T2 ALPHA recallZ-1621-2025 · Stryker GmbH · initiated Mar 28, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker GmbH or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The metal ring at the end of the Guide Wire with Ruler Tube may detach from the main body of the instrument during use in surgery.
FDA root cause category: Device Design
What the company is doing
On March 28, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions needed by Customers and Distributors: Our records indicate that you may have received one or more of the subject devices. It is Stryker's responsibility as the manufacturer to ensure that customers who may have received these affected products also receive this important communication. We therefore request that you read this notice carefully and complete the following actions. 1. Immediately check your internal inventory to locate the product(s) listed on the attached Business Reply Form, remove them from their point of use, and isolate/quarantine the unit(s) to prevent accidental use. 2. Sign and return the enclosed Business Reply Form by email to to confirm receipt of this notification/documenting product segregation. a. Response is required, even if you may not have any physical inventory on site anymore. It may be that you no longer have any physical inventory on site. Completing this form will allow us to update our records and will also eliminate the need for us to send further unnecessary communications on this matter. Therefore, please complete even if you no longer have any of the subject devices in your physical inventory. 3. Upon receipt of the completed Business Reply Form, Stryker will contact you to arrange for the return of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. 5. If you have further distributed the affected product, please notify the applicable parties at once about this recall. You may copy and distribute this notification letter. a. If possible, inform us if any of the subject devices have been distributed to other organizations, including contact details. b. If you are a distributor, note that you are responsible for notifying your affected customers. 6. Please inform us of any adverse events and/or report them to the Health/C
Recall details
- Recall number
- Z-1621-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 28, 2025
- Posted by the FDA
- Apr 17, 2025
- Recalling firm
- Stryker GmbH
- Quantity
- 6407 units
- Product code
- LXH
- Distribution
- US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NY, OH, TN, TX, WV.
Product description
Guide Wire with Ruler Tube 3x1000 mm DIA; Catalog number/REF: 2351-3100S;
Code information (models, lots, serial numbers)
Catalog number/REF: 2351-3100S; UDI-DI: 07613327361803; Lot #: KU164143, KU164755, KU164756, KU164757, KU164758, KU164759, KU164760, KU164761, KU164762, KU164763, KU164764, KU164765, KU164766, KU164767, KU164768, KU164769, KU164770, KU164771, KU164772, KU164773, KU164774, KU164775, KU164776, KU164777, KU164778, KU164779, KU164780, KU164781, KU164782, KU164783, KU164784, KU165161, KU165162, KU165163, KU165164, KU165165, KU165166, KU165167, KU165168, KU165169, KU165170, KU165171, KU165172, KU165173, KU165174, KU165175, KU184035, KU184036, KU184039, KU184041, KU184042, KU184044, KU184045, KU184046, KU184047, KU184049.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- T2 ALPHA 2351-3100SStryker GmbH · Orthopedic manual surgical instrument1 recall2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96555).
- Z-1620-2025 · Open, ClassifiedGuide Wire with Ruler Tube 3x800 mm DIA; Catalog number/REF: 2351-3080S;
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page