LinkSymphoKnee recallZ-1701-2025 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Apr 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery. For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1. Assembly/Replacement 3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment. 4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps. Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws. Type of Action by the Company: LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries. Other Information: Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio ([email protected]). Please contact LinkBio at 973-625-1333 or [email protected] for any additional questions.
Recall details
- Recall number
- Z-1701-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 3, 2025
- Posted by the FDA
- Apr 30, 2025
- Recalling firm
- Waldemar Link GmbH & Co. KG (Mfg Site)
- Quantity
- 441 units
- Product code
- JWH
- FDA submissions
- K202924
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI.
Product description
L-Shaped Femoral Augment; Item Numbers: (1) 880-320/12 (Sz. 0 - cemented sm), (2) 880-320/22 (Sz. 0 - cemented lg), (3) 880-321/12 (Sz. 1-2 - cemented sm), (4) 880-321/22 (Sz. 1-2 - cemented lg), (5) 880-323/13 (Sz. 3-4 - cemented sm), (6) 880-323/23 (Sz. 3-4 - cemented lg), (7) 880-325/13 (Sz. 5-6 - cemented sm), (8) 880-325/23 (Sz. 5-6 - cemented lg), (9) 880-327/13 (Sz. 7-8 - cemented sm), (10) 880-327/23 (Sz. 7-8 - cemented lg), (11) 880-329/13 (Sz. 9-10 - cemented sm), (12) 880-329/23 (Sz. 9-10 - cemented lg).
Code information (models, lots, serial numbers)
Item Number/UDI-DI: (1) 880-320/12 (04026575258048), (2) 880-320/22 (04026575258055), (3) 880-321/12 (04026575258062), (4) 880-321/22 (04026575258079), (5) 880-323/13 (04026575258086), (6) 880-323/23 (04026575258109), (7) 880-325/13 (04026575258123), (8) 880-325/23 (04026575258147), (9) 880-327/13 (04026575258154), (10) 880-327/23 (04026575258161), (11) 880-329/13 (04026575258178), (12) 880-329/23 (04026575258185). Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025.
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- LinkSymphoKnee 880-325/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER2 recalls1 report
- LinkSymphoKnee 880-320/12WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-320/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-321/12WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-321/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-323/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-323/23WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-325/23WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-327/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-327/23WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-329/13WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-329/23WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
Product lines
Other devices under the same FDA submission
105 devices cleared under K202924 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- LinkSymphoKnee 880-307/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- LinkSymphoKnee 880-307/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER2 recalls2 reports
- LinkSymphoKnee 880-100/50WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-305/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-313/22WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-315/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall1 report
- LinkSymphoKnee 880-300/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-301/21WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
- LinkSymphoKnee 880-303/11WALDEMAR LINK GmbH & Co. KG · PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96643).
- Z-1699-2025 · Open, ClassifiedDistal Femoral Augment; Item Numbers: (1) 880-300/11 (Sz. 0 - cemented sm), (2) 880-300/21 (Sz. 0 - cemented lg), (3) 880-301/11 (Sz. 1-2 - cemented sm), (4) 880-301/21 (Sz. 1-2 - …
- Z-1700-2025 · Open, ClassifiedPosterior Femoral Augment; Item Numbers: (1) 880-310/11 (Sz. 0 - cemented sm), (2) 880-310/21 (Sz. 0 - cemented lg), (3) 880-311/11 (Sz. 1-2 - cemented sm), (4) 880-311/21 (Sz. 1-2…
- Z-1702-2025 · Open, ClassifiedTibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented sm), (3) 880-331/13 (Sz. 1-2 - cemented md), (4) 880-331/21 (Sz. 1-2 - ce…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page