TS 7500Recalls & reported problems across all models
Baxter Medical Systems GmbH + Co. KG · 16 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
14 open recalls name at least one TS 7500 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls
- 14
We found 16 models of TS 7500 by Baxter Medical Systems GmbH + Co. KG in the FDA's device registry (GUDID). Together they are named in 4 adverse event reports in the FDA's MAUDE database, received between Jul 23, 2025 and Nov 10, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).
14 FDA recall records name at least one of these models by its device identifier, 14 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
TS 7500 recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1422-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1421-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1420-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1419-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1418-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1417-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1416-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1415-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1414-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1413-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1412-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1411-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1410-2025 · Root cause: Software design · Recall details · FDA record ↗
- Feb 25, 2025Open, Classified
There is a software issue which causes the upper back section to not be operable/adjustable when the "emergency mode" function is enabled.
Z-1409-2025 · Root cause: Software design · Recall details · FDA record ↗
Most reported TS 7500 problems
Device problems
- Device Dislodged or Dislocated1
- Device Fell1
- Key or Button Unresponsive/not Working1
- Unintended Movement1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
TS 7500 models
16 models, most reported first. Each model has its own page with its reports, recalls and device details.
- TS 7500 1730720Table, radiologic1 recall1 report
- TS 7500 1717023TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall1 report
- TS 7500 4091000TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 2067886TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 1854088Table, radiologic1 recall0 reports
- TS 7500 1854087TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1854086Table, radiologic1 recall0 reports
- TS 7500 1854085TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 1773204Table, radiologic1 recall0 reports
- TS 7500 1730732Table, radiologic1 recall0 reports
- TS 7500 1730731Table, radiologic1 recall0 reports
- TS 7500 1717021TABLE, OPERATING-ROOM, ELECTRICAL1 recall0 reports
- TS 7500 1717020Table, radiologic1 recall0 reports
- TS 7500 1704695TABLE, SURGICAL WITH ORTHOPEDIC ACCESSORIES, AC-POWERED1 recall0 reports
- TS 7500 2064888Table, radiologic1 report
- TS 7500 1909793TABLE, OPERATING-ROOM, ELECTRICAL1 report
Recent TS 7500 adverse event reports
The 4 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2025Key or Button Unresponsive/not Working; Key or Button Unresponsive/not Working
BAXTER RECEIVED A REPORT STATING THAT TRUSYSTEM 7500 U MOBILE COLUMN HAD NO FUNCTIONS. THE TABLE WAS LOCATED AT THE ACCOUNT AND OCCURRED BEFORE USE. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511614 ↗MalfunctionAug 22, 2025Device Fell; Device Fell
BAXTER RECEIVED A REPORT STATING THAT DURING THE OPERATION, TRUSYSTEM 7500 TABLE TOP COLLAPSED WITHOUT ANY INPUT FROM THE NURSING STAFF. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22874958 ↗MalfunctionJul 25, 2025Unintended Movement; Unintended Movement
BAXTER RECEIVED A REPORT STATING THAT OPERATING TABLETOP MIDDLE BACK SECTION WAS MOVING DOWN WITHOUT ANY USER INPUT. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22612578 ↗InjuryJul 23, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED, THE HEAD PLATE OF AN OR TABLETOP U26 H V COLLAPSED. DURING AN UNSPECIFIED BREAST PROCEDURE AND JUST BEFORE WAKING THE PATIENT FROM ANESTHESIA, THE HEAD PLATE OF THE TRUSYSTEM 7500 OPERATING TABLE COLLAPSED TO THE FLOOR WITHOUT WARNING. THE PATIENT¿S HEAD DROPPED RAPIDLY BACKWARDS IN HYPEREXTENSION. ONE OF THE HOSPITAL EMPLOYEES MANAGED TO GRAB THE PATIENT¿S HEAD BEFORE IT DROPPED FURTHER DOWN AND IMMEDIATELY STRAIGHTENED THE NECK TO ITS NATURAL POSITION. THE CUSTOMER STATED THAT THE OPERATION WENT WELL AND INITIALLY THE INCIDENT WAS NOT CONSIDERED TO HAVE HAD ANY IMPACT ON THE PATIENT. THE INCIDENT WAS RECORDED, AND THE PATIENT WAS INFORMED BEFORE BEING DISCHARGED. THE PATIENT DID NOT HAVE ANY SYMPTOMS OR DISCOMFORT AT THAT TIME. THREE DAYS AFTER THE INCIDENT, THE PATIENT REPORTEDLY SOUGHT MEDICAL CARE IN THEIR HOMETOWN DUE TO PAIN AND DISCOMFORT IN THE BACK OF THEIR NECK, AND WHIPLASH SYMPTOMS. THE PATIENT WAS ADVISED TO WEAR A NECK COLLAR AND TAKE UNREPORTED PAIN MEDICATIONS. IT WAS ALSO NOTED THAT THE PATIENT UNDERWENT AN MRI SCAN IN A DIFFERENT HEALTHCARE FACILITY AND THE RESULTS WERE NOT PROVIDED. AN INSPECTION PERFORMED BY THE HOSPITAL¿S TECHNICIANS NOTED THA
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 22575876 ↗
Other Baxter Medical Systems GmbH + Co. KG products
All Baxter Medical Systems GmbH + Co. KG devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page