Skip to content

CODMAN CERTAS recallZ-0978-2025 · Integra LifeSciences Corp. · initiated Dec 16, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for certain SKUs and lot numbers were distributed with the incorrect product description on the internal and external labels.

FDA root cause category: Labeling Change Control

What the company is doing

Integra LifeSciences notified consignees on about 12/16/2024 via letter. Consignees were instructed to review and understand the information provided in the letter, complete and return the provided Customer Acknowledgement Form, remove any affected units from service and quarantine them, arrange for return of affected units, and forward the notification to anyone who utilizes the affected units. Distributors were instructed to follow the same directions but also requested to notify any customers if product was further distributed, and return affected units.

Recall details

Recall number
Z-0978-2025
Status
Open, Classified
Initiated by the company
Dec 16, 2024
Posted by the FDA
Jan 17, 2025
Recalling firm
Integra LifeSciences Corp.
Quantity
170 units
Product code
JXG
FDA submissions
K223330
Distribution
US Nationwide distribution in the states of FL, NC, MA, MN, OH, DC, VA, CA, WI, NE, TX, NY, MI, OR, NJ, SC, AZ, UT, RI, IN, GA, CO and the countries of Canada, Australia, New Zealand, Cyprus, Sweden, Finland, Slovenia and Great Britain.

Product description

CODMAN CERTAS Plus Programmable Valve Inline Small Valve Only. Model Number: 828810PL. Implantable device that provides constant intraventricular pressure and drainage of CSF for the management of hydrocephalus.

Code information (models, lots, serial numbers)

UDI-DI: 10381780535881. Lot Numbers: 7416790, 7432471, 7434929, 7434930, 7438312 and 7438313.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

7 devices cleared under K223330 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates