IOPS™ ATW-2Recalls & reported problems
Centerline Biomedical, Inc. · Indications for Use: The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. IOPS™ Guidewire 2 is a sensorized wire used to navigate through vasculature to facilitate placement of a catheter. The guidewire is a single-use, externally communicating device that comes into contact with circulating blood for a duration of less than or equal to 24 hours. · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843152102037), received between Jun 13, 2023 and Feb 4, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Frayed” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 11, 2025Open, Classified
Centerline Biomedical has determined that the coating on the guidewires for lot 2404-2005 may delaminate during use of the guidewire.
Z-1760-2025 · Root cause: Under Investigation by firm · FDA record ↗
Most reported problems
Device problems
- Material Frayed2
- Difficult to Remove1
- Peeled/Delaminated1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 4, 2025Peeled/Delaminated; Peeled/Delaminated
DURING A BEVAR (BRANCHED ENDOVASCULAR AORTIC REPAIR) PROCEDURE, AN IOPS ATW-2 GUIDEWIRE WAS DAMAGED UPON REMOVAL FROM THE PATIENT. DURING REMOVAL, THE PHYSICIAN NOTED SOME RESISTANCE BUT CONTINUED TO RETRACT THE GUIDEWIRE. AFTER THE GUIDEWIRE WAS REMOVED, IT WAS FOUND THAT THE HYDROPHILIC COATING ON THE DISTAL TIP HAD BEGUN TO DELAMINATE. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21306703 ↗MalfunctionJun 13, 2023Material Frayed; Material Frayed
DURING A MINIMALLY INVASIVE VASCULAR SURGERY PERFORMED AT VANDERBILT UNIVERSITY MEDICAL CENTER, DR. ADVANCED THE IOPS GUIDEWIRE ATW-2 THROUGH A CATHETER AND SHEATH (BOTH FROM ANOTHER MANUFACTURER) THAT HAD ALREADY CANNULATED THE PMEG FENESTRATION TO ACCESS THE SMA. THE IOPS WIRE WAS SWITCHED OUT FOR ANOTHER WIRE FROM ANOTHER MANUFACTURER. BEFORE INSERTING IT BACK INTO THE SHEATH TO ACCESS THE CELIAC ARTERY, DR. NOTICED THAT THE WIRE COVERING OVER THE SENSOR WAS BECOMING DETACHED AND DECIDED NOT TO PROCEED ANY FURTHER. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17119007 ↗MalfunctionJun 13, 2023Material Frayed; Difficult to Remove; Material Frayed; Difficult to Remove
DURING A MINIMALLY INVASIVE VASCULAR SURGERY PERFORMED AT VANDERBILT UNIVERSITY MEDICAL CENTER, THE IOPS GUIDEWIRE ATW-2 WAS USED IN CONJUNCTION WITH A 7FR STEERABLE SHEATH AND A 90 CM CATHETER FROM ANOTHER MANUFACTURER. THE GOAL WAS TO CANNULATE THE CELIAC TRUNK OFF THE AORTA. THE GUIDEWIRE WAS IN THE CATHETER, AND THE CATHETER/GUIDEWIRE WERE IN THE STEERABLE SHEATH. THE SHEATH WAS PLACED INTO POSITION AND STEERED TO ENGAGE THE CELIAC OSTIUM SUCH THAT THE GUIDEWIRE CAN BE ADVANCED TO ATTEMPT TO CANNULATE THE VESSEL. THE DOCTOR FELT RESISTANCE TRYING TO PULL THE GUIDEWIRE BACK INTO THE CATHETER/SHEATH. ONCE REMOVED FROM THE PATIENT, THE HYDROPHILIC COATING ON THE TIP OF THE GUIDEWIRE WAS FOUND TO BE DAMAGE UPON INSPECTION. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17119000 ↗
Other Computer, Diagnostic, Programmable devices
- ZIO AT Patch A100A100XIrhythm Technologies, Inc. · Computer, Diagnostic, Programmable1 recall14 deaths629 reports
- CARTO 3 FG540000Biosense Webster (Israel) Ltd. · Computer, diagnostic, programmable1 death437 reports
- Sensis 10764561Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE5 recalls8 reports
- Sensis Vibe Hemo 11007641Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death22 reports
- Sensis Vibe Combo 11007642Siemens Healthcare GmbH · COMPUTER, DIAGNOSTIC, PROGRAMMABLE4 recalls1 death6 reports
- Change Healthcare Cardiology Hemo 15.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable4 recalls1 report
- Change Healthcare Cardiology Hemo 15.0.1Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls1 death1 report
- Change Healthcare Cardiology Hemo 14.3.2Change Healthcare Canada Company · Computer, Diagnostic, Programmable3 recalls0 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.