PROLENE recallZ-1096-2025 · Ethicon, LLC · initiated Dec 20, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ethicon, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.
FDA root cause category: Process control
What the company is doing
On 12/20/2024, Johnson & Johnson Med Tech mailed recall notices to Operating Room Supervisors, Materials Management Personnel, Chief of Surgery, and distributors who were asked to do the following: 1) Examine your inventory immediately to determine if you have product subject to this recall on hand and quarantine such product(s). 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return. Please consider including a copy of this recall letter when communicating. 3) Contact Sedgwick at 888-548-8526 to arrange affected product return. 4) Keep this notice visibly posted for awareness until all product subject to this recall has been returned to Sedgwick. 5) Complete and return the Business Reply Form via email to [email protected] If you have additional questions regarding this communication contact the firm's Resource Department at 1-877-ETHICON (1-877-384-4266) open Monday through Friday, 8:00 AM to 5:00 PM ET.
Recall details
- Recall number
- Z-1096-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 20, 2024
- Posted by the FDA
- Feb 4, 2025
- Recalling firm
- Ethicon, LLC
- Quantity
- 15,948
- Product code
- GAW
- FDA submissions
- K133356
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE, JP, LU, NL, NO, PL, SE, SG, ZA.
Product description
PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurosurgical procedures.
Code information (models, lots, serial numbers)
REF/UDI-DI-GTIN/Lot(expiration): EH7585H/10705031097254/103DQR(8/31/2029); 8706H/10705031019430/103E22(8/31/2029), 103EMJ(8/31/2029); 8833H/10705031021037/103HC4(8/31/2029); 8963H/10705031091931/103E8Z(8/31/2029); 8935H/10705031022089/103HBP(8/31/2029)
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- PROLENE 8706HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE2 recalls294 reports
- PROLENE EH7585HEthicon, LLC · Suture, nonabsorbable, synthetic, polypropylene1 recall63 reports
- PROLENE 8833HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall49 reports
- PROLENE 8935HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall21 reports
- PROLENE 8963HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall0 reports
Product lines
Other devices under the same FDA submission
337 devices cleared under K133356 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- PROLENE 8702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths254 reports
- PROLENE 8556HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death240 reports
- PROLENE 8805HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 death209 reports
- PROLENE 8521HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE3 deaths183 reports
- PROLENE 8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE163 reports
- PROLENE 8711HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE1 recall89 reports
- PROLENE 9702HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE130 reports
- PROLENE EP8735HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE128 reports
- PROLENE EP8732HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE123 reports
- PROLENE 8732HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE119 reports
- PROLENE 8557HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE109 reports
- PROLENE 8703HEthicon, LLC · SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE102 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96127).
- Z-1097-2025 · Open, ClassifiedPERMA-HAND Silk Suture, REF: K872H, W723H, 623H, K833H; Perma-hand Silk Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovas…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page