Skip to content

NA 2296003155Recalls & reported problems

Stryker Corporation · Precision Thin (9.0 x 0.38 x 18.5mm) · Class 1

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Jan 8, 2025Class IIOpen, Classified

    Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece.

    Z-1139-2025 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code GFA (blade, saw, general & plastic surgery, surgical) under 21 CFR 878.4820 as a Class I device. 21 CFR 878.4820 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the NA 2296003155

Has the NA 2296003155 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1139-2025 (Class II), began on Jan 8, 2025: Precision Thin blade attachments have a potential to be out of measurement specifications, preventing them from securely fitting into the compatible handpiece. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the NA 2296003155 FDA approved?

Devices of this type (product code GFA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Stryker Corporation.

Who makes the NA 2296003155, and is it still sold?

The FDA's device database lists Stryker Corporation as the company responsible for it (brand NA, model 2296003155). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices

All BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page