Revolution recallZ-0607-2025 · GE Healthcare (China) Co., Ltd. · initiated Nov 5, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Healthcare (China) Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
FDA root cause category: Component design/selection
What the company is doing
GE HealthCare issued an URGENT MEDICAL DEVICE notice to its consignees on 11/05/2024 via mailing using a traceable means. The notice explained the issue, potential safety issue, and requested the following actions be taken: " You can continue to use your system. When utilizing helical, cine, and cardiac scan modes, review the image data set for any potential rotation. When using these modes during interventional procedures or radiation therapy planning, adhere to standard clinical quality checks and workflows. If image rotation is detected, consult with the physician to determine whether a rescan is necessary or if the images need to be corrected during post-processing to ensure accurate interpretation and planning. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions." A GE HealthCare representative will contact the consignee to arrange for the correction. For questions or concerns: please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Recall details
- Recall number
- Z-0607-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 5, 2024
- Posted by the FDA
- Dec 4, 2024
- Recalling firm
- GE Healthcare (China) Co., Ltd.
- Quantity
- 4 units
- Product code
- JAK
- Distribution
- Worldwide distribution.
Product description
GE Healthcare Revolution Eagle, Model Number 5849504-3, X-ray/computed tomography system
Code information (models, lots, serial numbers)
UDI/DI 00195278385215, Serial/Sales Order Numbers: CBDPG2400006HM, CBDPG2400005HM, CBDPG2400004HM, CBDPG2400007HM
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Revolution AdvanceGE Hangwei Medical Systems Company Ltd. · System, x-ray, tomography, computed3 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95704).
- Z-0600-2025 · Open, ClassifiedGE Healthcare Discovery RT, Model Number 2374681-17, X-ray/computed tomography systems labeled as:
- Z-0601-2025 · Open, ClassifiedGE Healthcare Optima 520, Catalog number 5439126; X-ray/computed tomography systems
- Z-0602-2025 · Open, ClassifiedGE Healthcare Optima 680 Expert, Model number 6670000; X-ray/computed tomography system
- Z-0603-2025 · Open, ClassifiedGE Healthcare Revolution ACE, Model number 6670000-3; X-ray/computed tomography systems
- Z-0604-2025 · Open, ClassifiedGE Healthcare Revolution Advance, Model number 5849504; X-ray/computed tomography systems
- Z-0605-2025 · Open, ClassifiedGE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system
- Z-0606-2025 · Open, ClassifiedGE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
- Z-0608-2025 · Open, ClassifiedGE Healthcare Revolution EVO, Model Number 5454001-260; X-ray/computed tomography system
- Z-0609-2025 · Open, ClassifiedGE Healthcare Revolution Frontier, Model Numbers: 1) 5232085-213, 2) 5873009; X-ray/computed tomography system
- Z-0610-2025 · Open, ClassifiedGE Healthcare Revolution Maxima, Model Numbers: 1) 6665000-2, 2) 6670000-2, 3) 6670000-200, 4) 6670000-6; X-ray/computed tomography system
- Z-0611-2025 · Open, ClassifiedGE Healthcare Revolution Maxima M, Model Number, 6670000-2; X-ray/computed tomography system
- Z-0612-2025 · Open, ClassifiedGE Healthcare Revolution Maxima Select, Model Number, 6670000-55; X-ray/computed tomography system
- Z-0613-2025 · Open, ClassifiedGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B79632DA: EU: RT Refresh Package, Non-EU: Discovery RT, Optima CT580, or Discovery CT590 RT; X-ray/comput…
- Z-0614-2025 · Open, ClassifiedGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80212DA: EU: Operator Console (for BrightSpeed, Elite/BrightSpeed Elite Select), Non-EU: BrightSpeed Elite…
- Z-0615-2025 · Open, ClassifiedGE Healthcare Operator Console Upgrade Types, FRU Spare Parts Number B80522DA: China Only: Revolution Discovery CT
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.