IntelliSpace Cardiovascular recallZ-1740-2025 · Philips Medical Systems Nederland B.V. · initiated Apr 7, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An issue was identified with the software when using the Echo Module of ISCV, the issue may result in missing or incomplete information in the report.

FDA root cause category: Software in the Use Environment

What the company is doing

On April 7, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 1. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Verify that text added into the report conclusion matches the data in the work area interpret sheet (Refer to Attachment A Screenshot 2) " You may continue to use your system(s) in accordance with the intended use and by following the recommendation above. " Circulate this notice to all users of this device so that they are aware of the potential issue. " Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. " Please review the workflows described in Attachment A and identify whether either workflow is currently in use, or planned to be used, at your facility. Please note your response in the attached customer response form. " If these are not planned to be used at this time, please contact Philips Informatics Customer Support and ensure that this correction is implemented before beginning to use these workflows. " Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt of this letter via email to: [email protected]. 2. Actions planned by Philips to correct the problem Once all customer responses have been received, a Philips representative will contact customers who use or may use the affected workflows to schedule a time to install the software solution on your system(s) to resolve the issue (reference FCO-TBD). Please be assured that maintaining the highest level of safety and quality is our greatest priority. If you need any further information or support concerning this issue, please contact your local Philips representative: Philips Informatics Customer Support: 1-800-669-1328, Option 1. This notice has been reported to the appropriate Regulatory Agencies. Adverse reactions or quality problems experienced with the us

Recall details

Recall number
Z-1740-2025
Status
Open, Classified
Initiated by the company
Apr 7, 2025
Posted by the FDA
May 8, 2025
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
996 units (267 US, 729 OUS)
Product code
LLZ
FDA submissions
K153022
Distribution
Domestic: Nationwide Distribution International: Australia, Austria, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Hong Kong, Indonesia, Ireland, Italy, Japan, Luxembourg, Mexico, Netherlands, Norway, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.

Product description

IntelliSpace Cardiovascular software versions 6.x, 7.x, and 8.x; Model Number: 830089.

Code information (models, lots, serial numbers)

Model Number: 830089; Software Versions: 6x, 7.x, and 8.x; UDI-DIs: (01)00884838106970(11)211130(10)6.1.0.0 (Version 6.1.0.0), (01)00884838106970(11)220623(10)6.2.0.0 (Version 6.2.0.0), (01)00884838115378(11)221214(10)7.0.0.0 (Version 7.0.0.0), (01)00884838115378(11)230711(10)7.1.0.0 (Version 7.1.0.0), (01)00884838122000(11)240325(10)8.0.0.0 (Version 8.0.0.0).

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

6 devices cleared under K153022 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page