Why it was recalled

During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

FDA root cause category: Software design

What the company is doing

Terumo Cardiovascular issued an URGENT MEDICAL DEVICE REMOVAL notice to its consignees on 03/03/2025 via email. The notice explained the reason for the removal, potential hazard, and requested the following: 1. Review this Medical Device Correction and assure that all users have received notice of this issue. 2. Confirm receipt of this communication by completing and returning the attached Customer Response Form as indicated on the form. 3. Terumo CVS will contact users to discuss removal and correction activities. Note: Terumo CVS recommends that users continue using CDI OneView System H/S Probes while waiting for this removal

Recall details

Recall number
Z-1492-2025
Status
Open, Classified
Initiated by the company
Mar 3, 2025
Posted by the FDA
Mar 31, 2025
Recalling firm
Terumo Cardiovascular Systems Corporation
Quantity
37 units
Product code
DRY
FDA submissions
K234065
Distribution
US: IL, TX, NY, MI, FL, NV, and Australia, New Zealand

Product description

Terumo CDI OneView Monitoring System Hematocrit / Oxygen Saturation Probe, Catalog Number REF CD1754

Code information (models, lots, serial numbers)

UDI/DI 00886799001882, Serial Number range: H0050095-H0050110, H0050112-H0050116, H0050118-H0050124, H0050126, H0050131-H0050134, H0050136, H0050140-H0050142

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K234065 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.