Skip to content

iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen… recallZ-1727-2025 · Medical Depot Inc. · initiated Apr 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medical Depot Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT NOTICE MEDICAL DEVICE RECALL dated 4/11/25 was emailed to consignees. The recall notification instructs consignees on how to identify affected devices in their possession, destroy the affected charger cord, and provides instructions on how to obtain a replacement charger cord through Drive Devilbiss. If product was further distributed, consignees are to contact their customers; Devilbiss has a draft recall notification and recall website available for their consignees who further distributed product. For those units that are rentals, Devilbiss recommends that consignees check devices as they return from rental use. Dan Hinkbein is available to assist consignees who distributed affected devices fulfill recall actions and can be contacted at [email protected]. Consignees with any issues or questions can call Drive Devilbiss Healthcare at 888-770-7017 Monday through Friday from 7:30 AM to 5:00 PM CST. Questions can also be emailed to [email protected]. The firm has set up a website for this recall event at www.recallrtr.com/iGo2-car-cord.

Recall details

Recall number
Z-1727-2025
Status
Open, Classified
Initiated by the company
Apr 11, 2025
Posted by the FDA
May 9, 2025
Recalling firm
Medical Depot Inc.
Quantity
89,532 units
Product code
CAW
FDA submissions
K071397
Distribution
Worldwide distribution - US Nationwide and the countries of Albania, Austria, Bahrain, Belgium, Bosnia and Herzegovina, Bulgaria, Canada, China, Colombia, Croatia, Cyprus, Czech Republic, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, ISRAEL, Italy, JAPAN, Kenya, Kuwait, Latvia, Libya, Lithuania, Malaysia, Maldives, Nepal, Netherlands, Norway, Oman, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Slovak Republic, Slovenia, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Trinidad & Tobago, Turkey, Uganda, Ukraine, United Arab Emirates.

Product description

iGo2 DC Car Adapter to be used with iGo2 Portable Oxygen Concentrator Systems. Power cord allows the user to operate the device in a vehicle.

Code information (models, lots, serial numbers)

DC Cord Code Info: Catalog No. DV6X-619; UDI 885304020585; Cord Revision: Rev E; Lot No. beginning with 2023-01-01 through 2024-12-31. iGo2 Portable Oxygen Concentrator Systems Code Info: Catalog No. 125D, 125D-XB, 125D-BT, 125D-BT-XB, 125D-ARYA-XB; UDI 885304022237, 885304032250, 885304033257, 885304033240, 885304033516; Lot No. D23125001DS to D23A26121DS, D23706001DA to D23B16056DA, F23B06001DS to F25108143DS, F23313001DH to F25109259DH, F24212001DA to F25217150DA.

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

7 devices cleared under K071397 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates