Skip to content

Artix MT recallZ-1728-2025 · Inari Medical - Oak Canyon · initiated Apr 8, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Inari Medical - Oak Canyon or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Thrombectomy device specifies vessel range of 2.5-6 mm on the pouch, carton, and Instructions for Use rather than the correct vessel range of 3-6 mm, which could incrementally increase the radial force on the artery, which has the potential of vascular complications.

FDA root cause category: Labeling Change Control

What the company is doing

On 4/8/2025, correction and removal notices were distributed to customers who were asked to do the following: 1) Review notice and the accompanying copy of the device IFU (IU-01087 Rev. C) and disseminate it to any applicable personnel and device users at your facility and network to ensure they are aware of the unused device replacement. 2) Notice should be shared with any organization where affected devices may have been transferred. 3) Complete and return the acknowledgement reply form via email to If you have any questions, please contact your local sales representative, or the firm's Customer Service at 877-923-4747, or Quality department at [email protected].

Recall details

Recall number
Z-1728-2025
Status
Open, Classified
Initiated by the company
Apr 8, 2025
Posted by the FDA
May 6, 2025
Recalling firm
Inari Medical - Oak Canyon
Quantity
45
Product code
QEW
FDA submissions
K241894
Distribution
US Nationwide distribution in the states of FL, IL, WI, KY, TX, CA, OH, GA, AK, MI, PA, WA, CO, OH.

Product description

Artix MT Thrombectomy Device, REF: 32-102

Code information (models, lots, serial numbers)

UDI-DI: 00810123710584, Lot: 24100082, Expiration: 14 April 2025. IFU: IU-01087 Rev. B

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K241894 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates