Cartiva Synthetic Cartilage Implant (SCI) Catalog:… recallZ-0598-2025 · Cartiva, Inc · initiated Oct 31, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cartiva, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Patients implanted with synthetic cartilage implant, may experience a higher-than expected occurrence rate of the following hazards: revision, removal, implant subsidence, displacement, pain, nerve damage or fragmentation.
FDA root cause category: Device Design
What the company is doing
On 10/31/24 recall notices were mailed or emailed to customers asking them to do the following: 1) Continue to follow patients treated with an impacted product for new or worsening symptoms of pain, difficulty walking, skin reactions, stiffness, swelling, or weakness of the big toe joint, consistent with your follow up protocols. Per Instructions for Use: the long-term effects of cartilage replacement are not known; and the clinical and medical status of each patient should be considered when treating patients. 2) To help minimize complications, reference the information in the Instructions for Use and the information included in this notification. Per standard practice, continue to discuss all potential risks identified and discuss the benefits and risks of all relevant treatment options for first metatarsophalangeal joint osteoarthritis with your patients. 3) Check your internal inventory to locate the products listed on the attached business reply form, remove them from their point of use, isolate/quarantine the unit(s), and return. 4) Share and maintain awareness of this communication in your practice with until all required actions have been completed within your facility. 5) If you have further distributed the affected product, please notify the applicable parties about this notice. You may copy and distribute this notification letter. 6) If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details so that we can inform the recipients appropriately. 7) If you are a distributor, note that you are responsible for notifying your affected customers. 8) Complete and return the response form via email to [email protected] For questions or concerns, contact [email protected].
Recall details
- Recall number
- Z-0598-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 31, 2024
- Posted by the FDA
- Dec 4, 2024
- Recalling firm
- Cartiva, Inc
- Product code
- PNW
- FDA submissions
- P150017
- Distribution
- Worldwide - US Nationwide distribution in the states of CO, CT, FL, IL, MI, MS, NY, OH, RI, UT, VA, AL, AR, AZ, CA, IA, ID, IN, KS, KY, MA, MN, MO, MT, NC, NE, NM, NV, OK, OR, PA, TN, TX, WA, WI and the countries of Australia, Austria, Brazil, Canada, Chile, Cyprus, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Kuwait, Latvia, Lebanon, Netherlands, New Zealand, Panama, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Product description
Cartiva Synthetic Cartilage Implant (SCI) Catalog: CAR-06-US (6mm), CAR-08-US (8mm), CAR-10-US (10mm), CAR-12-US (12mm)
Code information (models, lots, serial numbers)
Catalog/UDI-DI: CAR-06-US/00852897002328, CAR-08-US/00852897002021, CAR-10-US/ 00852897002038, CAR-12-US/00852897002335. All lots Distributed from July 2016 to October 2024.
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- Cartiva Synthetic Cartilage Implant CAR-10-USCartiva, Inc · Prosthesis, Metatarsophalangeal Joint Cartilage Replacement Implant1 recall139 reports
- Cartiva Synthetic Cartilage Implant CAR-08-USCartiva, Inc · Prosthesis, Metatarsophalangeal Joint Cartilage Replacement Implant1 recall41 reports
- Cartiva Synthetic Cartilage Implant CAR-12-USCartiva, Inc · Prosthesis, Metatarsophalangeal Joint Cartilage Replacement Implant1 recall4 reports
- Cartiva Synthetic Cartilage Implant CAR-06-USCartiva, Inc · Prosthesis, Metatarsophalangeal Joint Cartilage Replacement Implant1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.