Xhibit Telemetry Receiver with Software, Model 96280 recallZ-0810-2025 · Spacelabs Healthcare, Inc. · initiated Nov 25, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Spacelabs Healthcare, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.

FDA root cause category: Software change control

What the company is doing

On 11/25/2024, correction notices were emailed to customers who were asked to do the following: Firm will contact you at the earliest possible date to schedule a convenient time to update affected devices to software version 1.4.2 for XTR systems which addresses this defect. You can also contact GTS to have an FSE paged out for discussing upgrade schedule: 1-800-522-7025 and select 2 for Technical Support. Complete and submit the customer acknowledgement form: https://spacelabshealthcare.com/support/recall-reply-form/recall-reply-form-xtr-15-nov-2024/ For additional information or technical assistance, or to report an adverse event, contact: Technical Support: 1-800-522-7025 and select 2

Recall details

Recall number
Z-0810-2025
Status
Open, Classified
Initiated by the company
Nov 25, 2024
Posted by the FDA
Jan 7, 2025
Recalling firm
Spacelabs Healthcare, Inc.
Quantity
434
Product code
MHX
FDA submissions
K141156
Distribution
US: MN, TX, TN, SC, OK, NM, GA, PA, OH, IA, AL, CO, AR, FL, MO, ME, LA, IL, WA, MA, WI, OR, MS, ID, NY, KS, PR. OUS: Canada, Taiwan, Czech Republic, France, Greece, Italy, Kuwait, Philippines, Poland.

Product description

Xhibit Telemetry Receiver with Software, Model 96280. Intended to provide monitoring system with patient data.

Code information (models, lots, serial numbers)

Software Versions: 1.4.0 and 1.4.1. UDI-DI: 10841522107177, 10841522100246