Sysmex PS-10 Sample Preparation System with software… recallZ-1113-2025 · Sysmex America, Inc. · initiated Jan 6, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sysmex America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.

FDA root cause category: Software design

What the company is doing

A Product Notification Field Correction letter was sent to customers beginning 1/6/25. Actions: 1. Until a Sysmex representative contacts you, please do not use the instrument running PS-10 software version 1.5 or 1.6 unless it is used only to create cocktails. 2. Review the content of this communication with your facility's physician and/or pathologist and retain this letter for any future reference. 3. Please fill out the form below and email it back to Sysmex using the email address: [email protected] Questions If you have any questions concerning this information, please contact the Technical Assistance Center at 1- 888-879-7639 An additional Product Notification Field Correction letter was sent to customers beginning 4/9/25. Software Correction The probe washing processes have been updated to incorporate essential washing steps during antibody pipetting. The PS-10 antibody pipetting washing sequence has been reinstated, ensuring a washing step is performed when switching between antibody vials. This update prevents any potential carry-over between antibody reagents. This correction addresses the potential risk of antibody carryover, as detailed in Field Correction Notice CF-08823, released in December 2024. Action 1. Please distribute this product notification as appropriate to your laboratory staff. 2. File this product notification as a part of your laboratory s quality system as required. Questions If you have questions concerning this information, please message the Sysmex Technical Assistance Center (TAC). Simply click on the support tab on the home page of the CRC, click the Create New Issue button under Help Me to send your question to TAC. Choose Technical Assistance-Equipment in the dropdown to complete the form and submit it. Sending a message to TAC is recommended for non-urgent requests. For Canadian customers, please note messages to TAC are only processed in English. For urgent requests, contact the Technical Assistance Center at 1-8

Recall details

Recall number
Z-1113-2025
Status
Open, Classified
Initiated by the company
Jan 6, 2025
Posted by the FDA
Feb 5, 2025
Recalling firm
Sysmex America, Inc.
Quantity
32 units
Product code
PER
Distribution
US Nationwide.

Product description

Sysmex PS-10 Sample Preparation System with software version 1.5 and 1.6., Catalog Number BQ716341. Product usage for automated pipetting, diluting, and specimen processing workstations for flow cytometric analysis.

Code information (models, lots, serial numbers)

UDI-DI: 04987562501908; Serial Numbers: 1908002275 1908002276 1908009775 1910009588 1910011295 1912002638 2103013913 2108004629 2108004630 2108004631 2108004632 2112005423 2112005425 2203014206 2203014207 2203014208 2205012732 2205012734 2206005973 2206012294 2208015835 2210003722 2210003723 2210003724 2212010048 1903011966 1903011968 1910009589 1912002640 2103013912 2208015834 1903011967

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.