Skip to content

EndoWrist recallZ-0863-2025 · Intuitive Surgical, Inc. · initiated Dec 19, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Intuitive Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to increased complaints regarding frayed or broken grip cables on reusable instruments with jaws.

FDA root cause category: Process change control

What the company is doing

On 12/19/2024, the firm send via email an "Urgent: Medical Device Correction" Letter informing customer that Intuitive Surgical has become aware of an increase in complaints regarding frayed or broken cables on some da Vinci X and Xi reusable instruments. Customer are instructed to: 1. Customers can continue using the products in accordance with the user manual. 2. As a reminder, when using da Vinci X and Xi reusable instruments, follow the Inspection Before Use and Warnings listed in the manual provided with your system to inspect for any broken cables. Refer to Figures C and D for examples. In addition, please refer to Appendix B for additional images for detection of grip cable failures. 3. If you observe any failed (frayed or broken) grip cables prior to use, during procedure, or during reprocessing, please stop use of instrument, remove from use and inform Intuitive via the standard complaint process. 4. Inform Intuitive of any Adverse Events*/Serious Incidents** or quality problems concerning the use of the subject device via the standard complaint process. For Question and assistance - " North and South America: (800) 876-1310, Option 3 (4 am to 5 pm PST) or mail: [email protected]. " Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) or [email protected] " South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) or [email protected] " Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) or [email protected] " India: +1-800-103-6952 (9 am to 6 pm IT) " Taiwan: +0800-86-8181 (9 am to 6 pm CT)

Recall details

Recall number
Z-0863-2025
Status
Open, Classified
Initiated by the company
Dec 19, 2024
Posted by the FDA
Jan 10, 2025
Recalling firm
Intuitive Surgical, Inc.
Quantity
783
Product code
NAY
FDA submissions
K131861
Distribution
U.S. - Nation-wide. Alabama, Alaska, Arizona, Arkansas, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, South Dakota, Tennessee, Texas, Utah, Vermont, Virginia, Washington, West Virginia, Wisconsin, Wyoming and Puerto Rico. O.U.S: Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Columbia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, French Guiana, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Mexico, Monaco, Netherlands, New Zealand, Norway, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Vietnam

Product description

8MM,DUAL BLADE RETRACTOR,IS4000 REF 470249

Code information (models, lots, serial numbers)

Model Number: 470249 Part Number: 470249-09 UDI-DI code: 00886874112403 Batch Numbers: K10230226 K10230316 K10230323 K10230331 K10230518 K10230824 K10230831 K10231130 K10231214 K10240118 K10240208 K10240222 K10240327 K10240509 K10240711 K10240725 K10240801 K11231116

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 95938).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page