Cardinal Health Presource Kits recallZ-1126-2025 · Cardinal Health 200, LLC · initiated Nov 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

FDA root cause category: Material/Component Contamination

What the company is doing

An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Recall details

Recall number
Z-1126-2025
Status
Open, Classified
Initiated by the company
Nov 8, 2024
Posted by the FDA
Feb 10, 2025
Recalling firm
Cardinal Health 200, LLC
Quantity
973,785 total units
Product code
LRO
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.

Product description

Cardinal Health Presource Kits: 1) KIT,DIEP FLAP , Catalog Number:PB23DFCM3; 2) KIT,DIEP FLAP , Catalog Number:PB23DFCM4; 3) KIT,DIEP FLAP , Catalog Number:PB23DFCM5; 4) KIT,DIEP FLAP , Catalog Number:PB23DFCM6; 5) KIT,DIEP FLAP , Catalog Number:PB23DFCM7; 6) KIT,DIEP FLAP , Catalog Number:PB23DFCMA; 7) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ10; 8) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ11; 9) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ12; 10) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ13; 11) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ14; 12) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ15; 13) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ16; 14) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ17; 15) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ18; 16) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ19; 17) KIT, HEAD AND NECK , Catalog Number:PE40HNAJ20; 18) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU1; 19) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU2; 20) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU3; 21) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU4; 22) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU5; 23) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU6; 24) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU7; 25) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU8; 26) KIT, HEAD AND NECK , Catalog Number:PE40HNAJU9; 27) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUA; 28) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUB; 29) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUC; 30) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUD; 31) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUE; 32) KIT, HEAD AND NECK , Catalog Number:PE40HNAJUF; 33) KIT, HEAD AND NECK , Catalog Number:PG40HNAJU2; 34) KIT, HEAD AND NECK , Catalog Number:PG40HNAJU3; 35) KIT, HEAD AND NECK , Catalog Number:PG40HNAJU4; 36) KIT, HEAD AND NECK , Catalog Number:PG40HNAJU5; 37) KIT, HEAD AND NECK , Catalog Number:PG40HNAJUC; 38) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ10; 39) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ11; 40) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ12; 41) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ13; 42) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ14; 43) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ15; 44) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ16; 45) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ17; 46) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ18; 47) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ19; 48) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ20; 49) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ21; 50) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ22; 51) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ23; 52) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJ24; 53) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU4; 54) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU5; 55) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU6; 56) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU7; 57) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU8; 58) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJU9; 59) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJUE; 60) KIT, FREE FLAP HEAD , Catalog Number:PG40HNJUF; 61) KIT, WILLIAMS SPINE , Catalog Number:PN23WGME1; 62) KIT, WILLIAMS SPINE , Catalog Number:PN23WGME2; 63) KIT, WILLIAMS SPINE , Catalog Number:PN23WGME3; 64) KIT, WILLIAMS SPINE , Cat

Code information (models, lots, serial numbers)

Show the full code information (92k characters)

1) PB23DFCM3, UDI-DI:10888439587050 (each) 50888439587058 (case), Lot Number: J37U13 ; 2) PB23DFCM4, UDI-DI:10888439601053 (each) 50888439601051 (case), Lot Number: J37U14 ; 3) PB23DFCM5, UDI-DI:10888439641790 (each) 50888439641798 (case), Lot Number: J46C08 ; 4) PB23DFCM6, UDI-DI:10888439836172 (each) 50888439836170 (case), Lot Number: J6050A ; 5) PB23DFCM7, UDI-DI:10888439853087 (each) 50888439853085 (case), Lot Number: J7704J ; 6) PB23DFCMA, UDI-DI:10887488531298 (each) 50887488531296 (case), Lot Number: J4952P ; 7) PB23DFCMA, UDI-DI:10887488531298 (each) 50887488531296 (case), Lot Number: J53W77 ; 8) PB23DFCMA, UDI-DI:10887488531298 (each) 50887488531296 (case), Lot Number: J5679N ; 9) PB23DFCMA, UDI-DI:10887488531298 (each) 50887488531296 (case), Lot Number: J84K06 ; 10) PB23DFCMA, UDI-DI:10887488531298 (each) 50887488531296 (case), Lot Number: J85W31 ; 11) PE40HNAJ10, UDI-DI:10195594746513 (each) 50195594746511 (case), Lot Number: 5918338 ; 12) PE40HNAJ11, UDI-DI:10195594777685 (each) 50195594777683 (case), Lot Number: 6279857 ; 13) PE40HNAJ12, UDI-DI:10195594808051 (each) 50195594808059 (case), Lot Number: 5649225 ; 14) PE40HNAJ13, UDI-DI:10195594878672 (each) 50195594878670 (case), Lot Number: 6279859 ; 15) PE40HNAJ14, UDI-DI:10197106057902 (each) 50197106057900 (case), Lot Number: 6487892 ; 16) PE40HNAJ15, UDI-DI:10197106116449 (each) 50197106116447 (case), Lot Number: 6487896 ; 17) PE40HNAJ16, UDI-DI:10197106138632 (each) 50197106138630 (case), Lot Number: 5 ; 18) PE40HNAJ16, UDI-DI:10197106138632 (each) 50197106138630 (case), Lot Number: 6560940 ; 19) PE40HNAJ17, UDI-DI:10197106224205 (each) 50197106224203 (case), Lot Number: 6568677 ; 20) PE40HNAJ18, UDI-DI:10197106480168 (each) 50197106480166 (case), Lot Number: 7197963 ; 21) PE40HNAJ19, UDI-DI:10197106571453 (each) 50197106571451 (case), Lot Number: 7280098 ; 22) PE40HNAJ20, UDI-DI:10197106587300 (each) 50197106587308 (case), Lot Number: 7287558 ; 23) PE40HNAJU1, UDI-DI:10195594142469 (each) 50195594142467 (case), Lot Number: J9602A ; 24) PE40HNAJU2, UDI-DI:10195594240776 (each) 50195594240774 (case), Lot Number: J9699U ; 25) PE40HNAJU3, UDI-DI:10195594358556 (each) 50195594358554 (case), Lot Number: 5324049 ; 26) PE40HNAJU4, UDI-DI:10195594448806 (each) 50195594448804 (case), Lot Number: 5533323 ; 27) PE40HNAJU5, UDI-DI:10195594541224 (each) 50195594541222 (case), Lot Number: 5565994 ; 28) PE40HNAJU6, UDI-DI:10195594585716 (each) 50195594585714 (case), Lot Number: 5664093 ; 29) PE40HNAJU7, UDI-DI:10195594629632 (each) 50195594629630 (case), Lot Number: 5775805 ; 30) PE40HNAJU8, UDI-DI:10195594666194 (each) 50195594666192 (case), Lot Number: 5918339 ; 31) PE40HNAJU9, UDI-DI:10195594700324 (each) 50195594700322 (case), Lot Number: 6022074 ; 32) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: J8393H ; 33) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 5858463 ; 34) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6702004 ; 35) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6702005 ; 36) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6909986 ; 37) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6416698 ; 38) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6987101 ; 39) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 7287561 ; 40) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 7280096 ; 41) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: J7905A ; 42) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: J8064N ; 43) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 5631065 ; 44) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 6279860 ; 45) PE40HNAJUA, UDI-DI:10888439948394 (each) 50888439948392 (case), Lot Number: 5858463 ; 46) PE40HNAJUB, UDI-DI:10197106566732 (each) 50197106566730 (case), Lot Number: 7280096 ; 47) PE40HNAJUC, UDI-DI:10197106630389 (each) 50197106630387 (case), Lot Number: 7287559 ; 48) PE40HNAJUC, UDI-DI:10197106630389 (each) 50197106630387 (case), Lot Number: 7331327 ; 49) PE40HNAJUD, UDI-DI:10197106663776 (each) 50197106663774 (case), Lot Number: 7332883 ; 50) PE40HNAJUE, UDI-DI:10197106692929 (each) 50197106692927 (case), Lot Number: 7332890 ; 51) PE40HNAJUE, UDI-DI:10197106692929 (each) 50197106692927 (case), Lot Number: 7287559 ; 52) PE40HNAJUF, UDI-DI:10197106755235 (each) 50197106755233 (case), Lot Number: 7332892 ; 53) PE40HNAJUF, UDI-DI:10197106755235 (each) 50197106755233 (case), Lot Number: 6779831 ; 54) PG40HNAJU2, UDI-DI:10888439568592 (each) 50888439568590 (case), Lot Number: J37U04 ; 55) PG40HNAJU3, UDI-DI:10888439786200 (each) 50888439786208 (case), Lot Number: J5678G ; 56) PG40HNAJU3, UDI-DI:10888439786200 (each) 50888439786208 (case), Lot Number: J5678J ; 57) PG40HNAJU3, UDI-DI:10888439786200 (each) 50888439786208 (case), Lot Number: J6278K ; 58) PG40HNAJU4, UDI-DI:10888439879537 (each) 50888439879535 (case), Lot Number: J7904Z ; 59) PG40HNAJU5, UDI-DI:10888439903829 (each) 50888439903827 (case), Lot Number: J7904Z ; 60) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J37U05 ; 61) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J4219Y ; 62) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J4219Z ; 63) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J46A94 ; 64) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J46A95 ; 65) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J4951Y ; 66) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J7498X ; 67) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J8393H ; 68) PG40HNAJUC, UDI-DI:10888439342864 (each) 50888439342862 (case), Lot Number: J8393K ; 69) PG40HNJ10, UDI-DI:10195594042929 (each) 50195594042927 (case), Lot Number: J9602A ; 70) PG40HNJ11, UDI-DI:10195594116903 (each) 50195594116901 (case), Lot Number: 5140303 ; 71) PG40HNJ12, UDI-DI:10195594252014 (each) 50195594252012 (case), Lot Number: 5140306 ; 72) PG40HNJ13, UDI-DI:10195594298098 (each) 50195594298096 (case), Lot Number: 5140335 ; 73) PG40HNJ14, UDI-DI:10195594401672 (each) 50195594401670 (case), Lot Number: 5478801 ; 74) PG40HNJ15, UDI-DI:10195594471101 (each) 50195594471109 (case), Lot Number: 5565993 ; 75) PG40HNJ16, UDI-DI:10195594531249 (each) 50195594531247 (case), Lot Number: J97W47 ; 76) PG40HNJ17, UDI-DI:10195594561390 (each) 50195594561398 (case), Lot Number: 5664091 ; 77) PG40HNJ18, UDI-DI:10195594631321 (each) 50195594631329 (case), Lot Number: 5140301 ; 78) PG40HNJ19, UDI-DI:10195594700232 (each) 50195594700230 (case), Lot Number: 6022074 ; 79) PG40HNJ20, UDI-DI:10195594773083 (each) 50195594773081 (case), Lot Number: 6025778 ; 80) PG40HNJ21, UDI-DI:10195594791742 (each) 50195594791740 (case), Lot Number: 6279857 ; 81) PG40HNJ22, UDI-DI:10195594834920 (each) 50195594834928 (case), Lot Number: 6025781 ; 82) PG40HNJ23, UDI-DI:10195594867980 (each) 50195594867988 (case), Lot Number: 6416697 ; 83) PG40HNJ24, UDI-DI:10197106013274 (each) 50197106013272 (case), Lot Number: 5970332 ; 84) PG40HNJU4, UDI-DI:10888439567311 (each) 50888439567319 (case), Lot Number: J37U04 ; 85) PG40HNJU5, UDI-DI:10888439616682 (each) 50888439616680 (case), Lot Number: J4219Z ; 86) PG40HNJU6, UDI-DI:10888439689907 (each) 50888439689905 (case), Lot Number: J5678H ; 87) PG40HNJU7, UDI-DI:10888439786446 (each) 50888439786444 (case), Lot Number: J4951Y ; 88) PG40HNJU8, UDI-DI:10888439854329 (each) 50888439854327 (case), Lot Number: J79H50 ; 89) PG40HNJU9, UDI-DI:10888439963571 (each) 50888439963579 (case), Lot Number: J8393H ; 90) PG40HNJUE, UDI-DI:10888439342833 (each) 50888439342831 (case), Lot Number: J46A95 ; 91) PG40HNJUE, UDI-DI:10888439342833 (each) 50888439342831 (case), Lot Number: J6366T ; 92) PG40HNJUE, UDI-DI:10888439342833 (each) 50888439342831 (case), Lot Number: J7904Z ; 93) PG40HNJUF, UDI-DI:10888439948400 (each) 50888439948408 (case), Lot Number: J6278P ; 94) PG40HNJUF, UDI-DI:10888439948400 (each) 50888439948408 (case), Lot Number: J7905A ; 95) PN23WGME1, UDI-DI:10197106553213 (each) 50197106553211 (case), Lot Number: 7055825 ; 96) PN23WGME1, UDI-DI:10197106553213 (each) 50197106553211 (case), Lot Number: 7280093 ; 97) PN23WGME2, UDI-DI:10197106594155 (each) 50197106594153 (case), Lot Number: 7280048 ; 98) PN23WGME2, UDI-DI:10197106594155 (each) 50197106594153 (case), Lot Number: 7280091 ; 99) PN23WGME3, UDI-DI:10197106735046 (each) 50197106735044 (case), Lot Number: 7332831 ; 100) PN23WGME3, UDI-DI:10197106735046 (each) 50197106735044 (case), Lot Number: 7349672 ; 101) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7055826 ; 102) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7280092 ; 103) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7280055 ; 104) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7287533 ; 105) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7280061 ; 106) PN23WGMEA, UDI-DI:10197106508008 (each) 50197106508006 (case), Lot Number: 7287534 ; 107) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7287466 ; 108) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331338 ; 109) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7287483 ; 110) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331319 ; 111) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331284 ; 112) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331323 ; 113) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331456 ; 114) PN23WGMEB, UDI-DI:10197106557570 (each) 50197106557578 (case), Lot Number: 7331313 ; 115) PN23WGMEC, UDI-DI:10197106770887 (each) 50197106770885 (case), Lot Number: 7280043 ; 116) PN23WGMEC, UDI-DI:10197106770887 (each) 50197106770885 (case), Lot Number: 7331320 ; 117) PN43SPUN1, UDI-DI:10197106570739 (each) 50197106570737 (case), Lot Number: 7287491 ; 118) PN43SPUN2, UDI-DI:10197106592663 (each) 50197106592661 (case), Lot Number: 7163141 ; 119) PN43SPUN3, UDI-DI:10197106647974 (each) 50197106647972 (case), Lot Number: 7331296 ; 120) PN43SPUNA, UDI-DI:10197106200025 (each) 50197106200023 (case), Lot Number: 6116185 ; 121) PN43SPUNB, UDI-DI:10197106542385 (each) 50197106542383 (case), Lot Number: 6701984 ; 122) PN43SPUNB, UDI-DI:10197106542385 (each) 50197106542383 (case), Lot Number: 7331296 ; 123) PN43SPUNB, UDI-DI:10197106542385 (each) 50197106542383 (case), Lot Number: 7331430 ; 124) PN43SPUNB, UDI-DI:10197106542385 (each) 50197106542383 (case), Lot Number: 7331430 ; 125) PN43SPUNB, UDI-DI:10197106542385 (each) 50197106542383 (case), Lot Number: 7332848 ; 126) PN4EACWIA, UDI-DI:10197106335192 (each) 50197106335190 (case), Lot Number: 7122012 ; 127) PN4EACWIA, UDI-DI:10197106335192 (each) 50197106335190 (case), Lot Number: 7160478 ; 128) PN4EACWIA, UDI-DI:10197106335192 (each) 50197106335190 (case), Lot Number: 7260071 ; 129) PN4EACWIB, UDI-DI:10197106525913 (each) 50197106525911 (case), Lot Number: 6560923 ; 130) PW40HNJU1, UDI-DI:10197106158913 (each) 50197106158911 (case), Lot Number: 6560940 ; 131) PW40HNJU2, UDI-DI:10197106226179 (each) 50197106226177 (case), Lot Number: 5 ; 132) PW40HNJU2, UDI-DI:10197106226179 (each) 50197106226177 (case), Lot Number: 6568677 ; 133) PW40HNJU3, UDI-DI:10197106304631 (each) 50197106304639 (case), Lot Number: 6702004 ; 134) PW40HNJU4, UDI-DI:10197106386293 (each) 50197106386291 (case), Lot Number: 6717949 ; 135) PW40HNJU5, UDI-DI:10197106454565 (each) 50197106454563 (case), Lot Number: 6779831 ; 136) PW40HNJU6, UDI-DI:10197106496633 (each) 50197106496631 (case), Lot Number: 6909988 ; 137) PW40HNJU7, UDI-DI:10197106579978 (each) 50197106579976 (case), Lot Number: 7280095 ; 138) PW40HNJUA, UDI-DI:10197106014189 (each) 50197106014187 (case), Lot Number: 7280097 ; 139) PW40HNJUA, UDI-DI:10197106014189 (each) 50197106014187 (case), Lot Number: 7331340 ; 140) PW40HNJUB, UDI-DI:10197106630341 (each) 50197106630349 (case), Lot Number: 7331327 ; 141) PW40HNJUC, UDI-DI:10197106663745 (each) 50197106663743 (case), Lot Number: 7332886 ; 142) PW40HNJUD, UDI-DI:10197106692875 (each) 50197106692873 (case), Lot Number: 7332883 ; 143) PW40HNJUE, UDI-DI:10197106758601 (each) 50197106758609 (case), Lot Number: 7332892 ; 144) SBA30MFOCA, UDI-DI:10195594120450 (each) 50195594120458 (case), Lot Number: J9699T ; 145) SBA30MFOCA, UDI-DI:10195594120450 (each) 50195594120458 (case), Lot Number: 5324049 ; 146) SBA30MFOCB, UDI-DI:10195594378011 (each) 50195594378019 (case), Lot Number: 5405624 ; 147) SBA30MFOCB, UDI-DI:10195594378011 (each) 50195594378019 (case), Lot Number: 5508057 ; 148) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 5858463 ; 149) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 6560942 ; 150) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 5565993 ; 151) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: J97W47 ; 152) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 5664093 ; 153) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 5664091 ; 154) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 5664094 ; 155) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 6022075 ; 156) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 6416700 ; 157) SBA30MFOCC, UDI-DI:10195594446321 (each) 50195594446329 (case), Lot Number: 6487896 ; 158) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 6717950 ; 159) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 6717949 ; 160) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 7280095 ; 161) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 7332887 ; 162) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 7332883 ; 163) SBA30MFOCD, UDI-DI:10197106288559 (each) 50197106288557 (case), Lot Number: 7332888 ; 164) SBA30PEMD1, UDI-DI:10195594762902 (each) 50195594762900 (case), Lot Number: 6133661 ; 165) SBA38FTVAC, UDI-DI:10888439938210 (each) 50888439938218 (case), Lot Number: J6786C ; 166) SBA38FTVAD, UDI-DI:10195594972226 (each) 50195594972224 (case), Lot Number: 6701941 ; 167) SBA38FTVAD, UDI-DI:10195594972226 (each) 50195594972224 (case), Lot Number: 6701940 ; 168) SBA38FTVAD, UDI-DI:10195594972226 (each) 50195594972224 (case), Lot Number: 5664019 ; 169) SBA38FTVAD, UDI-DI:10195594972226 (each) 50195594972224 (case), Lot Number: 5891283 ; 170) SBA38FTVAE, UDI-DI:10197106422212 (each) 50197106422210 (case), Lot Number: 7122097 ; 171) SBA38FTVAE, UDI-DI:10197106422212 (each) 50197106422210 (case), Lot Number: 6485588 ; 172) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7360139 ; 173) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7280055 ; 174) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6701938 ; 175) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7005442 ; 176) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6701970 ; 177) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7260067 ; 178) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7287457 ; 179) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7332803 ; 180) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: J4218Z ; 181) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 7332814 ; 182) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: J4680J ; 183) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: J4950U ; 184) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: J6277R ; 185) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 5775788 ; 186) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 5891293 ; 187) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6022056 ; 188) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6264448 ; 189) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6345500 ; 190) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6485576 ; 191) SBA38LMCID, UDI-DI:10887488575568 (each) 50887488575566 (case), Lot Number: 6683719 ; 192) SBA40FPJUB, UDI-DI:10197106722121 (each) 50197106722129 (case), Lot Number: 6779832 ; 193) SBA40FPJUB, UDI-DI:10197106722121 (each) 50197106722129 (case), Lot Number: 7332890 ; 194) SBA43BSUNA, UDI-DI:10197106212455 (each) 50197106212453 (case), Lot Number: 6485593 ; 195) SBA4JDFCB1, UDI-DI:10197106730447 (each) 50197106730445 (case), Lot Number: 7349697 ; 196) SBA4JDFCBA, UDI-DI:10197106527900 (each) 50197106527908 (case), Lot Number: 7320242 ; 197) SBA4JDFCBC, UDI-DI:10197106654477 (each) 50197106654475 (case), Lot Number: 7320243 ; 198) SBA4JFDCM1, UDI-DI:10197106731321 (each) 50197106731329 (case), Lot Number: 7349697 ; 199) SBA4JFDCMA, UDI-DI:10197106518977 (each) 50197106518975 (case), Lot Number: 7320241 ; 200) SBA51FFK10, UDI-DI:10195594738402 (each) 50195594738400 (case), Lot Number: 6232004 ; 201) SBA51FFK11, UDI-DI:10195594890940 (each) 50195594890948 (case), Lot Number: 6364783 ; 202) SBA51FFK12, UDI-DI:10197106025451 (each) 50197106025459 (case), Lot Number: 6416684 ; 203) SBA51FFK13, UDI-DI:10197106192658 (each) 50197106192656 (case), Lot Number: 6416685 ; 204) SBA51FFK14, UDI-DI:10197106252178 (each) 50197106252176 (case), Lot Number: 6748352 ; 205) SBA51FFK15, UDI-DI:10197106298336 (each) 50197106298334 (case), Lot Number: 6470287 ; 206) SBA51FFK16, UDI-DI:10197106367636 (each) 50197106367634 (case), Lot Number: 6928777 ; 207) SBA51FFK17, UDI-DI:10197106512166 (each) 50197106512164 (case), Lot Number: 7280089 ; 208) SBA51FFK18, UDI-DI:10197106553718 (each) 50197106553716 (case), Lot Number: 7280093 ; 209) SBA51FFK19, UDI-DI:10197106593585 (each) 50197106593583 (case), Lot Number: 7280090 ; 210) SBA51FFK20, UDI-DI:10197106661352 (each) 50197106661350 (case), Lot Number: 7332856 ; 211) SBA51FFK21, UDI-DI:10197106753675 (each) 50197106753673 (case), Lot Number: 7349671 ; 212) SBA51FFK22, UDI-DI:10197106779521 (each) 50197106779529 (case), Lot Number: 7332878 ; 213) SBA51FFKC3, UDI-DI:10888439755107 (each) 50888439755105 (case), Lot Number: J6278G ; 214) SBA51FFKC4, UDI-DI:10888439773248 (each) 50888439773246 (case), Lot Number: J6278J ; 215) SBA51FFKC5, UDI-DI:10195594967352 (each) 50195594967350 (case), Lot Number: J9602H ; 216) SBA51FFKC6, UDI-DI:10195594067557 (each) 50195594067555 (case), Lot Number: 20M067262 ; 217) SBA51FFKC6, UDI-DI:10195594067557 (each) 50195594067555 (case), Lot Number: 5135994 ; 218) SBA51FFKC7, UDI-DI:10195594167332 (each) 50195594167330 (case), Lot Number: 5138327 ; 219) SBA51FFKC8, UDI-DI:10195594310912 (each) 50195594310910 (case), Lot Number: 5405617 ; 220) SBA51FFKC9, UDI-DI:10195594700027 (each) 50195594700025 (case), Lot Number: 6022070 ; 221) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J37T95 ; 222) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J43U72 ; 223) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J46A82 ; 224) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J74A02 ; 225) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J4951U ; 226) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J8064R ; 227) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J86N66 ; 228) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J70X09 ; 229) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: J70X08 ; 230) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5533310 ; 231) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5631062 ; 232) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5405618 ; 233) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5508052 ; 234) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5508054 ; 235) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5664049 ; 236) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 5533321 ; 237) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 6470285 ; 238) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 6470287 ; 239) SBA51FFKCE, UDI-DI:10885425756261 (each) 50885425756269 (case), Lot Number: 6563975 ; 240) SBA51FFKCF, UDI-DI:10197106344910 (each) 50197106344918 (case), Lot Number: 7005466 ; 241) SBA51MFKO1, UDI-DI:10197106800713 (each) 50197106800711 (case), Lot Number: 7331351 ; 242) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 6779960 ; 243) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 6788670 ; 244) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7160529 ; 245) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7163105 ; 246) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7266055 ; 247) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7160538 ; 248) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7160538 ; 249) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7280116 ; 250) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7320241 ; 251) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7280112 ; 252) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7332922 ; 253) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7332921 ; 254) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7349698 ; 255) SBA51MFKOA, UDI-DI:10197106223819 (each) 50197106223817 (case), Lot Number: 7349699 ; 256) SBA51PMKO9, UDI-DI:10197106490426 (each) 50197106490424 (case), Lot Number: 7265984 ; 257) SBA51PMKOE, UDI-DI:10197106223901 (each) 50197106223909 (case), Lot Number: 6775048 ; 258) SBA51PMKOF, UDI-DI:10197106527207 (each) 50197106527205 (case), Lot Number: 7332874 ; 259) SBA51PMKOF, UDI-DI:10197106527207 (each) 50197106527205 (case), Lot Number: 7280090 ; 260) SBA51PMKOF, UDI-DI:10197106527207 (each) 50197106527205 (case), Lot Number: 7287539 ; 261) SBA51PMKOF, UDI-DI:10197106527207 (each) 50197106527205 (case), Lot Number: 7332861 ; 262) SBA5FDLMM2, UDI-DI:10195594883492 (each) 50195594883490 (case), Lot Number: 5530913 ; 263) SBA5FDLMM3, UDI-DI:10197106248997 (each) 50197106248995 (case), Lot Number: 6907763 ; 264) SBA5FDLMMB, UDI-DI:10885425502578 (each) 50885425502576 (case), Lot Number: J4218Z ; 265) SBA5FDLMMB, UDI-DI:10885425502578 (each) 50885425502576 (case), Lot Number: J4950Y ; 266) SBA5FDLMMB, UDI-DI:10885425502578 (each) 50885425502576 (case), Lot Number: J4680A ; 267) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 7280059 ; 268) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 7055832 ; 269) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 7331287 ; 270) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 6560923 ; 271) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 7331457 ; 272) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 7332845 ; 273) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J7609N ; 274) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J7904N ; 275) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J8123N ; 276) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J8692M ; 277) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J96N39 ; 278) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: J9602N ; 279) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 5311244 ; 280) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 5138168 ; 281) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 5631042 ; 282) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 5664033 ; 283) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 5891282 ; 284) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 6025767 ; 285) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 6345502 ; 286) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 6560930 ; 287) SBA5FDLMMC, UDI-DI:10888439831726 (each) 50888439831724 (case), Lot Number: 6683719 ; 288) SBA69FFC18, UDI-DI:10888439587111 (each) 50888439587119 (case), Lot Number: J3593N ; 289) SBA69FFC19, UDI-DI:10888439650921 (each) 50888439650929 (case), Lot Number: J46C07 ; 290) SBA69FFC20, UDI-DI:10888439668124 (each) 50888439668122 (case), Lot Number: J4952N ; 291) SBA69FFC21, UDI-DI:10888439738261 (each) 50888439738269 (case), Lot Number: J53W77 ; 292) SBA69FFC22, UDI-DI:10888439852011 (each) 50888439852019 (case), Lot Number: J7704J ; 293) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J37U13 ; 294) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J53W78 ; 295) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J43U79 ; 296) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J7963U ; 297) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J7963U ; 298) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J85W32 ; 299) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J6278W ; 300) SBA69FFCIZ, UDI-DI:10888439507966 (each) 50888439507964 (case), Lot Number: J6368X ; 301) SBA69FFKW2, UDI-DI:10195594163426 (each) 