Z1 Hip System recallZ-1329-2025 · Zimmer, Inc. · initiated Feb 12, 2025
Class II Open, Classified All recalls posted in March 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential design issue with the Z1 Offset Stem Inserter leading to fracturing of the distal tip.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification dated 2/12/25 was sent to customers. Our records indicate that you may have received one or more of the affected products. The affected units were distributed between September 2024 and January 2025. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. For cases where the Z1 Offset Stem Inserter was used, increased clinical or radiographic monitoring may be required if the device fractured during use. If the Z1 Offset Stem Inserter was not used or if there was no evidence of fracture during use, there are no additional specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1
Recall details
- Recall number
- Z-1329-2025
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Feb 12, 2025
- Posted by the FDA
- Mar 24, 2025
- Recalling firm
- Zimmer, Inc.
- Quantity
- 35 units
- Product code
- LZO
- FDA submissions
- K233476
- Distribution
- US Nationwide distribution.
Product description
Z1 Femoral Hip System, Product Number 611777612
Code information (models, lots, serial numbers)
UDI-DI: 00887868577888; Lot Numbers: 102646 102777 102778 104095 104324 105277
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Z1™ Hip System 611777612Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall2 reports
Product lines
Other devices under the same FDA submission
15 devices cleared under K233476 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Z1™ Hip System 611201010Zimmer, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate3 reports
- Z1™ Hip System 611201020Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented2 reports
- Z1™ Hip System 611201040Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer2 reports
- Z1™ Hip System 611201050Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer2 reports
- Z1™ Hip System 611201060Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented2 reports
- Z1™ Hip System 611202040Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented2 reports
- Z1™ Hip System 611201000Zimmer, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 report
- Z1™ Hip System 611201030Zimmer, Inc. · Prosthesis, hip, constrained, cemented or uncemented, metal/polymer1 report
- Z1™ Hip System 611201070Zimmer, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 report
- Z1™ Hip System 611202010Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- Z1™ Hip System 611202030Zimmer, Inc. · Prosthesis, hip, semi-constrained, uncemented, metal / polymer, non-porous, calcium phosphate1 report
- Z1™ Hip System 611202070Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.