DeVilbiss DV6X-619Recalls & reported problems
Devilbiss Healthcare LLC · DC car adaptor · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00885304020585), received between Apr 17, 2025 and Apr 17, 2025 — classified as 1 malfunction.
The most frequently reported specific device problem is “Component Misassembled” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
2 other recalls cover devices cleared under the same FDA submission (K071397); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Apr 11, 2025Open, Classified
Potential for DC Power Supply housing to become hot to the touch and deform the plastic housing.
Z-1727-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Most reported problems
Device problems
- Component Misassembled1
Patient problems
- Superficial (First Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2025Component Misassembled; Component Misassembled
DEVILBISS HEALTHCARE WAS NOTIFIED OF AN INCIDENT INVOLVING A REPORT OF A DC CAR CHARGER BY THE PROVIDER, WHO STATED THAT THE CHARGER TIP BURNED THE END USER'S HAND DURING USE. THE END USER PLACED ICE ON THE CONTACT SPOT. THERE WAS NO SERIOUS INJURY THAT REQUIRED MEDICAL ATTENTION. ALTHOUGH THE END USER SUSTAINED A BURN RELATED TO THE USE OF THE DEVICE, THE INFORMATION PROVIDED INDICATES NO MEDICAL INTERVENTION WAS REQUIRED AND THUS NOT A REPORTABLE SERIOUS INJURY. THIS DEVICE IS BEING RECALLED. THE RES RECALL NUMBER HAS YET TO BE ISSUED. THE ROOT CAUSE IS THAT THE SUPPLIER OF THE CHARGER SWITCHED THE LOCATION OF THE FUSE AND THE SPRING DURING ASSEMBLY. THERE HAVE BEEN NO COMPLAINTS OF SERIOUS INJURY TO DATE ASSOCIATED WITH THIS ISSUE. THERE HAS BEEN ONE OTHER SUBJECTIVE COMPLAINT OF MINOR THERMAL INJURY TO FINGERS/HAND REPORTED TO DATE, WHICH RESULTED IN MINOR TRANSIENT PAIN, AND NO NEED FOR MEDICAL INTERVENTION. THESE TWO COMPLAINTS REPRESENT (B)(4) OF THE AT RISK POPULATION OF DEVICES.
Patient problems: Superficial (First Degree) Burn
MAUDE report 21854445 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K071397. One submission often covers many models, so check the product and lot details in each record.
- Z-1558-2023 · Mar 9, 2023 · Open, ClassifiedDEVILBISS HEALTHCARE 1025 SERIES OXYGEN CONCENTRATOR, Model Numbers 1025DS, 1025UK, 1025KS, and 1025KS-AR
- Z-0724-2023 · Nov 18, 2022 · Open, ClassifiedDeVilbiss, 525 5-liter Oxygen Concentrator Models 525DS, 525DS-Q
Other Generator, oxygen, portable devices
- DeVilbiss 525DSDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls14 deaths287 reports
- Inogen One G5 Oxygen Concentrator - Bluetooth 1260 BluetoothInogen, Inc. · Generator, Oxygen, Portable3 deaths235 reports
- EverFlo EverFlo w/O2 Percentage Indicator, 230V, EURespironics, Inc. · Generator, oxygen, portable5 deaths112 reports
- Inogen One G3 Oxygen Concentrator 1050Inogen, Inc. · Generator, Oxygen, Portable2 deaths111 reports
- DeVilbiss 125DDevilbiss Healthcare LLC · Generator, Oxygen, Portable2 recalls1 death3 reports
- EverFlo EVERFLO NON-O2 Percentage Indicator, 120V, US,CARespironics, Inc. · Generator, oxygen, portable19 deaths80 reports
- EverFlo EVERFLO w/Oxygen Percentage Indicator, 120V US, CARespironics, Inc. · Generator, oxygen, portable2 deaths71 reports
- SimplyGo Mini SimplyGo Mini, USRespironics, Inc. · Generator, oxygen, portable1 recall1 death5 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.