Tempus Pro recallZ-1666-2025 · Remote Diagnostic Technologies Ltd. · initiated Apr 16, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Remote Diagnostic Technologies Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.

FDA root cause category: Device Design

What the company is doing

On April 16, 2025, Remote Diagnostic Technologies Ltd (subsidiary of Philips) issued a Urgent Medical Device Correction to affected consignees via mail. Philips ask consignees to take the following: 1. DO NOT connect the Video Laryngoscope USB probe to the Tempus Pro during device boot-up. Once Patient Details dialog appears, enter patient data, then plug in the Video Laryngoscope USB probe. 2. If the Tempus Pro Monitor displays the error dialog screen (where the user is asked to restart the device), press the Shutdown Button and restart the Tempus Pro Monitor. 3. Keep a copy of this letter with or near your Tempus Pro Monitor. 4. Complete and return the Urgent Field Safety Notification Response Form included with this letter within 30 days of receipt of this notice. 5. This notice must be shared with all relevant personnel within your organization and with any organization where the potentially affected devices have been transferred.

Recall details

Recall number
Z-1666-2025
Status
Open, Classified
Initiated by the company
Apr 16, 2025
Posted by the FDA
Apr 24, 2025
Recalling firm
Remote Diagnostic Technologies Ltd.
Quantity
6,961 units
Product code
MHX
FDA submissions
K130773, K201746
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.

Product description

Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]

Code information (models, lots, serial numbers)

All Lots/ UDI-DI(s): 00-1004-R - 05060472440020 & 05060472442901; 00-1007-R - 05060472440013 & 05060472442918; 00-1024-R - 05060472441027 & 05060472442925; 00-1026-R - 05060472441058 & 05060472442932; 00-1004 - 05060472440624; 00-1007 - 05060472440655

Devices named in this recall

10 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page