EndoWrist recallZ-0849-2025 · Intuitive Surgical, Inc. · initiated Dec 19, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.
FDA root cause category: Process change control
What the company is doing
On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)
Recall details
- Recall number
- Z-0849-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 19, 2024
- Posted by the FDA
- Jan 10, 2025
- Recalling firm
- Intuitive Surgical, Inc.
- Quantity
- 10,603
- Product code
- NAY
- FDA submissions
- K131861
- Distribution
- U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and Puerto Rico. O.U.S: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam
Product description
8MM, LARGE NEEDLE DRIVER, IS4000 REF 470006
Code information (models, lots, serial numbers)
Model Number: 470006 Part Number: 470006-12 UDI-DI code: 00886874112151 Batch Numbers: K10230120 K10230118 K12230119 K11230118 K10230217 Part Number: 470006-13 UDI-DI code: 00886874112151 Batch Numbers: K10230623 K10240506 K10240815
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EndoWrist 470006Intuitive Surgical, Inc. · System,surgical,computer controlled instrument1 recall289 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 95938).
- Z-0847-2025 · Open, Classified8MM, MONOPOLAR CURVED SCISSORS, IS4000 REF 470179
- Z-0848-2025 · Open, Classified8MM, POTTS SCISSORS, IS4000 REF 470001
- Z-0850-2025 · Open, Classified8MM, ROUND TIP SCISSORS, IS4000 REF 470007
- Z-0851-2025 · Open, Classified8MM,BLACK DIAMOND MICRO FORCEPS,IS4000 REF 470033
- Z-0852-2025 · Open, Classified8MM,DEBAKEY FORCEPS,IS4000 REF 470036
- Z-0853-2025 · Open, Classified8MM,CADIERE FORCEPS,IS4000 REF 470049
- Z-0854-2025 · Open, Classified8MM,PROGRASP FORCEPS,IS4000 REF 470093
- Z-0855-2025 · Open, Classified8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 470171
- Z-0856-2025 · Open, Classified8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 470172
- Z-0857-2025 · Open, Classified8MM,RESANO FORCEPS,IS4000 REF 470181
- Z-0858-2025 · Open, Classified8MM,MEGA NEEDLE DRIVER,IS4000 REF 470194
- Z-0859-2025 · Open, Classified8MM,FENESTRATED BIPOLAR FORCEPS,IS4000 REF 470205
- Z-0860-2025 · Open, Classified8MM,TENACULUM FORCEPS,IS4000 REF 470207
- Z-0861-2025 · Open, Classified8MM,LARGE HEM-O-LOK CLIP APPLIER,IS4000 REF 470230
- Z-0862-2025 · Open, Classified8MM,ATRIAL RETRACTOR SHORT RIGHT,IS4000 REF 470246
- Z-0863-2025 · Open, Classified8MM,DUAL BLADE RETRACTOR,IS4000 REF 470249
- Z-0864-2025 · Open, Classified8MM,LARGE SUTURECUT NEEDLE DRIVER,IS4000 REF 470296
- Z-0865-2025 · Open, Classified8MM,MEGA SUTURECUT ND,IS4000 REF 470309
- Z-0866-2025 · Open, Classified8MM,SMALL GRASPING RETRACTOR,IS4000 REF 470318
- Z-0867-2025 · Open, Classified8MM,MEDIUM-LARGE CLIP APPLIER,IS4000 REF 470327
- Z-0868-2025 · Open, Classified8MM,CURVED BIPOLAR DISSECTOR,IS4000 REF 470344
- Z-0869-2025 · Open, Classified8MM,TIP-UP FENESTRATED GRASPER,IS4000 REF 470347
- Z-0870-2025 · Open, Classified8MM,HORIZON SMALL CLIP APPLIER,IS4000 REF 470401
- Z-0871-2025 · Open, Classified8MM,FORCE BIPOLAR,IS4000 REF 470405
- Z-0872-2025 · Open, Classified8MM,LONG TIP FORCEPS,IS4000 REF 471048
- Z-0873-2025 · Open, Classified8MM,CADIERE FORCEPS,IS4000 REF 471049
- Z-0874-2025 · Open, Classified8MM,PROGRASP FORCEPS,IS4000 REF 471093
- Z-0875-2025 · Open, Classified8MM,MICRO BIPOLAR FORCEPS,IS4000 REF 471171
- Z-0876-2025 · Open, Classified8MM,MARYLAND BIPOLAR FORCEPS,IS4000 REF 471172
- Z-0877-2025 · Open, Classified8MM,COBRA GRASPER,IS4000 REF 471190
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page