Allura Xper FD recallZ-1077-2025 · Philips Medical Systems Nederland B.V. · initiated Jan 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
FDA root cause category: Device Design
What the company is doing
Philips issued URGENT Medical Device Correction Letter via Certified Mail (USPS) on 1/13/25. Letter states reason for recall, health risk and action to take: " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions when positioning the patient table included in Appendix B. These instructions include information of potential finger entrapment areas when moving the table. The information of this Appendix is being included in an Addendum to the Instructions for Use. Additionally, Philips is creating a Quick Reference Card. The Addendum and the Quick Reference Card will be available as from end of March 2025. Please refer to Appendix C on how to download the Addendum to the Instructions for Use and the Quick Reference Card. " In case that the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through the email account: [email protected] " Please complete and return the Response form attached to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction Letter and understanding of the issue and required actions to be taken. " Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative (1- 800-722-9377). Actions planned by Philips IGT-S to correct the problem Philips is releasing an addendum to the instructions for use and a Quick Reference Card that will be available for customers at the end of March 2025. Philips is working on the design of additional brackets for the AD7 tables, and a new redesigned table cover for AD7X tables (ref. FCOs: FCO72200516, FCO72200552, FCO72200553, FCO72200554, FCO72200555, FCO72200556). Philips will contact you to schedule a visit once the ne
Recall details
- Recall number
- Z-1077-2025
- Status
- Open, Classified
- Initiated by the company
- Jan 13, 2025
- Posted by the FDA
- Feb 4, 2025
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 474 units US; 1908 units OUS
- Product code
- OWB
- FDA submissions
- K162859, K041949, K031333
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, JAMAICA, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritania, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Yemen.
Product description
Allura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026
Code information (models, lots, serial numbers)
UDI: (1) 722003 00884838059191(10); (2) 722010 00884838059030(21); (3)722026 00884838054189(21).
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System23 recalls2 deaths2,015 reports
- Allura Xper FD Allura Xper FD10Philips Medical Systems Nederland B.V. · Interventional Fluoroscopic X-Ray System20 recalls2 deaths1,837 reports
- Interventional Workspot Interventional Workspot - 1Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological13 recalls24 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96120).
- Z-1078-2025 · Open, ClassifiedAlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027
- Z-1080-2025 · Open, ClassifiedAlluraXper FD20 System Code: (1) 722006 (2) 722012 (3) 722028
- Z-1081-2025 · Open, ClassifiedAlluraXper FD20 Biplane System Code: (1) 722008 (2) 722013
- Z-1084-2025 · Open, ClassifiedAlluraXperFD20/10 System Code: (1) 722029
- Z-1085-2025 · Open, ClassifiedAllura Xper FD20/20 System Code: (1) 722038
- Z-1086-2025 · Open, ClassifiedAlluraXperFD20/15 System Code: (1) 722058
- Z-1087-2025 · Open, ClassifiedAzurion 3 M12 System Code: (1) 722063 (2) 722221
- Z-1088-2025 · Open, ClassifiedAzurion 3 M15 System Code: (1) 722064 (2) 722222 (3) 722280
- Z-1089-2025 · Open, ClassifiedAzurion 5 M12 System Code: (1) 722227 (2) 722231
- Z-1090-2025 · Open, ClassifiedAzurion 5 M20 System Code: (1) 722228 (2) 722232 (3) 722281
- Z-1091-2025 · Open, ClassifiedAzurion 7 B12 System Code: (1) 722067 (2) 722225 (3) 722235
- Z-1092-2025 · Open, ClassifiedAzurion 7 B20 System Code: (1) 722068 (2) 722226 (3) 722236
- Z-1093-2025 · Open, ClassifiedAzurion 7 M12 System Code: (1) 722078 (2) 722223 (3) 722233
- Z-1094-2025 · Open, ClassifiedAzurion 7 M20 System code: (1) 722224 (2) 722079 (3) 722234 (4) 722282 System Code: (1) 722078 (2) 722223 (3) 722233
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page