Indeflator recallZ-0733-2025 · Abbott Vascular Inc · initiated Nov 6, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Inflation Device used during cardiovascular procedures may leak in the indeflator at the hose rotating luer or at the gauge, which could lead to air embolism.
FDA root cause category: Component change control
What the company is doing
On 11/6/2024, recall notices were distributed to customers asking them to take the following actions: 1) Stop using affected devices. 2) Complete and return the acknowledgement form. 3) Return unused devices to the firm. 4) Share this notification with relevant personnel in your organization. 5) If you have further distributed/transferred the affected products, notify those customers. If you have any questions contact firm customer service at 800-227-9902
Recall details
- Recall number
- Z-0733-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 6, 2024
- Posted by the FDA
- Dec 20, 2024
- Recalling firm
- Abbott Vascular Inc
- Quantity
- 4,975
- Product code
- MAV
- FDA submissions
- K961471
- Distribution
- US: IA, MD, DC, IN, CA, NV, TX, CT, GA, MS, NY, OH, FL, NC, MI, KY, PA, VA, IL, OK, TN, UT, WA, WI, MT, OR, LA, HI, NE, AR, MO, MA, KS, ID, AL, AZ, PR, CO, SC. OUS: IN, KR, SG, TH, VN, AR, BR, CA, CO, PA, UY, BE, CZ, DE, ES, FR, GB, IT, PT, RE, SE, SK, TM, JP.
Product description
20/30 Priority Pack Kit/.115 RHV, REF: 1000186-115, used for cardiovascular procedures.
Code information (models, lots, serial numbers)
UDI-DI: 08717648015274, Lots: 60564920, 60564933, 60566431, 60566432, 60566433, 60566667
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Indeflator 1000186-115Abbott Vascular Inc. · Syringe, balloon inflation2 recalls124 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95707).
- Z-0730-2025 · Open, Classified20/30 INDEFLATOR, REF: 1000184, for use during cardiovascular procedures in conjunction with interventional devices such as balloon dilatation catheters to create and monitor press…
- Z-0731-2025 · Open, Classified20/30 Priority Pack Accessory Kit/.096 RHV, REF: 1000186, or use during vascular procedures in conjunction with interventional and / or diagnostic devices (e.g., balloon dilatation…
- Z-0732-2025 · Open, Classified20/30 Priority Pack Accessory Kit w/Copilot, REF: 1003327, used for cardiovascular procedures.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page