Brand Name: Rover recallZ-1578-2025 · Micro-X Ltd. · initiated Mar 28, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Micro-X Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
FDA root cause category: Device Design
What the company is doing
On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: [email protected] and copy [email protected].
Recall details
- Recall number
- Z-1578-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 28, 2025
- Posted by the FDA
- Apr 16, 2025
- Recalling firm
- Micro-X Ltd.
- Quantity
- 31 units
- Product code
- IZL
- FDA submissions
- K211423
- Distribution
- US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US territory).
Product description
Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Serial Number/UDI: MSN0334 (01)09357123000051(11)220921(21)00334; MSN0350 (01)09357123000037(11)230328(21)00350 ; MSN0352 (01)09357123000051(11)230823(21)00352; MSN0364 01)09357123000051(11)230705(21)00364; MSN0382 (01)09357123000051(11)230809(21)00382; MSN0383 (01)09357123000051(11)230814(21)00383; MSN0384 (01)09357123000051(11)230816(21)00384; MSN0385 (01)09357123000051(11)230906(21)00385; MSN0386 (01)09357123000051(11)230908(21)00386; MSN0387 (01)09357123000051(11)230914(21)00387; MSN0388 (01)09357123000051(11)230916(21)00388; MSN0389 (01)09357123000051(11)230918(21)00389; MSN0390 (01)09357123000051(11)231013(21)00390; MSN0395 (01)09357123000051(11)231106(21)00395; MSN0397 (01)09357123000051(11)231118(21)00397; MSN0399 (01)09357123000051(11)240122(21)00399; MSN0400 (01)09357123000051(11)240207(21)00400; MSN0401 (01)09357123000051(11)240131(21)00401; MSN0402 01)09357123000051(11)240202(21)00402; MSN0403 (01)09357123000051(11)240205(21)00403; MSN0405 (01)09357123000051(11)240206(21)00405; MSN0408 (01)09357123000051(11)240308(21)00408; MSN0409 (01)09357123000051(11)240312(21)00409; MSN0410 (01)09357123000051(11)240318(21)00410; MSN0411 (01)09357123000051(11)240319(21)00411; MSN0422 (01)09357123000051(11)240724(21)00422; MSN0423 (01)09357123000051(11)240808(21)00423; MSN0424 (01)09357123000051(11)240809(21)00424; MSN0425 (01)09357123000051(11)240812(21)00425; MSN0426 (01)09357123000051(11)240812(21)00426; MSN0427 (01)09357123000051(11)240812(21)00427;
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)7 recalls0 reports
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)3 recalls0 reports
Other devices under the same FDA submission
1 device cleared under K211423 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Micro-X RoverMicro-X Limited · Solid State X-Ray Imager (Flat Panel/Digital Imager)2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96518).
- Z-1579-2025 · Open, ClassifiedBrand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV35 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorpo…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page