Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use… recallZ-0599-2025 · Siemens Healthcare Diagnostics, Inc. · initiated Oct 21, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for falsely depressed auto-diluted results for samples above the measuring interval of 18.00 mg/L, up to 30.00 mg/L, when using the Atellica CH 2-Microglobulin (B2M) reagent.

FDA root cause category: Under Investigation by firm

What the company is doing

Siemens issued Urgent Medical Device Correction (UMDC) ACHC25-02.A.US was issued to US affected customers by FedEx beginning on 10/28/2024 and Urgent Field Safety Notice (UFSN) ACHC25-02.A.OUS. Letter states reason for recall, health risk and action to take: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. " Manually dilute patient samples that are > 18.00 mg/L: 1. Navigate to the CH Test Definition screen. 2. Under Measuring Interval, uncheck Repeat when Outside Measuring Interval. 3. For samples that generate test results > 18.00 mg/L, prepare a 1:2 manual dilution with saline solution (0.9% unbuffered Isotonic saline). 4. From the Create Patient Order screen, order the sample and enter 2 in the manual dilution field on the left of the screen. Note: The system will apply the 1:2 dilution factor and calculate the final result. 5. Process the manually diluted sample. "Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. "Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution The Atellica CH B2M assay IFU will be updated with the above steps to support manual 1:2 dilution only with saline for samples > 18.00 mg/L. Please see Appendix, Table 2, for data using these revised instructions. Contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.

Recall details

Recall number
Z-0599-2025
Status
Open, Classified
Initiated by the company
Oct 21, 2024
Posted by the FDA
Dec 4, 2024
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
11330 units
Product code
JZG
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hungary, India, Iran, Israel, Italy, Ivory Coast, Japan, Jordan, Kuwait, Mexico, Morocco, Oman, P.R. China, Pakistan, Paraguay, Poland, Portugal, Saudi Arabia, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Switzerland, Thailand, Turkey, U.A.E., Uganda, United Kingdom, Uruguay, Uzbekistan, Vietnam, Zaire.

Product description

Atellica CH 2-Microglobulin (B2M) - In vitro diagnostic use in the quantitative determination of 2-microglobulin in human serum and plasma. Siemens Material Number (SMN): 11097635

Code information (models, lots, serial numbers)

UDI-DI: 00630414595566 All in-date and future Atellica CH B2M reagent lots