Skip to content

LNCS-II rainbow recallZ-1184-2025 · Physio-Control, Inc. · initiated Jan 21, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Physio-Control, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.

FDA root cause category: Device Design

What the company is doing

On 1/21/2025, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day mail to customers to inform them that the LIFEPAK 15 V4 displays an error where "SpO2: Sensor does not support SpCO or SpMET". The error results in preventing users from utilizing LIFEPAK 15 V4 SpCO and SpMet functionalities when connected to specific lots of Masimo Rainbow Sensors. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form and remove them from their point of use. 2. Return the enclosed business reply form by email to [email protected] to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please keep this communication close to the affected items until replacement. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. For questions or concerns - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at [email protected].

Recall details

Recall number
Z-1184-2025
Status
Open, Classified
Initiated by the company
Jan 21, 2025
Posted by the FDA
Feb 19, 2025
Recalling firm
Physio-Control, Inc.
Quantity
622
Product code
MKJ
FDA submissions
P160026
Distribution
U.S. Nationwide distribution including in the states of CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, MA, MD, MI, MO, MS, NC, NH, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WV.

Product description

LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

Code information (models, lots, serial numbers)

Catalog Number: 11996-000515 UDI-DI/GTIN code: 00843997010863 Lot Number: 23G79 Catalog Number: 11996-000519 UDI-DI/GTIN code: 00843997015608 Lot Numbers: 23HER 23HPT 23JBN 23JTA Catalog Number: 11996-000520 UDI-DI/GTIN code: 00843997015615 Lot Numbers: 23HNV 23JBP

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

27 devices cleared under P160026 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page