Hyperthermia Pump Procedure Kit. Model Number: 902-00045 recallZ-1674-2025 · Belmont Instrument LLC · initiated Mar 31, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Belmont Instrument LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.
FDA root cause category: Process design
What the company is doing
Belmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any product found defective or at the request of the user any affected product that has not been used.
Recall details
- Recall number
- Z-1674-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 31, 2025
- Posted by the FDA
- Apr 25, 2025
- Recalling firm
- Belmont Instrument LLC
- Quantity
- 158 sets
- Product code
- LGZ
- FDA submissions
- K152208
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, CT, DC, IL, IN, KS, KY, MA, ME, MN, NC, ND, NJ, NM, OH, OR, PA, SC, TN, TX, VA, VT, WA, WI, WV and the countries of Canada, Hungary, Korea (the Republic of), Latvia, Malaysia.
Product description
Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) consists of 4.4 Liter Reservoir (P/N 902-00034P), Heat Exchanger set (P/N 902-00006P) and HP Procedure Pack (P/N 902-00048P). The three components are packaged separately in individual pouches and boxes and placed in the same case for shipment. All three components are connected prior to use. The female quick connector that may be cracked is located on the Heat Exchanger set (P/N 902-00006P).
Code information (models, lots, serial numbers)
Model Number: 902-00045. UDI-DI: 00896128002589. Lot Numbers: 20241005, 20241108
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Hyperthermia Pump Procedure Kit 902-00045Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall4 reports
Other devices under the same FDA submission
7 devices cleared under K152208 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Straight In/Out Pt Line Kit 902-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall3 reports
- Straight Inflow/Bifurcated Outflow Pt Line Kit 902-00038Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall3 reports
- Hyperthermia Pump, 120V 902-00001Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid24 reports
- HEAT EXCHANGER SET (1 Piece) 902-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid2 reports
- Hyperthermia Pump St In/Out Pt Line 902-00039PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid2 reports
- English, Hyperthermia Pump, 230V 902-00001ABelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid2 reports
- 4.4 Liter Reservoir 902-00034Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 96617).
- Z-1675-2025 · Open, ClassifiedStraight Inflow/Outflow Patient Line Kit. Model Number: 902-00037. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient Line…
- Z-1676-2025 · Open, ClassifiedStraight Inflow/Bifurcated Outflow Line Kit. Model Number: 902-00038. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient L…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