Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285.… recallZ-1074-2025 · Getinge Usa Sales Inc · initiated Jan 15, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

FDA root cause category: Device Design

What the company is doing

Getinge notified consignees on 01/15/2024 via URGENT MEDICAL DEVICE RECALL letter. Consignees were provided additional information which is complementary to the existing user manual and were instructed to ensure all affected product inventory has been quarantined, consignees are entitled to credit/replacement, the recall information is forwarded to affected personnel and any customers if it was further distributed and complete and return the provided URGENT MEDICAL DEVICE RECALL RESPONSE FORM.

Recall details

Recall number
Z-1074-2025
Status
Open, Classified
Initiated by the company
Jan 15, 2025
Posted by the FDA
Feb 7, 2025
Recalling firm
Getinge Usa Sales Inc
Quantity
6 units
Product code
BSZ
FDA submissions
K191027
Distribution
Domestic distribution to Kentucky and Connecticut.

Product description

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Code information (models, lots, serial numbers)

Model Number: 6682285. UDI-DI: 07325710001141. Serial Numbers: 3761, 7635, 7633, 7630, 7632, 7631.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

6 devices cleared under K191027 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.