MISHA Knee System Implant Small, Left, REF: 2-1001 recallZ-0762-2025 · Moximed, Inc. · initiated Dec 2, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Moximed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Knee implant may fracture due to a supplier-related issue with one component that has nonconforming geometry at the distal end of the absorber component that may result in a reduced wall thickness and microcracks, which may lead to new or worsening symptoms of discomfort, swelling / edema, pain or stiffness, which may necessitate removal of the device.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On 12/2/2034, recall notices were sent to customers who were asked to do the following: Recipients of implants from the Affected Lots can continue to use their implant as usual. However, you should evaluate patients who present with any new or worsening symptoms of discomfort, swelling / edema, pain or stiffness. If a device issue is suspected or you or your patient is concerned, X-rays may be useful to evaluate the device. Decisions about removing the device should be made on a case-by-case basis between health care providers and patient or caregiver. For patients who are identified as having a fractured absorber component, the surgeon should consider explantation surgery based on their independent clinical judgment. Firm is providing surgeons with a draft letter for their patients who have been implanted with affected Knee Systems from the Affected Lots. Firm recommends that surgeons customize the letter, send it to patients implanted with affected devices, and discuss with patients as appropriate. Share this Notice with all device users within your facility and network to ensure they are aware of this recall. This notification should also be shared with any organization if the care of patients with potentially affected devices has been transferred. Firm is prepared to assist patients with certain out-of-pocket costs related to this recall. Please correspond with the email address, [email protected], or call us at 877-204-2922. Complete and return the response form via email to [email protected]
Recall details
- Recall number
- Z-0762-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 2, 2024
- Posted by the FDA
- Dec 23, 2024
- Recalling firm
- Moximed, Inc.
- Quantity
- 115
- Product code
- QVV
- Distribution
- US Nationwide distribution in the states of CA, OR, NY, OH, TX, CO, NY, VA.
Product description
MISHA Knee System Implant Small, Left, REF: 2-1001. MISHA Knee System Implant Small, Right, REF: 2-1002.
Code information (models, lots, serial numbers)
REF/UDI-DI/Lot(Expiration): 2-1001/00856047005795/23032901(29-Mar-25), 23092102(21-Sep-25), 24032601(26-Mar-26). 2-1002/00856047005801/23032902(29-Mar-25), 23092503(25-Sep-25), 24040801(8-Apr-26).
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- MISHA® Knee System, Small, Left 2-1001Moximed, Inc. · Medial Knee Implanted Shock Absorber1 recall8 reports
- MISHA® Knee System, Small, Right 2-1002Moximed, Inc. · Medial Knee Implanted Shock Absorber1 recall6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.