Skip to content

T2 ALPHA recallZ-1620-2025 · Stryker GmbH · initiated Mar 28, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.

FDA root cause category: Device Design

What the company is doing

On March 28, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions needed by Customers and Distributors: Our records indicate that you may have received one or more of the subject devices. It is Stryker's responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Immediately check your internal inventory to locate the product(s) listed on the attached Business Reply Form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Sign and return the enclosed Business Reply Form by email to to confirm receipt of this notification/documenting product segregation. a. Response is required, even if you may not have any physical inventory on site anymore. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also eliminate the need for us to send further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 3. Upon receipt of the completed Business Reply Form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details. b. If you are a distributor, note that you are responsible for notifying your affected customers. 6. Please inform us of any adverse events and/or report them to the Health/C

Recall details

Recall number
Z-1620-2025
Status
Open, Classified
Initiated by the company
Mar 28, 2025
Posted by the FDA
Apr 17, 2025
Recalling firm
Stryker GmbH
Quantity
1588 units
Product code
LXH
Distribution
US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NY, OH, TN, TX, WV.

Product description

Guide Wire with Ruler Tube 3x800 mm DIA; Catalog number/REF: 2351-3080S;

Code information (models, lots, serial numbers)

Catalog number/REF: 2351-3080S; UDI-DI: 07613327361797; Lot #: KU155697, KU155698, KU155699, KU155700, KU155701, KU155702, KU155703, KU155704, KU155705, KU155706, KU155707, KU155708, KU155709, KU155710, KU155711, KU155712.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96555).

Sources and updates