BD Alaris Systems Manager , REF 9601 Infusion Safety… recallZ-1304-2025 · CareFusion 303, Inc. · initiated Feb 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).

FDA root cause category: Under Investigation by firm

What the company is doing

On February 18, 2025, Becton, Dickinson issued a " Urgent Medical Device Product Correction" Notification. to affected consignees via Fed-Ex and E-Mail. 1. Users should review all infusion parameters pre-populated and confirm they are correct before pressing START as stated in the BD Alaris User Manual. 2. Disseminate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 3. BD Project Managers will engage with you to address any questions regarding this notice, and specific next steps for your facility. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form, regardless of whether you have any of the impacted material. This will allow BD to confirm receipt of this notification. 5. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program.

Recall details

Recall number
Z-1304-2025
Status
Open, Classified
Initiated by the company
Feb 18, 2025
Posted by the FDA
Mar 13, 2025
Recalling firm
CareFusion 303, Inc.
Product code
PHC
FDA submissions
K211218
Distribution
US: AZ, CA, DC, FL, GA, IL, IN, MI, MO, NC, NJ, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI OUS: None

Product description

BD Alaris Systems Manager , REF 9601 Infusion Safety Management Software

Code information (models, lots, serial numbers)

Software v12.5.1 or v12.5.2, and prior versions (4.33, 12.0.1, 12.0.2, 12.1, 12.1.2, 12.3) UDIs: 10885403960123 10885403519666 (10885403960116)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96323).