AnchorFast SlimFit, Oral Endotracheal Tube Fastener, REF… recallZ-1179-2025 · Hollister Incorporated · initiated Dec 23, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Hollister Incorporated or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Hollister received reports of decreased skin barrier wear time which could can lead to tube migration
FDA root cause category: Process change control
What the company is doing
Hollister issued an URGENT - MEDICAL DEVICE RECALL notice to its consignees on 12/23/2024 via FedEx and email. The notice explained the reason for the recall and potential risk and instructed customers to quarantine affected inventory, discard the product per their local procedures for product destruction, and Please complete and return the response form attached as soon as possible, even if you do not have affected product in your possession and send to [email protected]. For product credit requests: If you are a direct customer of Hollister, contact Hollister Incorporated Customer Service for product credit at: 1-800-323-4060 (prompt #1). If you purchase product from a distributor, please contact your distributor for product credit. Hollister has posted this recall on their website with the link to the page.
Recall details
- Recall number
- Z-1179-2025
- Status
- Open, Classified
- Initiated by the company
- Dec 23, 2024
- Posted by the FDA
- Feb 19, 2025
- Recalling firm
- Hollister Incorporated
- Quantity
- 45 box / 540 eaches
- Product code
- CBH
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Product description
AnchorFast SlimFit, Oral Endotracheal Tube Fastener, REF 9787,
Code information (models, lots, serial numbers)
UDI/DI 00610075095834, Lot Numbers: 4L172, 4L182
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- AnchorFast SlimFit 9787Hollister Incorporated · DEVICE, FIXATION, TRACHEAL TUBE1 recall4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96181).
- Z-1178-2025 · Open, ClassifiedAnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700,
- Z-1180-2025 · Open, ClassifiedAnchorFast, Oral Endotracheal Tube Fastener, REF 9799,
- Z-1181-2025 · Open, ClassifiedAnchorFast Guard, Oral Endotracheal Tube Fastener, REF 9800,
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.