Skip to content

Indeflator recallZ-0730-2025 · Abbott Vascular Inc · initiated Nov 6, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.

FDA root cause category: Component change control

What the company is doing

On 11/6/2024, recall notices were distributed to customers asking them to take the following actions: 1) Stop using affected devices. 2) Complete and return the acknowledgement form. 3) Return unused devices to the firm. 4) Share this notification with relevant personnel in your organization. 5) If you have further distributed/transferred the affected products, notify those customers. If you have any questions contact firm customer service at 800-227-9902

Recall details

Recall number
Z-0730-2025
Status
Open, Classified
Initiated by the company
Nov 6, 2024
Posted by the FDA
Dec 20, 2024
Recalling firm
Abbott Vascular Inc
Quantity
1,418
Product code
MAV
FDA submissions
K961471
Distribution
US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.

Product description

20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatation catheters to create and monitor pressure in the balloon.

Code information (models, lots, serial numbers)

UDI-DI: 08717648013591, Lots: 60418466, 60418467, 60418468

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95707).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page