IntelliSpace Cardiovascular recallZ-1740-2025 · Philips Medical Systems Nederland B.V. · initiated Apr 7, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.
FDA root cause category: Software in the Use Environment
What the company is doing
On April 7, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 1. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Verify that text added into the report conclusion matches the data in the work area interpret sheet (Refer to Attachment A Screenshot 2) " You may continue to use your system(s) in accordance with the intended use and by following the recommendation above. " Circulate this notice to all users of this device so that they are aware of the potential issue. " Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Please review the workflows described in Attachment A and identify whether either workflow is currently in use, or planned to be used, at your facility. Please note your response in the attached customer response form. " If these are not planned to be used at this time, please contact Philips Informatics Customer Support and ensure that this correction is implemented before beginning to use these workflows. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: [email protected]. 2. Actions planned by Philips to correct the problem Once all customer responses have been received, a Philips representative will contact customers who use or may use the affected workflows to schedule a time to install the software solution on your system(s) to resolve the issue (reference FCO-TBD). Please be assured that maintaining the highest level of safety and quality is our greatest priority. If you need any further information or support concerning this issue, please contact your local Philips representative: Philips Informatics Customer Support: 1-800-669-1328, Option 1. This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the us
Recall details
- Recall number
- Z-1740-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 7, 2025
- Posted by the FDA
- May 8, 2025
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 996 units (267 US, 729 OUS)
- Product code
- LLZ
- FDA submissions
- K153022
- Distribution
- Domestic: Nationwide Distribution International: Australia, Austria, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Indonesia, Ireland, Italy, Japan, Luxembourg, Mexico, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Product description
IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.
Code information (models, lots, serial numbers)
Model Number: 830089; Software Versions: 6x, 7.x, and 8.x; UDI-DIs: (01)00884838106970(11)211130(10)6.1.0.0 (Version 6.1.0.0), (01)00884838106970(11)220623(10)6.2.0.0 (Version 6.2.0.0), (01)00884838115378(11)221214(10)7.0.0.0 (Version 7.0.0.0), (01)00884838115378(11)230711(10)7.1.0.0 (Version 7.1.0.0), (01)00884838122000(11)240325(10)8.0.0.0 (Version 8.0.0.0).
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
- IntelliSpace Cardiovascular 7Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls99 reports
- IntelliSpace Cardiovascular 6Philips Medical Systems Nederland B.V. · System, image processing, radiological1 recall118 reports
Product lines
Other devices under the same FDA submission
6 devices cleared under K153022 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- IntelliSpace Cardiovascular 5Philips Medical Systems Nederland B.V. · System, image processing, radiological52 reports
- IntelliSpace Cardiovascular 4.2Philips Medical Systems Nederland B.V. · System, image processing, radiological17 reports
- IntelliSpace Cardiovascular IntelliSpace Cardiovascular 4.1Philips Medical Systems Nederland B.V. · System, image processing, radiological8 reports
- IntelliSpace Cardiovascular IntelliSpace Cardiovascular 3.1Philips Medical Systems Nederland B.V. · System, image processing, radiological5 reports
- IntelliSpace Cardiovascular 3.2Philips Medical Systems Nederland B.V. · System, image processing, radiological5 reports
- IntelliSpace Cardiovascular 2.3Philips Medical Systems Nederland B.V. · System, image processing, radiological1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page