Deltec recallZ-1280-2025 · Smiths Medical ASD, Inc. · initiated Feb 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Smiths Medical ASD, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The plastic port housing and port reservoir of the ProPort Plastic Implantable Ports may separate because of a manufacturing defect.
FDA root cause category: Process control
What the company is doing
Smiths medical issued an URGENT MEDICAL DEVICE RECALL to its consignees on 02/13/2025 via letter delivered by traceable means. The notice explained the issue, potential risk, and requested the following: Required Actions for Customers 1. Please discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility. 2. Please return all affected product. 3. Inform all potential users of the product in your organization of this notification and complete and return the attached response form to [email protected] within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product and/or has already been used. 4. If you have distributed the product further, immediately notify your accounts that received the affected product and ask them to contact Sedgwick at 1-888-345-2656 (MF, 8am-5pm ET) to obtain a response form.
Recall details
- Recall number
- Z-1280-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 13, 2025
- Posted by the FDA
- Mar 18, 2025
- Recalling firm
- Smiths Medical ASD, Inc.
- Quantity
- 9506 units
- Product code
- LJT
- FDA submissions
- K994216
- Distribution
- Worldwide distribution.
Product description
smiths medical ProPort Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm I.D., 6Fr Introducer, REF 21-4153-24
Code information (models, lots, serial numbers)
UDI/DI 10610586012563, Lot Numbers: 3926121, 3928189, 3953709, 3957303, 4006476, 4008521, 4036894, 4046803, 4096337, 4108664, 4114114, 4114115, 4136363, 4146466, 4153872, 4173601, 4232305, 4235508, 4235598, 4271453, 4291483, 4294059, 4317075, 4317092, 4358052, 4366693, 4366694, 4387733, 4434229, 4449244, 4453603, 4453604, 4453605, 6001912, 6026649, 6051324, 6062012, 6071257, 6077416.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Deltec 21-4153-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall8 reports
Product lines
Other devices under the same FDA submission
12 devices cleared under K994216 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Deltec 21-4155-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall13 reports
- Deltec 21-4183-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall3 reports
- Deltec 21-4187-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall3 reports
- Deltec 21-4152-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall1 report
- Deltec 21-4150-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4151-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4165-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4171-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4172-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4173-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4182-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 recall0 reports
- Deltec 21-4170-24Icu Medical, Inc. · PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 96294).
- Z-1277-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Silicone Catheter, 2.87,,O.D. (8.4Fr) s 1.0mm I.D., REF 21-4150-24
- Z-1278-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Silicone Catheter, 2.8mm (8.4Fr) O.D. x 1.0mm I.D., 9 Fr Introducer, REF 21-4151-24
- Z-1279-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm O.D. (5.8 Fr) x 1.0mm I.D., REF 21-4152-24
- Z-1281-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4155-24
- Z-1282-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm (7.8Fr) O.D. x 1.6mm I.D., 8.6Fr Introducer, REF 21-4165-24
- Z-1283-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, PolyFlow Catheter, 2.6 (7.8Fr) O.D. x 1.6mm I.D., 8.5Fr Introducer, REF 21-4171-24
- Z-1284-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm O.D. (7.8Fr) x 1.6mm I.D., REF 21-4172-24
- Z-1285-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Pre-assembled with PolyFlow Polyurethane Catheter, 2.6mm (7.8Fr) O.D. x 1.6mm I.D., 8.5Fr Introducer, REF 21-4173-24
- Z-1286-2025 · Open, Classifiedsmiths medical ProPort Plastic Venous Access System, Implantable Venous Access Systems, 82-24
- Z-1287-2025 · Open, Classifiedsmiths medical ProPort Low Profile Plastic Venous Access System, PolyFlow Polyurethane Catheter, 1.9mm (5.8Fr) O.D. x 1.0mm., 6Fr Introducer, REF 21-4183-24
- Z-1288-2025 · Open, Classifiedsmiths medical ProPort Low Profile Plastic Venous Access System, Silicone Catheter, 2.8mm (8.4Fr) O.D. x 1.0mm I.D., 9Fr Introducer, REF 21-4187-24
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page