Eclipse PRO REF 98700 recallZ-0684-2025 · Spacelabs Healthcare, Ltd. · initiated Nov 25, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Spacelabs Healthcare, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
FDA root cause category: Process control
What the company is doing
On 11/25/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers inform them that the Eclipse Pro and Eclipse MINI cardiac Holter monitors may have mismatched serial numbers between the label and what is programmed in the devices. This is due to a manufacturing error. Customers are instructed to: complete the following steps for each Eclipse device under your control. Following these steps will allow you to check if your device is affected. 1. Visit https://eclipseutils.cardiology.spacelabs.com/ 2. The first time you use this site you will need to register for an account. Click Register and create a username and password. 3. Plug Eclipse (Mini or Pro) device into the computer via USB. 4. Select Serial Check in the menu and click CONNECT NEW RECORDER 5. Select the device from the list and click Connect 6. Click VERIFY SERIAL NUMBER once the device is connected 7. When prompted, enter the serial number printed on the label of the device and click CONTINUE 8. If the serial number you entered on the website does not match the serial number loaded into the device, you will see the warning. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, click GO BACK . If correctly entered the serial number click CONFIRM . 9. If the serial number entered on the website matches the serial number loaded into the device, you will see the confirmation below. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, please click GO BACK . If you have correctly entered the serial number click CONFIRM . 10. Repeat the previous steps for all your eclipse devices. Questions or assistance - contact Spacelabs service at 1-800-522-7025 or visit us at https://spacelabshealthcare.com/support/ Follow-up Updated Customer Letter sent 04/04/2025
Recall details
- Recall number
- Z-0684-2025
- Status
- Open, Classified
- Initiated by the company
- Nov 25, 2024
- Posted by the FDA
- Dec 17, 2024
- Recalling firm
- Spacelabs Healthcare, Ltd.
- Quantity
- 13,202 devices
- Product code
- MWJ
- FDA submissions
- K211651
- Distribution
- U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerland Bolivia Germany Netherlands Taiwan Brazil Greece Oman Trinidad and Tobago Brunei Darussalam Guatemala Panama Turkey Canada Honduras Peru United Arab Emirates Chile Hong Kong Poland United Kingdom China Iceland Portugal Columbia India Venezuela Congo Ireland Qatar Viet Nam Costa Rica Israel Republic of Macedonia Croatia Italy Roumanie Czech Republic Korea, Republic of Saudi Arabia
Product description
Eclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
Code information (models, lots, serial numbers)
Model Number: 98700 UDI-DI code: 10841522128851 Serial Numbers: All serial numbers manufactured between April 2021 to November 2024. Serial Numbers pending.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Eclipse ProSpacelabs Healthcare (Washington), Inc · Electrocardiograph, ambulatory (without analysis)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 95786).
- Z-0685-2025 · Open, ClassifiedEclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch recorder intended to record the patient s electrocardiogram.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page