Numerous models of nonsterile hemostatic forceps recallZ-1312-2025 · Aesculap Inc · initiated Jan 31, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Aesculap Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The forceps have been used in ways not covered by the design resulting in breakage of the clamps.

FDA root cause category: Use error

What the company is doing

The recalling firm issued letters dated 1/29/2025 on 1/31/2025 via certified mail return-receipt. The letter explained the reason for the correction, identified the product impacted, provided the risk to health, and the actions required for the customer/user. The actions included: (1) Review the Urgent Medical Device Correction Notification in its entirety and ensure all users within the consignee organization are informed about the correction. Post the notification where the affected products are stored. If the consignee is a distributor, they are to forward the notification to their customers. The correction is to be extended to the end user/customer level; (2) Determine the current inventory of the affected items and comply with the following instructions: (a) Use the products identified for clamping of tissue and small vessels only; (b) Inspect the products for end-of-life indicators prior to use; (c) Visually inspect the products which should be clean and free of any signs of corrosion and cracks. Check the joint and jaw in the open position from both sides (if needed, use magnification; (d) Test the function of the device, including the jaw area, for deformation and loss of tension. An Acknowledgment Form was enclosed for completion and return via FAX or email within 2 weeks of receipt, even if the total inventory in the consignee location is 0.

Recall details

Recall number
Z-1312-2025
Status
Open, Classified
Initiated by the company
Jan 31, 2025
Posted by the FDA
Mar 11, 2025
Recalling firm
Aesculap Inc
Quantity
47,078 forceps
Product code
KAE
Distribution
Distribution was made nationwide and to PR. There was also government/military distribution. Foreign distribution was made to Canada, Bermuda, Bahamas, and Qatar.

Product description

Numerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM; (2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM; (3) REF BH957R, Schnidt Tonsil Forceps SLT-CVD185MM; (4) REF BH959R, Schnidt Tonsil Forceps CVD 185MM; (5) REF BH961R, Schnidt Tonsil FCPSOPEN-RINGCVD 185MM; (6) REF BH963R, Schnidt Tonsil FCPSOPEN-RINGCVD185MM; and (7) REF BH965R, Wilson Tonsil Forceps 190MM.

Code information (models, lots, serial numbers)

GTIN numbers: (1) REF BH951R - 04038653027026; (2) REF BH952R - 04038653139200; (3) REF BH957R - 04038653027033; (4) REF BH959R - 04038653027040; (5) REF BH961R - 04038653027057; (6) REF BH963R - 04038653027064; and (7) REF BH965R - 04038653027071.

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 96223).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page