TS 7500 recallZ-1418-2025 · Baxter Healthcare Corporation · initiated Feb 25, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
FDA root cause category: Software design
What the company is doing
Baxter issued an UREGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/25/2025 by letter sent via traceable means. The notice explained the issue, hazard involved, and requested the following actions be taken: "Actions to be Taken by Customers 1. Baxter will contact you to install a software upgrade in the impacted surgical table systems. 2. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply form instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 3. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 4. Please forward a copy of this communication to the Medical Director/Office Manager, Director of Pharmacy, Medication Safety Officer, Purchasing Director/Head of Central Supply and any other departments within your institution who use the affected product. 5. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal." "For general questions regarding this communication, please contact Baxter Technical Support at 800-445-3720 (select option 2, then option 7) between 8:00 am and 6:00 pm Eastern Time, Monday through Friday, or email Baxter at [email protected]."
Recall details
- Recall number
- Z-1418-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 25, 2025
- Posted by the FDA
- Mar 21, 2025
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 8 units
- Product code
- JEA
- Distribution
- US Nationwide distribution.
Product description
Baxter TruSystem 7500 Hybrid (SC), Product Code 1854086
Code information (models, lots, serial numbers)
UID/DI 00887761974074, All serial numbers manufactured until 11/14/2024.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- TS 7500 1854086Baxter Medical Systems GmbH + Co. KG · Table, radiologic1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96286).
- Z-1409-2025 · Open, ClassifiedBaxter Operating table column TS7500 MOBIUS, Product Code 1704695
- Z-1410-2025 · Open, ClassifiedBaxter Stationary column TruSystem 7500, Product Code 1717020
- Z-1411-2025 · Open, ClassifiedBaxter Floor mounting column TruSystem 7500, Product Code 1717021
- Z-1412-2025 · Open, ClassifiedBaxter Mobile column TruSystem 7500, Product Code 1717023
- Z-1413-2025 · Open, ClassifiedBaxter Mobile column TruSystem 7500 U, Product Code 1730720
- Z-1414-2025 · Open, ClassifiedBaxter Stationary column TruSystem 7500 U, Product Code 1730731
- Z-1415-2025 · Open, ClassifiedBaxter Floor mounting column TS 7500 U, Product Code 1730732
- Z-1416-2025 · Open, ClassifiedBaxter TruSystem 7500 Hybrid (MC), Product Code 1773204
- Z-1417-2025 · Open, ClassifiedBaxter TruSystem 7500 Hybrid (FC), Product Code 1854085
- Z-1419-2025 · Open, ClassifiedBaxter TruSystem 7500 Hybrid Plus (FC), Product Code 1854087
- Z-1420-2025 · Open, ClassifiedBaxter TruSystem 7500 Hybrid Plus (SC), Product Code 1854088
- Z-1421-2025 · Open, ClassifiedBaxter TruSystem 7500 Hybrid MR IMRIS, Product Code 2067886
- Z-1422-2025 · Open, ClassifiedBaxter TruSystem 7500, Product Code 4091000
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page