Skip to content

Allura Xper FDRecalls & reported problems across all models

Philips Medical Systems Nederland B.V. · 19 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

82 open recalls name at least one Allura Xper FD model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
15,187
Deaths reported
13
Injuries reported
52
Recalls
84

We found 19 models of Allura Xper FD by Philips Medical Systems Nederland B.V. in the FDA's device registry (GUDID). Together they are named in 15,187 adverse event reports in the FDA's MAUDE database, received between Jun 15, 2017 and Sep 30, 2026 — classified as 13 deaths, 52 injuries, 15,122 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (7,167 reports).

84 FDA recall records name at least one of these models by its device identifier, 82 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 407 reportsDec 2023: 391 reportsJan 2024: 362 reportsFeb 2024: 530 reportsMar 2024: 371 reportsApr 2024: 308 reportsMay 2024: 252 reportsJun 2024: 298 reportsJul 2024: 302 reportsAug 2024: 327 reportsSep 2024: 360 reportsOct 2024: 373 reportsNov 2024: 240 reportsDec 2024: 89 reportsJan 2025: 182 reportsFeb 2025: 331 reportsMar 2025: 170 reportsApr 2025: 253 reportsMay 2025: 301 reportsJun 2025: 469 reportsJul 2025: 419 reportsAug 2025: 886 reportsSep 2025: 839 reportsOct 2025: 904 reportsNov 2025: 779 reportsDec 2025: 73 reportsJan 2026: 70 reportsFeb 2026: 40 reportsMar 2026: 42 reportsApr 2026: 17 reportsMay 2026: 49 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 1 reportsSep 2026: 22 reportsOct 2026: 0 reports202420252026904
Last 36 months. The most recent month may be incomplete.

Allura Xper FD recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 29, 2026Open, Classified

    Firmware issue potentially leading to loss of geometry movements.

    Z-2928-2026 · Root cause: Software design · names 7 of these models · Recall details · FDA record ↗

  2. Jun 17, 2026Open, Classified

    Potential Loss of imaging functionality, loss of motorized movement, or loss of data due to unauthorized Hard Disk Drives.

    Z-2891-2026 · Root cause: Under Investigation by firm · names 18 of these models · Recall details · FDA record ↗

  3. Apr 30, 2026Open, Classified

    It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

    Z-2259-2026 · Root cause: Under Investigation by firm · names 17 of these models · Recall details · FDA record ↗

  4. Apr 28, 2026Open, Classified

    The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

    Z-2228-2026 · Root cause: Component design/selection · names 14 of these models · Recall details · FDA record ↗

  5. Mar 19, 2026Open, Classified

    Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

    Z-1805-2026 · Root cause: Labeling design · names 8 of these models · Recall details · FDA record ↗

  6. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1672-2026 · Root cause: Device Design · Recall details · FDA record ↗

  7. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1671-2026 · Root cause: Device Design · Recall details · FDA record ↗

  8. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1670-2026 · Root cause: Device Design · Recall details · FDA record ↗

  9. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1669-2026 · Root cause: Device Design · Recall details · FDA record ↗

  10. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1668-2026 · Root cause: Device Design · Recall details · FDA record ↗

  11. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1667-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  12. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1666-2026 · Root cause: Device Design · Recall details · FDA record ↗

  13. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1665-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  14. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1664-2026 · Root cause: Device Design · Recall details · FDA record ↗

  15. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1663-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  16. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1662-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  17. Mar 3, 2026Open, Classified

    Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.

    Z-1661-2026 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗

  18. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1077-2026 · Root cause: Other · Recall details · FDA record ↗

  19. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1076-2026 · Root cause: Other · Recall details · FDA record ↗

  20. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1075-2026 · Root cause: Other · Recall details · FDA record ↗

  21. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1074-2026 · Root cause: Other · Recall details · FDA record ↗

  22. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1073-2026 · Root cause: Other · Recall details · FDA record ↗

  23. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1072-2026 · Root cause: Other · Recall details · FDA record ↗

  24. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1071-2026 · Root cause: Other · Recall details · FDA record ↗

  25. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1070-2026 · Root cause: Other · Recall details · FDA record ↗

  26. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1069-2026 · Root cause: Other · Recall details · FDA record ↗

  27. Dec 15, 2025Open, Classified

    Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.

    Z-1068-2026 · Root cause: Other · Recall details · FDA record ↗

  28. Sep 3, 2025Open, Classified

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

    Z-0013-2026 · Root cause: Device Design · Recall details · FDA record ↗

  29. Sep 3, 2025Open, Classified

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

    Z-0012-2026 · Root cause: Device Design · Recall details · FDA record ↗

  30. Sep 3, 2025Open, Classified

    The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

    Z-0011-2026 · Root cause: Device Design · Recall details · FDA record ↗

…and 54 older recalls, listed on each model's page.

Most reported Allura Xper FD problems

Device problems

  • Activation Failure7,167
  • Radiation Output Failure4,806
  • Positioning Failure1,615
  • No Display/Image1,150
  • Detachment of Device or Device Component96
  • Display or Visual Feedback Problem80
  • Failure to Power Up72
  • Unintended System Motion56

Patient problems

  • No Clinical Signs, Symptoms or Conditions15,030
  • Insufficient Information169
  • No Consequences Or Impact To Patient18
  • Hair Loss11
  • No Known Impact Or Consequence To Patient10
  • Fall9
  • Cardiac Arrest6
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4

Allura Xper FD models

19 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Allura Xper FD adverse event reports