Eclipse Mini REF 98900 recallZ-0685-2025 · Spacelabs Healthcare, Ltd. · initiated Nov 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.

FDA root cause category: Process control

What the company is doing

On 11/25/2024, the firm sent an "URGENT MEDICAL DEVICE CORRECTION" Letter to customers inform them that the Eclipse Pro and Eclipse MINI cardiac Holter monitors may have mismatched serial numbers between the label and what is programmed in the devices. This is due to a manufacturing error. Customers are instructed to: complete the following steps for each Eclipse device under your control. Following these steps will allow you to check if your device is affected. 1. Visit https://eclipseutils.cardiology.spacelabs.com/ 2. The first time you use this site you will need to register for an account. Click Register and create a username and password. 3. Plug Eclipse (Mini or Pro) device into the computer via USB. 4. Select Serial Check in the menu and click CONNECT NEW RECORDER 5. Select the device from the list and click Connect 6. Click VERIFY SERIAL NUMBER once the device is connected 7. When prompted, enter the serial number printed on the label of the device and click CONTINUE 8. If the serial number you entered on the website does not match the serial number loaded into the device, you will see the warning. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, click GO BACK . If correctly entered the serial number click CONFIRM . 9. If the serial number entered on the website matches the serial number loaded into the device, you will see the confirmation below. Confirm that you do not have a typo in the serial number you entered on the website. If you have a typo, please click GO BACK . If you have correctly entered the serial number click CONFIRM . 10. Repeat the previous steps for all your eclipse devices. Questions or assistance - contact Spacelabs service at 1-800-522-7025 or visit us at https://spacelabshealthcare.com/support/ Follow-up Updated Customer Letter sent 04/04/2025

Recall details

Recall number
Z-0685-2025
Status
Open, Classified
Initiated by the company
Nov 25, 2024
Posted by the FDA
Dec 17, 2024
Recalling firm
Spacelabs Healthcare, Ltd.
Quantity
1,779 devices
Product code
MWJ
FDA submissions
K212317
Distribution
U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerland Bolivia Germany Netherlands Taiwan Brazil Greece Oman Trinidad and Tobago Brunei Darussalam Guatemala Panama Turkey Canada Honduras Peru United Arab Emirates Chile Hong Kong Poland United Kingdom China Iceland Portugal Columbia India Venezuela Congo Ireland Qatar Viet Nam Costa Rica Israel Republic of Macedonia Croatia Italy Roumanie Czech Republic Korea, Republic of Saudi Arabia

Product description

Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch recorder intended to record the patient s electrocardiogram.

Code information (models, lots, serial numbers)

Model Number: 98900 UDI-DI code: 10841522128769 Serial Numbers: All serial numbers manufactured between May 2021 to November 2024. Serial Numbers pending.