50195594163424 (case), Lot Number: J9850H ; 302) SBA69FFKW3, UDI-DI:10195594381714 (each) 50195594381712 (case), Lot Number: 5508060 ; 303) SBA69FFKW4, UDI-DI:10195594717445 (each) 50195594717443 (case), Lot Number: 5970335 ; 304) SBA69FFKW5, UDI-DI:10197106209110 (each) 50197106209118 (case), Lot Number: 6701854 ; 305) SBA69FFKW6, UDI-DI:10197106395783 (each) 50197106395781 (case), Lot Number: 7160529 ; 306) SBA69FFKW7, UDI-DI:10197106716670 (each) 50197106716678 (case), Lot Number: 7332924 ; 307) SBA69FFKWM, UDI-DI:10888439589658 (each) 50888439589656 (case), Lot Number: J4952N ; 308) SBA69FFKWM, UDI-DI:10888439589658 (each) 50888439589656 (case), Lot Number: J5679N ; 309) SBA69FFKWM, UDI-DI:10888439589658 (each) 50888439589656 (case), Lot Number: J8693M ; 310) SBA69FFKWM, UDI-DI:10888439589658 (each) 50888439589656 (case), Lot Number: 5664108 ; 311) SBA69FFKWN, UDI-DI:10195594529284 (each) 50195594529282 (case), Lot Number: 6788668 ; 312) SBA69FFKWN, UDI-DI:10195594529284 (each) 50195594529282 (case), Lot Number: 6133662 ; 313) SBA69FFKWP, UDI-DI:10197106563731 (each) 50197106563739 (case), Lot Number: 7332920 ; 314) SBA69MFSU1, UDI-DI:10195594057909 (each) 50195594057907 (case), Lot Number: J9602W ; 315) SBA69MFSU2, UDI-DI:10195594353841 (each) 50195594353849 (case), Lot Number: 5140385 ; 316) SBA69MFSU3, UDI-DI:10195594448370 (each) 50195594448378 (case), Lot Number: 5533341 ; 317) SBA69MFSU4, UDI-DI:10197106058350 (each) 50197106058358 (case), Lot Number: 6560947 ; 318) SBA69MFSU5, UDI-DI:10197106104736 (each) 50197106104734 (case), Lot Number: 6560948 ; 319) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6788668 ; 320) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6788669 ; 321) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6779959 ; 322) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7005491 ; 323) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7163105 ; 324) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7163105 ; 325) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7266056 ; 326) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7269982 ; 327) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7280113 ; 328) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 7280115 ; 329) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: J9602U ; 330) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5135998 ; 331) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5140357 ; 332) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5664108 ; 333) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5664107 ; 334) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5664108 ; 335) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5841390 ; 336) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5533340 ; 337) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 5970336 ; 338) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6022136 ; 339) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6025793 ; 340) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 202109016-KSH ; 341) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6264572 ; 342) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6264573 ; 343) SBA69MFSUB, UDI-DI:10195594949167 (each) 50195594949165 (case), Lot Number: 6702034 ; 344) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7280116 ; 345) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7320243 ; 346) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7331347 ; 347) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7332920 ; 348) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7332924 ; 349) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7349697 ; 350) SBA69MFSUC, UDI-DI:10197106513606 (each) 50197106513604 (case), Lot Number: 7365541 ; 351) SBA69PBK10, UDI-DI:10195594584252 (each) 50195594584250 (case), Lot Number: 5664033 ; 352) SBA69PBKKD, UDI-DI:10888439676594 (each) 50888439676592 (case), Lot Number: J5678M ; 353) SBA69PBKKD, UDI-DI:10888439676594 (each) 50888439676592 (case), Lot Number: J6365X ; 354) SBA69PBKKD, UDI-DI:10888439676594 (each) 50888439676592 (case), Lot Number: J79F30 ; 355) SBA7CFPYN2, UDI-DI:10195594017668 (each) 50195594017666 (case), Lot Number: J9602W ; 356) SBA7CFPYN3, UDI-DI:10195594169565 (each) 50195594169563 (case), Lot Number: J9850M ; 357) SBA7CFPYN4, UDI-DI:10195594340896 (each) 50195594340894 (case), Lot Number: 5140385 ; 358) SBA7CFPYN5, UDI-DI:10197106173770 (each) 50197106173778 (case), Lot Number: 6702034 ; 359) SBA7CFPYN6, UDI-DI:10197106531037 (each) 50197106531035 (case), Lot Number: 7280116 ; 360) SBA7CFPYN6, UDI-DI:10197106531037 (each) 50197106531035 (case), Lot Number: 7320240 ; 361) SBA7CFPYN7, UDI-DI:10197106563182 (each) 50197106563180 (case), Lot Number: 7320242 ; 362) SBA7CFPYN8, UDI-DI:10197106799932 (each) 50197106799930 (case), Lot Number: 7331349 ; 363) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J46C07 ; 364) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J4952N ; 365) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J53W78 ; 366) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J53W80 ; 367) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J7963Y ; 368) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J7963X ; 369) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J84K06 ; 370) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J6368X ; 371) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J6368X ; 372) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: 5533338 ; 373) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: 5631076 ; 374) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: 5664105 ; 375) SBA7CFPYNT, UDI-DI:10888439532609 (each) 50888439532607 (case), Lot Number: J6050C ; 376) SBA7CFPYNU, UDI-DI:10195594678852 (each) 50195594678850 (case), Lot Number: 5533340 ; 377) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 6788668 ; 378) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 6788671 ; 379) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 7005491 ; 380) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 7163083 ; 381) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 7163149 ; 382) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 7163035 ; 383) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 6133662 ; 384) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 6345526 ; 385) SBA7CFPYNV, UDI-DI:10195594734244 (each) 50195594734242 (case), Lot Number: 6560947 ; 386) SBA7CFPYNW, UDI-DI:10197106607817 (each) 50197106607815 (case), Lot Number: 7349699 ; 387) SBA7CFPYNW, UDI-DI:10197106607817 (each) 50197106607815 (case), Lot Number: 7280112 ; 388) SBA7CFPYNW, UDI-DI:10197106607817 (each) 50197106607815 (case), Lot Number: 7365541 ; 389) SBA7CFPYNW, UDI-DI:10197106607817 (each) 50197106607815 (case), Lot Number: 7332923 ; 390) SBA7CFPYNX, UDI-DI:10197106764459 (each) 50197106764457 (case), Lot Number: 7163105 ; 391) SBA95BMHH1, UDI-DI:10195594018535 (each) 50195594018533 (case), Lot Number: J96N38 ; 392) SBA95BMHH1, UDI-DI:10195594018535 (each) 50195594018533 (case), Lot Number: J9601R ; 393) SBA95BMHH2, UDI-DI:10195594300517 (each) 50195594300515 (case), Lot Number: 5324048 ; 394) SBA95BMHH2, UDI-DI:10195594300517 (each) 50195594300515 (case), Lot Number: 5138269 ; 395) SBA95BMHH3, UDI-DI:10195594361716 (each) 50195594361714 (case), Lot Number: 5138226 ; 396) SBA95BMHH3, UDI-DI:10195594361716 (each) 50195594361714 (case), Lot Number: 5508047 ; 397) SBA95BMHH4, UDI-DI:10197106178065 (each) 50197106178063 (case), Lot Number: 6701891 ; 398) SBA95BMHH4, UDI-DI:10197106178065 (each) 50197106178063 (case), Lot Number: 6560936 ; 399) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J3986G ; 400) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J43U62 ; 401) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J30T14 ; 402) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J4679F ; 403) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J30T15 ; 404) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J412H ; 405) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J5678Y ; 406) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J4949D ; 407) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J72P49 ; 408) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J7497Y ; 409) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J8392X ; 410) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: N/A ; 411) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J8123A ; 412) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J6786C ; 413) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J8670R ; 414) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J6910W ; 415) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: J6278C ; 416) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 5631025 ; 417) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 5135982 ; 418) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 5664025 ; 419) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 5618929 ; 420) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6345496 ; 421) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6169293 ; 422) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6485583 ; 423) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6264453 ; 424) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6687082 ; 425) SBA95BMHHD, UDI-DI:10888439203721 (each) 50888439203729 (case), Lot Number: 6560934 ; 426) SBA95BMHHE, UDI-DI:10197106164433 (each) 50197106164431 (case), Lot Number: 6907768 ; 427) SBA95BMHHE, UDI-DI:10197106164433 (each) 50197106164431 (case), Lot Number: 6701982 ; 428) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7280045 ; 429) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7332850 ; 430) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7121998 ; 431) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 6701983 ; 432) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7055835 ; 433) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 6987010 ; 434) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7160429 ; 435) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7005451 ; 436) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7260060 ; 437) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7068424 ; 438) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7280044 ; 439) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7280083 ; 440) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7287460 ; 441) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7287492 ; 442) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7287483 ; 443) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7287493 ; 444) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7280050 ; 445) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7331296 ; 446) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7332810 ; 447) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7331297 ; 448) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7332809 ; 449) SBA95BMHHF, UDI-DI:10197106333402 (each) 50197106333400 (case), Lot Number: 7331297 ; 450) SBACGACAU2, UDI-DI:10195594101602 (each) 50195594101600 (case), Lot Number: J84J90 ; 451) SBACGACAU2, UDI-DI:10195594101602 (each) 50195594101600 (case), Lot Number: J6278P ; 452) SBACGACAU3, UDI-DI:10195594171285 (each) 50195594171283 (case), Lot Number: 5138240 ; 453) SBACGACAU3, UDI-DI:10195594171285 (each) 50195594171283 (case), Lot Number: J9699R ; 454) SBACGACAU4, UDI-DI:10195594213206 (each) 50195594213204 (case), Lot Number: 5138246 ; 455) SBACGACAU4, UDI-DI:10195594213206 (each) 50195594213204 (case), Lot Number: J9699R ; 456) SBACGACAU5, UDI-DI:10195594375102 (each) 50195594375100 (case), Lot Number: 5138222 ; 457) SBACGACAU5, UDI-DI:10195594375102 (each) 50195594375100 (case), Lot Number: 5405624 ; 458) SBACGACAU6, UDI-DI:10195594826987 (each) 50195594826985 (case), Lot Number: 6411399 ; 459) SBACGACAU6, UDI-DI:10195594826987 (each) 50195594826985 (case), Lot Number: 6345515 ; 460) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J4678J ; 461) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J25G75 ; 462) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J4678W ; 463) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J46A94 ; 464) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J6049R ; 465) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J5678G ; 466) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J6049N ; 467) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J6278K ; 468) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J72P49 ; 469) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J74A17 ; 470) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J7498U ; 471) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J8123T ; 472) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J7498Z ; 473) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J86T75 ; 474) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J8671D ; 475) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J96N36 ; 476) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J9601Y ; 477) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5530904 ; 478) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5508057 ; 479) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5631025 ; 480) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5631064 ; 481) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: J53W15 ; 482) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5664091 ; 483) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5775786 ; 484) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 5775805 ; 485) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6345493 ; 486) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6025781 ; 487) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6563916 ; 488) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6416699 ; 489) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6701884 ; 490) SBACGACAUI, UDI-DI:10888439378528 (each) 50888439378526 (case), Lot Number: 6560941 ; 491) SBACGDEM10, UDI-DI:10197106183175 (each) 50197106183173 (case), Lot Number: 6701937 ; 492) SBACGDEM11, UDI-DI:10197106266779 (each) 50197106266777 (case), Lot Number: 6907763 ; 493) SBACGDEM12, UDI-DI:10197106420492 (each) 50197106420490 (case), Lot Number: 7122068 ; 494) SBACGDEM13, UDI-DI:10197106546888 (each) 50197106546886 (case), Lot Number: 7280037 ; 495) SBACGDEME1, UDI-DI:10888439601916 (each) 50888439601914 (case), Lot Number: J4678P ; 496) SBACGDEME2, UDI-DI:10195594475017 (each) 50195594475015 (case), Lot Number: 5631035 ; 497) SBACGDEME3, UDI-DI:10195594559151 (each) 50195594559159 (case), Lot Number: J6277X ; 498) SBACGDEME4, UDI-DI:10195594595371 (each) 50195594595379 (case), Lot Number: 5775786 ; 499) SBACGDEME5, UDI-DI:10195594770433 (each) 50195594770431 (case), Lot Number: 6025767 ; 500) SBACGDEME6, UDI-DI:10195594786977 (each) 50195594786975 (case), Lot Number: 6345503 ; 501) SBACGDEME7, UDI-DI:10195594810603 (each) 50195594810601 (case), Lot Number: 6411399 ; 502) SBACGDEME8, UDI-DI:10197106002568 (each) 50197106002566 (case), Lot Number: 6485589 ; 503) SBACGDEME9, UDI-DI:10197106048306 (each) 50197106048304 (case), Lot Number: 6563907 ; 504) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J4679U ; 505) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J4950Y ; 506) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J6049M ; 507) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J6049Y ; 508) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J7904P ; 509) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J86T75 ; 510) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J6366F ; 511) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J6366D ; 512) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J6910T ; 513) SBACGDEMEA, UDI-DI:10888439270082 (each) 50888439270080 (case), Lot Number: J9602N ; 514) SBACGDEMEB, UDI-DI:10195594063061 (each) 50195594063069 (case), Lot Number: 5136034 ; 515) SBACGDEMEB, UDI-DI:10195594063061 (each) 50195594063069 (case), Lot Number: 5631025 ; 516) SBACGDEMEF, UDI-DI:10197106489420 (each) 50197106489428 (case), Lot Number: 7286184 ; 517) SBADHBPR10, UDI-DI:10195594550363 (each) 50195594550361 (case), Lot Number: 5135995 ; 518) SBADHBPR11, UDI-DI:10195594591083 (each) 50195594591081 (case), Lot Number: 5664093 ; 519) SBADHBPR12, UDI-DI:10195594646738 (each) 50195594646736 (case), Lot Number: 5140305 ; 520) SBADHBPR13, UDI-DI:10195594715335 (each) 50195594715333 (case), Lot Number: 6025777 ; 521) SBADHBPR14, UDI-DI:10195594779429 (each) 50195594779427 (case), Lot Number: 5918338 ; 522) SBADHBPR15, UDI-DI:10197106102824 (each) 50197106102822 (case), Lot Number: 6279857 ; 523) SBADHBPR16, UDI-DI:10197106147092 (each) 50197106147090 (case), Lot Number: 6487895 ; 524) SBADHBPR17, UDI-DI:10197106267516 (each) 50197106267514 (case), Lot Number: 6702005 ; 525) SBADHBPR18, UDI-DI:10197106374535 (each) 50197106374533 (case), Lot Number: 6909986 ; 526) SBADHBPRC2, UDI-DI:10888439591583 (each) 50888439591581 (case), Lot Number: J4219Y ; 527) SBADHBPRC2, UDI-DI:10888439591583 (each) 50888439591581 (case), Lot Number: J46A89 ; 528) SBADHBPRC3, UDI-DI:10888439629163 (each) 50888439629161 (case), Lot Number: J46A94 ; 529) SBADHBPRC4, UDI-DI:10888439665987 (each) 50888439665985 (case), Lot Number: J53W55 ; 530) SBADHBPRC5, UDI-DI:10888439834833 (each) 50888439834831 (case), Lot Number: J6278M ; 531) SBADHBPRC7, UDI-DI:10195594060954 (each) 50195594060952 (case), Lot Number: 5136014 ; 532) SBADHBPRC8, UDI-DI:10195594172978 (each) 50195594172976 (case), Lot Number: J9699R ; 533) SBADHBPRC9, UDI-DI:10195594288266 (each) 50195594288264 (case), Lot Number: 5140306 ; 534) SBADHBPRCC, UDI-DI:10888439603132 (each) 50888439603130 (case), Lot Number: J5678G ; 535) SBADHBPRCC, UDI-DI:10888439603132 (each) 50888439603130 (case), Lot Number: J79H51 ; 536) SBADHBPRCC, UDI-DI:10888439603132 (each) 50888439603130 (case), Lot Number: J85W18 ; 537) SBADHBPRCC, UDI-DI:10888439603132 (each) 50888439603130 (case), Lot Number: J7498Z ; 538) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332892 ; 539) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 6717951 ; 540) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 6914936 ; 541) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 6909987 ; 542) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7110269 ; 543) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7280098 ; 544) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7287561 ; 545) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332882 ; 546) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7331326 ; 547) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332883 ; 548) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332883 ; 549) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332887 ; 550) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 7332890 ; 551) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 5533322 ; 552) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 5631064 ; 553) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 6416697 ; 554) SBADHBPRCD, UDI-DI:10195594118754 (each) 50195594118752 (case), Lot Number: 6487891 ; 555) SBAOCFFK10, UDI-DI:10197106598863 (each) 50197106598861 (case), Lot Number: 7287492 ; 556) SBAOCFFK11, UDI-DI:10197106623213 (each) 50197106623211 (case), Lot Number: 7287492 ; 557) SBAOCFFK12, UDI-DI:10197106635469 (each) 50197106635467 (case), Lot Number: 7068414 ; 558) SBAOCFFK13, UDI-DI:10197106675632 (each) 50197106675630 (case), Lot Number: 6485594 ; 559) SBAOCFFK14, UDI-DI:10197106723265 (each) 50197106723263 (case), Lot Number: 7331297 ; 560) SBAOCFFK15, UDI-DI:10197106765067 (each) 50197106765065 (case), Lot Number: 7332848 ; 561) SBAOCFFKE1, UDI-DI:10195594145552 (each) 50195594145550 (case), Lot Number: J9699M ; 562) SBAOCFFKE2, UDI-DI:10195594522773 (each) 50195594522771 (case), Lot Number: 5533304 ; 563) SBAOCFFKE3, UDI-DI:10197106065259 (each) 50197106065257 (case), Lot Number: 6345505 ; 564) SBAOCFFKE4, UDI-DI:10197106152027 (each) 50197106152025 (case), Lot Number: 6560935 ; 565) SBAOCFFKE5, UDI-DI:10197106310977 (each) 50197106310975 (case), Lot Number: 6485593 ; 566) SBAOCFFKE6, UDI-DI:10197106366646 (each) 50197106366644 (case), Lot Number: 6701983 ; 567) SBAOCFFKE7, UDI-DI:10197106422823 (each) 50197106422821 (case), Lot Number: 6928774 ; 568) SBAOCFFKE8, UDI-DI:10197106472996 (each) 50197106472994 (case), Lot Number: 7280081 ; 569) SBAOCFFKE9, UDI-DI:10197106542927 (each) 50197106542925 (case), Lot Number: 7280083 ; 570) SBAOCFFKEA, UDI-DI:10888439908299 (each) 50888439908297 (case), Lot Number: J84J94 ; 571) SBAOCFFKEA, UDI-DI:10888439908299 (each) 50888439908297 (case), Lot Number: J8670T ; 572) SBAOCFFKEA, UDI-DI:10888439908299 (each) 50888439908297 (case), Lot Number: J6365C ; 573) SBAOCFFKEB, UDI-DI:10195594969882 (each) 50195594969880 (case), Lot Number: J6365G ; 574) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5135992 ; 575) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5135982 ; 576) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5533305 ; 577) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5618928 ; 578) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5618928 ; 579) SBAOCFFKEC, UDI-DI:10195594130510 (each) 50195594130518 (case), Lot Number: 5618929 ; 580) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 6568663 ; 581) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 5664036 ; 582) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 5858462 ; 583) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 6022063 ; 584) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 6169294 ; 585) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 6345506 ; 586) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 2042026 ; 587) SBAOCFFKED, UDI-DI:10195594611668 (each) 50195594611666 (case), Lot Number: 6560932 ; 588) SBAOCFFKEE, UDI-DI:10197106233856 (each) 50197106233854 (case), Lot Number: 7280082 ; 589) SEN30HNCH3, UDI-DI:10195594165680 (each) 50195594165688 (case), Lot Number: J96N43 ; 590) SEN30HNCH4, UDI-DI:10195594198343 (each) 50195594198341 (case), Lot Number: 5138330 ; 591) SEN30HNCH5, UDI-DI:10195594319656 (each) 50195594319654 (case), Lot Number: 5405620 ; 592) SEN30HNCH6, UDI-DI:10195594610876 (each) 50195594610874 (case), Lot Number: 5891299 ; 593) SEN30HNCH7, UDI-DI:10195594811853 (each) 50195594811851 (case), Lot Number: 6133659 ; 594) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J37T97 ; 595) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J30T08 ; 596) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J3986R ; 597) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J2649U ; 598) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: H1515655 ; 599) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J43U72 ; 600) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J332169 ; 601) SEN30HNCHH, UDI-DI:10888439509045 (each) 50888439509043 (case), Lot Number: J43U80 ; 602) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J6278H ; 603) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J43U79 ; 604) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J65K11 ; 605) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J5679R ; 606) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J74A02 ; 607) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J5679T ; 608) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J7904X ; 609) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J7704J ; 610) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J8045R ; 611) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J76D61 ; 612) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J70X09 ; 613) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J6278X ; 614) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J9424P ; 615) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: J6368U ; 616) SEN30HNCHI, UDI-DI:10888439785562 (each) 50888439785560 (case), Lot Number: 2680B61 ; 617) SEN30HNCHJ, UDI-DI:10195594150792 (each) 50195594150790 (case), Lot Number: 5533311 ; 618) SEN30HNCHJ, UDI-DI:10195594150792 (each) 50195594150790 (case), Lot Number: 5664049 ; 619) SEN30HNCHJ, UDI-DI:10195594150792 (each) 50195594150790 (case), Lot Number: 5775804 ; 620) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6563977 ; 621) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 20834C2285 ; 622) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6748355 ; 623) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6416685 ; 624) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 7005466 ; 625) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 7287535 ; 626) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 7331319 ; 627) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 7287534 ; 628) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 7331313 ; 629) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6022070 ; 630) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6416680 ; 631) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6416682 ; 632) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6470296 ; 633) SEN30HNCHK, UDI-DI:10195594705688 (each) 50195594705686 (case), Lot Number: 6560939 ; 634) SEN43LRUNA, UDI-DI:10197106199947 (each) 50197106199945 (case), Lot Number: 6701848 ; 635) SEN51MFK12, UDI-DI:10195594774912 (each) 50195594774910 (case), Lot Number: 6133662 ; 636) SEN51MFK13, UDI-DI:10195594801564 (each) 50195594801562 (case), Lot Number: 6264574 ; 637) SEN51MFK14, UDI-DI:10195594872229 (each) 50195594872227 (case), Lot Number: 6487906 ; 638) SEN51MFK17, UDI-DI:10197106031520 (each) 50197106031528 (case), Lot Number: 6487915 ; 639) SEN51MFK18, UDI-DI:10197106065877 (each) 50197106065875 (case), Lot Number: 6487917 ; 640) SEN51MFKO7, UDI-DI:10888439577396 (each) 50888439577394 (case), Lot Number: J30T07 ; 641) SEN51MFKO8, UDI-DI:10888439607246 (each) 50888439607244 (case), Lot Number: J37U13 ; 642) SEN51MFKO9, UDI-DI:10195594749781 (each) 50195594749789 (case), Lot Number: 5970337 ; 643) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J46C07 ; 644) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J46C08 ; 645) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 2879AC1 ; 646) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J4952P ; 647) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J43U80 ; 648) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J5679N ; 649) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J7963Y ; 650) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J85W30 ; 651) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J85W32 ; 652) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J6278X ; 653) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J9602W ; 654) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J9850D ; 655) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: J9850H ; 656) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5140378 ; 657) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5140379 ; 658) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5533341 ; 659) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5631078 ; 660) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5664107 ; 661) SEN51MFKOB, UDI-DI:10887488705194 (each) 50887488705192 (case), Lot Number: 5858464 ; 662) SEN51MFKOC, UDI-DI:10197106072738 (each) 50197106072736 (case), Lot Number: 6701854 ; 663) SEN51MFKOC, UDI-DI:10197106072738 (each) 50197106072736 (case), Lot Number: 6560948 ; 664) SEN51MFKOC, UDI-DI:10197106072738 (each) 50197106072736 (case), Lot Number: 6702023 ; 665) SEN51MFKOC, UDI-DI:10197106072738 (each) 50197106072736 (case), Lot Number: 6702024 ; 666) SEN73SEKDA, UDI-DI:10197106137710 (each) 50197106137718 (case), Lot Number: 6702024 ; 667) SEN73SEKDA, UDI-DI:10197106137710 (each) 50197106137718 (case), Lot Number: 6788668 ; 668) SEN73SEKDA, UDI-DI:10197106137710 (each) 50197106137718 (case), Lot Number: 6788670 ; 669) SENHDAPJH1, UDI-DI:10197106571910 (each) 50197106571918 (case), Lot Number: 7287537 ; 670) SENHDAPJHB, UDI-DI:10197106190234 (each) 50197106190232 (case), Lot Number: 7332878 ; 671) SENHDAPJHB, UDI-DI:10197106190234 (each) 50197106190232 (case), Lot Number: 6775066 ; 672) SENHDAPJHB, UDI-DI:10197106190234 (each) 50197106190232 (case), Lot Number: 7269986 ; 673) SENHDAPJHB, UDI-DI:10197106190234 (each) 50197106190232 (case), Lot Number: 7331322 ; 674) SNE13HTMG4, UDI-DI:10195594288563 (each) 50195594288561 (case), Lot Number: 5311245 ; 675) SNE13HTMG4, UDI-DI:10195594288563 (each) 50195594288561 (case), Lot Number: 5324047 ; 676) SNE13HTMG5, UDI-DI:10195594381226 (each) 50195594381224 (case), Lot Number: 5405611 ; 677) SNE13HTMG6, UDI-DI:10195594580988 (each) 50195594580986 (case), Lot Number: 5664027 ; 678) SNE13HTMGC, UDI-DI:10195594436681 (each) 50195594436689 (case), Lot Number: 5631025 ; 679) SNE23WGME1, UDI-DI:10197106308790 (each) 50197106308798 (case), Lot Number: PO394501 ; 680) SNE23WGME1, UDI-DI:10197106308790 (each) 50197106308798 (case), Lot Number: 6775066 ; 681) SNE23WGME2, UDI-DI:10197106399378 (each) 50197106399376 (case), Lot Number: 6701940 ; 682) SNE23WGME2, UDI-DI:10197106399378 (each) 50197106399376 (case), Lot Number: 6470296 ; 683) SNE23WGME3, UDI-DI:10197106460696 (each) 50197106460694 (case), Lot Number: 7284052 ; 684) SNE23WGME3, UDI-DI:10197106460696 (each) 50197106460694 (case), Lot Number: 7265982 ; 685) SNE23WGME4, UDI-DI:10197106519561 (each) 50197106519569 (case), Lot Number: 7260066 ; 686) SNE23WGME4, UDI-DI:10197106519561 (each) 50197106519569 (case), Lot Number: 7163081 ; 687) SNE2ASPBA1, UDI-DI:10195594285111 (each) 50195594285119 (case), Lot Number: 5324046 ; 688) SNE2ASPBA2, UDI-DI:10197106211786 (each) 50197106211784 (case), Lot Number: 6701939 ; 689) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J4218Z ; 690) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J53W12 ; 691) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J6049N ; 692) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J6365T ; 693) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J84J86 ; 694) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J84J87 ; 695) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: J96N39 ; 696) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: 5530908 ; 697) SNE2ASPBAA, UDI-DI:10888439460223 (each) 50888439460221 (case), Lot Number: 5775788 ; 698) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6701894 ; 699) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6907761 ; 700) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 7121998 ; 701) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 5631029 ; 702) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 5664033 ; 703) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6022054 ; 704) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6485591 ; 705) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6563908 ; 706) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6701957 ; 707) SNE30SHACN, UDI-DI:10195594445126 (each) 50195594445124 (case), Lot Number: 6022059 ; 708) SNE31SPKKJ, UDI-DI:10887488869384 (each) 50887488869382 (case), Lot Number: J4219W ; 709) SNE31SPKKJ, UDI-DI:10887488869384 (each) 50887488869382 (case), Lot Number: J4877C ; 710) SNE42NCMM1, UDI-DI:10197106666364 (each) 50197106666362 (case), Lot Number: 6563793 ; 711) SNE42NCMM2, UDI-DI:10197106666524 (each) 50197106666522 (case), Lot Number: 7280050 ; 712) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7005445 ; 713) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7055822 ; 714) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7122055 ; 715) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7284052 ; 716) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7280037 ; 717) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7122012 ; 718) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7280059 ; 719) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7287460 ; 720) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7331457 ; 721) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7331285 ; 722) SNE42NCMMA, UDI-DI:10197106314876 (each) 50197106314874 (case), Lot Number: 7332841 ; 723) SNE45SPTMA, UDI-DI:10197106087688 (each) 50197106087686 (case), Lot Number: 6907770 ; 724) SNE45SPTMA, UDI-DI:10197106087688 (each) 50197106087686 (case), Lot Number: 6748352 ; 725) SNE45SPTMA, UDI-DI:10197106087688 (each) 50197106087686 (case), Lot Number: 6701966 ; 726) SNE45SPTMA, UDI-DI:10197106087688 (each) 50197106087686 (case), Lot Number: 6470291 ; 727) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7005436 ; 728) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 6806109 ; 729) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7160479 ; 730) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7280091 ; 731) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7287456 ; 732) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7160550 ; 733) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7287474 ; 734) SNE45SPTMB, UDI-DI:10197106294994 (each) 50197106294992 (case), Lot Number: 7331317 ; 735) SNE53NPSO1, UDI-DI:10197106409893 (each) 50197106409891 (case), Lot Number: 7055826 ; 736) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 6264449 ; 737) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7260068 ; 738) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7260070 ; 739) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7280077 ; 740) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7280048 ; 741) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7287481 ; 742) SNE53NPSOA, UDI-DI:10197106153567 (each) 50197106153565 (case), Lot Number: 7331284 ; 743) SNE53NPSOB, UDI-DI:10197106698242 (each) 50197106698240 (case), Lot Number: 7332839 ; 744) SNE53NPSOB, UDI-DI:10197106698242 (each) 50197106698240 (case), Lot Number: 7332833 ; 745) SNE53NPSOB, UDI-DI:10197106698242 (each) 50197106698240 (case), Lot Number: 7332843 ; 746) SNE53NPSOC, UDI-DI:10197106769010 (each) 50197106769018 (case), Lot Number: 7280056 ; 747) SNE60LGRM1, UDI-DI:10197106424650 (each) 50197106424658 (case), Lot Number: 7005492 ; 748) SNE60LGRM1, UDI-DI:10197106424650 (each) 50197106424658 (case), Lot Number: 7005495 ; 749) SNE60LGRM2, UDI-DI:10197106590898 (each) 50197106590896 (case), Lot Number: 7331477 ; 750) SNE60LGRM2, UDI-DI:10197106590898 (each) 50197106590896 (case), Lot Number: 7163125 ; 751) SNE60LGRMA, UDI-DI:10195594738099 (each) 50195594738097 (case), Lot Number: 5664109 ; 752) SNE60LGRMA, UDI-DI:10195594738099 (each) 50195594738097 (case), Lot Number: 6251073 ; 753) SNE60LGRMA, UDI-DI:10195594738099 (each) 50195594738097 (case), Lot Number: 6416696 ; 754) SNE60LGRMA, UDI-DI:10195594738099 (each) 50195594738097 (case), Lot Number: 6560953 ; 755) SNE60LGRMD, UDI-DI:10197106349618 (each) 50197106349616 (case), Lot Number: 6560950 ; 756) SNE60LGRMD, UDI-DI:10197106349618 (each) 50197106349616 (case), Lot Number: 7160540 ; 757) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7333439 ; 758) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7349703 ; 759) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7333441 ; 760) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7333447 ; 761) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7333439 ; 762) SNE60LGRME, UDI-DI:10197106653746 (each) 50197106653744 (case), Lot Number: 7333448 ; 763) SNE69LPKV1, UDI-DI:10195594503208 (each) 50195594503206 (case), Lot Number: 5631025 ; 764) SNE69LPKV1, UDI-DI:10195594503208 (each) 50195594503206 (case), Lot Number: 5533324 ; 765) SNE69LPKV2, UDI-DI:10197106232453 (each) 50197106232451 (case), Lot Number: 6907749 ; 766) SNE69LPKV2, UDI-DI:10197106232453 (each) 50197106232451 (case), Lot Number: 6717950 ; 767) SNE69LPKV3, UDI-DI:10197106296141 (each) 50197106296149 (case), Lot Number: 22174311 ; 768) SNE69LPKV3, UDI-DI:10197106296141 (each) 50197106296149 (case), Lot Number: 7005431 ; 769) SNE69LPKV3, UDI-DI:10197106296141 (each) 50197106296149 (case), Lot Number: 2260596 ; 770) SNE69LPKV3, UDI-DI:10197106296141 (each) 50197106296149 (case), Lot Number: 6717950 ; 771) SNE69LPKVA, UDI-DI:10195594466220 (each) 50195594466228 (case), Lot Number: J53W15 ; 772) SNE69LPKVA, UDI-DI:10195594466220 (each) 50195594466228 (case), Lot Number: 5649225 ; 773) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6264447 ; 774) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6487895 ; 775) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7121998 ; 776) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6756777 ; 777) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7045817 ; 778) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6416698 ; 779) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7055833 ; 780) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6909987 ; 781) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7163139 ; 782) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7110269 ; 783) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7280039 ; 784) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7287561 ; 785) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7260072 ; 786) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7287561 ; 787) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7287471 ; 788) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7287560 ; 789) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7331285 ; 790) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7331476 ; 791) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6560923 ; 792) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7332882 ; 793) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7332809 ; 794) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7332883 ; 795) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7332844 ; 796) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7349682 ; 797) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7045816 ; 798) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 7332892 ; 799) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 5891284 ; 800) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 5140301 ; 801) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6025761 ; 802) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 5970332 ; 803) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6025760 ; 804) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 5888370 ; 805) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6411401 ; 806) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6025781 ; 807) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6560926 ; 808) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6279860 ; 809) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6563915 ; 810) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6487892 ; 811) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6560926 ; 812) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6487890 ; 813) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6701957 ; 814) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6416699 ; 815) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6701963 ; 816) SNE69LPKVB, UDI-DI:10195594578916 (each) 50195594578914 (case), Lot Number: 6560940 ; 817) SNE73SFKA1, UDI-DI:10197106242001 (each) 50197106242009 (case), Lot Number: 6568677 ; 818) SNE73SFKA2, UDI-DI:10197106400845 (each) 50197106400843 (case), Lot Number: 6416698 ; 819) SNE73SFKA3, UDI-DI:10197106624715 (each) 50197106624713 (case), Lot Number: 7287559 ; 820) SNE73SFKA4, UDI-DI:10197106659830 (each) 50197106659838 (case), Lot Number: 7331340 ; 821) SNE73SFKA5, UDI-DI:10197106688878 (each) 50197106688876 (case), Lot Number: 7332883 ; 822) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 7110269 ; 823) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 6909988 ; 824) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 7287559 ; 825) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 7280095 ; 826) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 7332888 ; 827) SNE73SFKAA, UDI-DI:10197106200476 (each) 50197106200474 (case), Lot Number: 7332890 ; 828) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6264449 ; 829) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6560942 ; 830) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6986974 ; 831) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6702004 ; 832) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7005436 ; 833) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6779831 ; 834) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7045817 ; 835) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6909987 ; 836) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7284052 ; 837) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6987101 ; 838) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7286184 ; 839) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6909988 ; 840) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7280062 ; 841) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7280098 ; 842) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7331284 ; 843) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7331476 ; 844) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7287461 ; 845) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7331326 ; 846) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7332810 ; 847) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7332883 ; 848) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7332839 ; 849) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 7332890 ; 850) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6701946 ; 851) SNECGADAUA, UDI-DI:10197106138977 (each) 50197106138975 (case), Lot Number: 6560940 ; 852) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6701939 ; 853) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6560942 ; 854) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6907769 ; 855) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6702005 ; 856) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6986975 ; 857) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6717951 ; 858) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6907762 ; 859) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6756777 ; 860) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7122055 ; 861) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6806111 ; 862) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7122098 ; 863) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 6914935 ; 864) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7160424 ; 865) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7197963 ; 866) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7260064 ; 867) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7163118 ; 868) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7280039 ; 869) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7280094 ; 870) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7260072 ; 871) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7287561 ; 872) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7287483 ; 873) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7332882 ; 874) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7331284 ; 875) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7331326 ; 876) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7332804 ; 877) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7332886 ; 878) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7332810 ; 879) SNECGLAAUJ, UDI-DI:10197106157015 (each) 50197106157013 (case), Lot Number: 7332883 ; 880) SNECGLAAUK, UDI-DI:10197106721971 (each) 50197106721979 (case), Lot Number: 7332824 ; 881) SNECGLAAUK, UDI-DI:10197106721971 (each) 50197106721979 (case), Lot Number: 6779832 ; 882) SNECGLAAUK, UDI-DI:10197106721971 (each) 50197106721979 (case), Lot Number: 7332812 ; 883) SNECGLAAUK, UDI-DI:10197106721971 (each) 50197106721979 (case), Lot Number: 7332887 ; 884) SNEY3NPWH2, UDI-DI:10197106227138 (each) 50197106227136 (case), Lot Number: 6701998 ; 885) SNEY3NPWHA, UDI-DI:10195594722241 (each) 50195594722249 (case), Lot Number: 7332876 ; 886) SNEY3NPWHA, UDI-DI:10195594722241 (each) 50195594722249 (case), Lot Number: 7269985 ; 887) SNEY3NPWHA, UDI-DI:10195594722241 (each) 50195594722249 (case), Lot Number: 7280092 ; 888) SNEY3NPWHA, UDI-DI:10195594722241 (each) 50195594722249 (case), Lot Number: 7332858 ; 889) SNEY3NPWHA, UDI-DI:10195594722241 (each) 50195594722249 (case), Lot Number: 6232009 ; 890) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J2649C ; 891) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: PO J2649C ; 892) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J30R91 ; 893) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J3986D ; 894) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J30T13 ; 895) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J37T98 ; 896) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4678Y ; 897) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J37T75 ; 898) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4219R ; 899) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4950X ; 900) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4950Y ; 901) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J43U68 ; 902) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J46A85 ; 903) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J46A86 ; 904) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J46A79 ; 905) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4877T ; 906) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J53W17 ; 907) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J30T15 ; 908) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J30T18 ; 909) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J4877P ; 910) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J53W47 ; 911) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J6365X ; 912) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J5677R ; 913) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: N/A ; 914) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J6786N ; 915) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J68T51 ; 916) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J5677T ; 917) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J6786P ; 918) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J6365X ; 919) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J5677T ; 920) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J70X10 ; 921) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J82M03 ; 922) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J8064K ; 923) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J8064R ; 924) SOP41BKHMQ, UDI-DI:10888439390728 (each) 50888439390726 (case), Lot Number: J8064U ;

Devices named in this recall

278 devices in the FDA device registry whose identifier appears in the recall's code information; the 100 most reported are shown.

Other products in the same recall event

The FDA groups related products into one recall event (event 95947).