Skip to content

Cardinal Health Presource Kits recallZ-1129-2025 · Cardinal Health 200, LLC · initiated Nov 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for endotoxin contamination in non-sterile surgical strips and patties in Presource Kits.

FDA root cause category: Material/Component Contamination

What the company is doing

An URGENT: MEDICAL DEVICE CORRECTION OVER LABEL notification letter dated 11/8/24 was sent to customers. If you locate affected product within your facility, please complete the following: 1. QUARANTINE affected kits & ALERT your clinicians of this action 2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all CODMAN Patties & Strips included with the Presource product. 3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648. Should you have any questions, require additional labels or desire special assistance relating to this recall, please feel free to contact Cardinal Health Quality Systems at 800.292.9332.

Recall details

Recall number
Z-1129-2025
Status
Open, Classified
Initiated by the company
Nov 8, 2024
Posted by the FDA
Feb 10, 2025
Recalling firm
Cardinal Health 200, LLC
Quantity
973,785 total units
Product code
OEZ
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Saudi Arabia.

Product description

Cardinal Health Presource Kits: 1) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV20110B; 2) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV20110D 3) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA103 4) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10T 5) KIT,CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10U 6) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10X 7) KIT, CAROTID ENDARTERECTOMY, Catalog Number:PVV2CA10Y 8) SLH AAA PACK, Catalog Number:SCV13AASLA 9) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAF10 10) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH4 11) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH5 12) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH6 13) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH7 14) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH8 15) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFH9 16) MICROVASCULAR FREE FLAP, Catalog Number:SCV30VAFHP 17) MICROVASCULAR FREE FLAP, Catalog Number:SCV30VAFHQ 18) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFHR 19) CALGARY MICROVASCULAR FREEFLAP, Catalog Number:SCV30VAFHS 20) NEURO EMBOLISM PACK, Catalog Number:SNE11AN102 21) NEURO EMBOLISM PACK, Catalog Number:SNE11AN10E

Code information (models, lots, serial numbers)

Show the full code information (6k characters)

1) PVV20110B, UDI-DI:10197106291412 (each) 50197106291410 (case), Lot Number: 6788690 ; 2) PVV20110B, UDI-DI:10197106291412 (each) 50197106291410 (case), Lot Number: 6806122 ; 3) PVV20110B, UDI-DI:10197106291412 (each) 50197106291410 (case), Lot Number: 7160517 ; 4) PVV20110D, UDI-DI:10197106546505 (each) 50197106546503 (case), Lot Number: 7287574 ; 5) PVV2CA103, UDI-DI:10888439714081 (each) 50888439714089 (case), Lot Number: J46C12 ; 6) PVV2CA10T, UDI-DI:10888439349467 (each) 50888439349465 (case), Lot Number: J46C11 ; 7) PVV2CA10T, UDI-DI:10888439349467 (each) 50888439349465 (case), Lot Number: J43W16 ; 8) PVV2CA10T, UDI-DI:10888439349467 (each) 50888439349465 (case), Lot Number: J6022K ; 9) PVV2CA10T, UDI-DI:10888439349467 (each) 50888439349465 (case), Lot Number: J7708R ; 10) PVV2CA10T, UDI-DI:10888439349467 (each) 50888439349465 (case), Lot Number: J79Z50 ; 11) PVV2CA10U, UDI-DI:10888439693607 (each) 50888439693605 (case), Lot Number: J8241D ; 12) PVV2CA10U, UDI-DI:10888439693607 (each) 50888439693605 (case), Lot Number: J84K11 ; 13) PVV2CA10X, UDI-DI:10888439967265 (each) 50888439967263 (case), Lot Number: J8998R ; 14) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 5140389 ; 15) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 5664115 ; 16) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 5858468 ; 17) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 5858469 ; 18) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 6116189 ; 19) PVV2CA10Y, UDI-DI:10195594175795 (each) 50195594175793 (case), Lot Number: 6251075 ; 20) SCV13AASLA, UDI-DI:10197106514481 (each) 50197106514489 (case), Lot Number: 7055826 ; 21) SCV13AASLA, UDI-DI:10197106514481 (each) 50197106514489 (case), Lot Number: 7287469 ; 22) SCV13AASLA, UDI-DI:10197106514481 (each) 50197106514489 (case), Lot Number: 7331456 ; 23) SCV13AASLA, UDI-DI:10197106514481 (each) 50197106514489 (case), Lot Number: 7332825 ; 24) SCV30VAF10, UDI-DI:10197106231142 (each) 50197106231140 (case), Lot Number: 6702001 ; 25) SCV30VAFH4, UDI-DI:10195594091989 (each) 50195594091987 (case), Lot Number: J8045P ; 26) SCV30VAFH5, UDI-DI:10195594154073 (each) 50195594154071 (case), Lot Number: 5138326 ; 27) SCV30VAFH6, UDI-DI:10195594243616 (each) 50195594243614 (case), Lot Number: J9699Y ; 28) SCV30VAFH7, UDI-DI:10195594324940 (each) 50195594324948 (case), Lot Number: 5405620 ; 29) SCV30VAFH8, UDI-DI:10195594381400 (each) 50195594381408 (case), Lot Number: 5138335 ; 30) SCV30VAFH9, UDI-DI:10195594616182 (each) 50195594616180 (case), Lot Number: D212939 ; 31) SCV30VAFH9, UDI-DI:10195594616182 (each) 50195594616180 (case), Lot Number: 5775803 ; 32) SCV30VAFHP, UDI-DI:10888439248883 (each) 50888439248881 (case), Lot Number: J3986T ; 33) SCV30VAFHP, UDI-DI:10888439248883 (each) 50888439248881 (case), Lot Number: J4219R ; 34) SCV30VAFHP, UDI-DI:10888439248883 (each) 50888439248881 (case), Lot Number: J4877C ; 35) SCV30VAFHQ, UDI-DI:10888439785494 (each) 50888439785492 (case), Lot Number: J6278H ; 36) SCV30VAFHQ, UDI-DI:10888439785494 (each) 50888439785492 (case), Lot Number: J6786K ; 37) SCV30VAFHQ, UDI-DI:10888439785494 (each) 50888439785492 (case), Lot Number: J76N07 ; 38) SCV30VAFHR, UDI-DI:10888439928877 (each) 50888439928875 (case), Lot Number: J86N63 ; 39) SCV30VAFHR, UDI-DI:10888439928877 (each) 50888439928875 (case), Lot Number: J86N65 ; 40) SCV30VAFHR, UDI-DI:10888439928877 (each) 50888439928875 (case), Lot Number: J70X09 ; 41) SCV30VAFHR, UDI-DI:10888439928877 (each) 50888439928875 (case), Lot Number: J9424P ; 42) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7332861 ; 43) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6701998 ; 44) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6748352 ; 45) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6748352 ; 46) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6470289 ; 47) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6775068 ; 48) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6775065 ; 49) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7005466 ; 50) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7280092 ; 51) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7280093 ; 52) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7005463 ; 53) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 7332855 ; 54) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6022069 ; 55) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6133658 ; 56) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6232003 ; 57) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6345513 ; 58) SCV30VAFHS, UDI-DI:10195594450007 (each) 50195594450005 (case), Lot Number: 6416679 ; 59) SNE11AN102, UDI-DI:10197106232293 (each) 50197106232291 (case), Lot Number: 6568688 ; 60) SNE11AN10E, UDI-DI:10197106067062 (each) 50197106067060 (case), Lot Number: 6560950 ;

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 95947).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page