Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use… recallZ-0599-2025 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 21, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.
FDA root cause category: Under Investigation by firm
What the company is doing
Siemens issued Urgent Medical Device Correction (UMDC) ACHC25-02.A.US was issued to US affected customers by FedEx beginning on 10/28/2024 and Urgent Field Safety Notice (UFSN) ACHC25-02.A.OUS. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Manually dilute patient samples that are > 18.00 mg/L: 1. Navigate to the CH Test Definition screen. 2. Under Measuring Interval, uncheck Repeat when Outside Measuring Interval. 3. For samples that generate test results > 18.00 mg/L, prepare a 1:2 manual dilution with saline solution (0.9% unbuffered Isotonic saline). 4. From the Create Patient Order screen, order the sample and enter 2 in the manual dilution field on the left of the screen. Note: The system will apply the 1:2 dilution factor and calculate the final result. 5. Process the manually diluted sample. "Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. "Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution The Atellica CH B2M assay IFU will be updated with the above steps to support manual 1:2 dilution only with saline for samples > 18.00 mg/L. Please see Appendix, Table 2, for data using these revised instructions. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Recall details
- Recall number
- Z-0599-2025
- Status
- Open, Classified
- Initiated by the company
- Oct 21, 2024
- Posted by the FDA
- Dec 4, 2024
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Quantity
- 11330 units
- Product code
- JZG
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.
Product description
Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635
Code information (models, lots, serial numbers)
UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atellica CH B2M 11097635Siemens Healthcare Diagnostics Inc. · System, test, beta-2-microglobulin immunological2 recalls0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page