Skip to content

Centerline Biomedical IOPS Guidewire 2, REF ATW-2 recallZ-1760-2025 · Centerline Biomedical Inc · initiated Apr 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Centerline Biomedical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Centerline Biomedical has determined that the coating on the guidewires for lot 2404-2005 may delaminate during use of the guidewire.

FDA root cause category: Under Investigation by firm

What the company is doing

Centerline Biomedical issued an "URGENT MEDICAL DEVICE RECALL" notice dated 4/11/2025 to its consignees on 04/11/2025 via hand delivery. The notice explained the issue with the device, potential risk to the patient and actions to be taken. The consignees were requested to Complete the Customer Acknowledgement letter in the possession of your Clinical Account Manager and return all affected products to Clinical Account Manager. For questions contact: Director Quality and Regulatory, Phone Number: (330) 577-5033, Email: [email protected].

Recall details

Recall number
Z-1760-2025
Status
Open, Classified
Initiated by the company
Apr 11, 2025
Posted by the FDA
May 19, 2025
Recalling firm
Centerline Biomedical Inc
Quantity
80 units
Product code
DQK
FDA submissions
K190106, K230309, K241243, K242133, K243842
Distribution
US distribution to states of: Pennsylvania, Florida, Texas, New York, North Carolina, New Mexico, Virginia, Tennessee

Product description

Centerline Biomedical IOPS Guidewire 2, REF ATW-2; Sensorized wire used to navigate through vasculature to facilitate placement of a catheter

Code information (models, lots, serial numbers)

UDI/DI 00843152102037, Lot Number 2404-2005

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K190106, K230309, K241243, K242133, K243842 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates