IOPS™ ATW-2Recalls & reported problems

Centerline Biomedical, Inc. · Indications for Use: The IOPS (Intra-Operative Positioning System) is intended for the evaluation of vascular anatomy as captured via 3D modeling from previously acquired scan data. It is intended for real time tip positioning and navigation using sensor-equipped compatible catheters and guidewires used in endovascular interventions in the peripheral, aortic and aortic side branch vasculature. The system is indicated for use as an adjunct to fluoroscopy. The IOPS does not make a diagnosis. IOPS™ Guidewire 2 is a sensorized wire used to navigate through vasculature to facilitate placement of a catheter. The guidewire is a single-use, externally communicating device that comes into contact with circulating blood for a duration of less than or equal to 24 hours. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00843152102037), received between Jun 13, 2023 and Feb 4, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Material Frayed” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 11, 2025Open, Classified

    Centerline Biomedical has determined that the coating on the guidewires for lot 2404-2005 may delaminate during use of the guidewire.

    Z-1760-2025 · Root cause: Under Investigation by firm · FDA record ↗

Most reported problems

Device problems

  • Material Frayed2
  • Difficult to Remove1
  • Peeled/Delaminated1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 4, 2025Peeled/Delaminated; Peeled/Delaminated

    DURING A BEVAR (BRANCHED ENDOVASCULAR AORTIC REPAIR) PROCEDURE, AN IOPS ATW-2 GUIDEWIRE WAS DAMAGED UPON REMOVAL FROM THE PATIENT. DURING REMOVAL, THE PHYSICIAN NOTED SOME RESISTANCE BUT CONTINUED TO RETRACT THE GUIDEWIRE. AFTER THE GUIDEWIRE WAS REMOVED, IT WAS FOUND THAT THE HYDROPHILIC COATING ON THE DISTAL TIP HAD BEGUN TO DELAMINATE. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21306703 ↗
  • MalfunctionJun 13, 2023Material Frayed; Material Frayed

    DURING A MINIMALLY INVASIVE VASCULAR SURGERY PERFORMED AT VANDERBILT UNIVERSITY MEDICAL CENTER, DR. ADVANCED THE IOPS GUIDEWIRE ATW-2 THROUGH A CATHETER AND SHEATH (BOTH FROM ANOTHER MANUFACTURER) THAT HAD ALREADY CANNULATED THE PMEG FENESTRATION TO ACCESS THE SMA. THE IOPS WIRE WAS SWITCHED OUT FOR ANOTHER WIRE FROM ANOTHER MANUFACTURER. BEFORE INSERTING IT BACK INTO THE SHEATH TO ACCESS THE CELIAC ARTERY, DR. NOTICED THAT THE WIRE COVERING OVER THE SENSOR WAS BECOMING DETACHED AND DECIDED NOT TO PROCEED ANY FURTHER. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17119007 ↗
  • MalfunctionJun 13, 2023Material Frayed; Difficult to Remove; Material Frayed; Difficult to Remove

    DURING A MINIMALLY INVASIVE VASCULAR SURGERY PERFORMED AT VANDERBILT UNIVERSITY MEDICAL CENTER, THE IOPS GUIDEWIRE ATW-2 WAS USED IN CONJUNCTION WITH A 7FR STEERABLE SHEATH AND A 90 CM CATHETER FROM ANOTHER MANUFACTURER. THE GOAL WAS TO CANNULATE THE CELIAC TRUNK OFF THE AORTA. THE GUIDEWIRE WAS IN THE CATHETER, AND THE CATHETER/GUIDEWIRE WERE IN THE STEERABLE SHEATH. THE SHEATH WAS PLACED INTO POSITION AND STEERED TO ENGAGE THE CELIAC OSTIUM SUCH THAT THE GUIDEWIRE CAN BE ADVANCED TO ATTEMPT TO CANNULATE THE VESSEL. THE DOCTOR FELT RESISTANCE TRYING TO PULL THE GUIDEWIRE BACK INTO THE CATHETER/SHEATH. ONCE REMOVED FROM THE PATIENT, THE HYDROPHILIC COATING ON THE TIP OF THE GUIDEWIRE WAS FOUND TO BE DAMAGE UPON INSPECTION. NO PATIENT ADVERSE EFFECTS WERE REPORTED DUE TO THIS OCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17119000 ↗