Skip to content

Alinity m recallZ-1665-2025 · Abbott Molecular, Inc. · initiated Mar 20, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Molecular, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

FDA root cause category: Device Design

What the company is doing

An Urgent Field Safety Notice/ Field Correction Recall notification letter was sent to customers. Necessary Actions In accordance with the Alinity m Operations Manual, users should always utilize Personal Protective Equipment when engaging with the instrument. In the event you observe a leak, please take appropriate precautions to prevent exposure and immediately contact your Abbott Molecular Representative for additional troubleshooting information. Please complete and return the associated Customer Reply Form. If you have forwarded this product to any other laboratories, please also forward this letter and customer reply form to that laboratory. Review this information with laboratory personnel and retain this communication for future reference. If you have any questions regarding this communication, please contact your local Abbott representative. We apologize for any inconvenience this may have created for your laboratory.

Recall details

Recall number
Z-1665-2025
Status
Open, Classified
Initiated by the company
Mar 20, 2025
Posted by the FDA
Apr 24, 2025
Recalling firm
Abbott Molecular, Inc.
Product code
OOI
Distribution
Worldwide - US Nationwide distribution.

Product description

The Alinity m System is designed to automate the steps for nucleic acid testing. Model Numbers included in this recall are 08N53-001, 08N53-002, and 08N53-032 (Refurbished).

Code information (models, lots, serial numbers)

All Serial Numbers. 08N53-001, UDI-DI: 00884999047389; 08N53-002, UDI-DI: 00884999048034; 08N53-032, UDI-DI: 00884999047587

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page