TerminatedTwinFix Ti 2.8 mm HS Suture Anchor - Product Usage: A fixation device intended to provide secure attachment of soft tissue to bone.
Other FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE devices
- TRUESPAN ORTHOCORD 228151Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,686 reports
- TRUESPAN ORTHOCORD 228152Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE1,398 reports
- OMNISPAN ORTHOCORD 228142Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE932 reports
- FOOTPRINT 72202901Smith & Nephew, Inc. · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE721 reports
- RIGIDLOOP 232447Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE633 reports
- Arthrex® AR-3636Arthrex, Inc. · Fastener, fixation, nondegradable, soft tissue624 reports
- OMNISPAN ORTHOCORD 228141Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE609 reports
- TRUESPAN ORTHOCORD 228162Medos International Sàrl · FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE460 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page