LinkSymphoKnee recallZ-1374-2025 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Feb 14, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A problem was discovered during the scanning process of the outer carton label because of an incorrect GTIN on the label, which is a GTIN for a different product.
FDA root cause category: Labeling Change Control
What the company is doing
On February 21, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken by the Customer/User: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801 Please contact LinkBio at 973-625-1333 or [email protected] for any questions. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail or by fax.
Recall details
- Recall number
- Z-1374-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 14, 2025
- Posted by the FDA
- Mar 17, 2025
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 20 units
- Product code
- JWH
- FDA submissions
- K202924
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IN, KS, MO, NJ, NV, OH TX, VA, WI and the countries of Germany, Italy, Spain, Switzerland.
Product description
Modular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 13 mm, Product Code: 880-601/13
Code information (models, lots, serial numbers)
Product Code: 880-601/13; UDI-DI: 04026575443949; Lot Numbers: 1943277, 1943280.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- LinkSymphoKnee 880-601/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
Product lines
Other devices under the same FDA submission
116 devices cleared under K202924 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LinkSymphoKnee 880-307/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- LinkSymphoKnee 880-325/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls1 report
- LinkSymphoKnee 880-100/50WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-315/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-300/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96386).
- Z-1372-2025 · Open, ClassifiedModular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 11 mm, Product Code: 880-601/11
- Z-1373-2025 · Open, ClassifiedModular Stem cylindrical, Tilastan, press-fit cementless, L= 80 mm, DIA = 12 mm, Product Code: 880-601/12
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page