stryker Precision Thin, Catalog Number REF 2296-003-155 recallZ-1139-2025 · Stryker Corporation · initiated Jan 8, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.
FDA root cause category: Process control
What the company is doing
Stryker issued an URGENT MEDICAL DEVICE recall notice to its consignees on 01/08/2025 via FedEx. The notice explained the issue, potential risk, and requested the affected items be quarantined pending return. Stryker will issue a shipping label to the consignee once the response form has been received. Distributors were directed to forward the notice to all those to whom the affected devices were distributed.
Recall details
- Recall number
- Z-1139-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 8, 2025
- Posted by the FDA
- Feb 11, 2025
- Recalling firm
- Stryker Corporation
- Quantity
- 238 units
- Product code
- GEI
- Distribution
- Worldwide distribution.
Product description
stryker Precision Thin, Catalog Number REF 2296-003-155; Micro Oscillating and Sagittal Blades
Code information (models, lots, serial numbers)
GTIN 04546540466549, Lot Number 23070017
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NA 2296003155Stryker Corporation · BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96098).
- Z-1138-2025 · Open, Classifiedstryker Precision Thin, Catalog Number REF 2296-003-108; Micro Oscillating and Sagittal Blades
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page