Tempus Pro recallZ-1729-2025 · Remote Diagnostic Technologies Ltd. · initiated Apr 15, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Remote Diagnostic Technologies Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
FDA root cause category: Software design
What the company is doing
On 4/15/2025, correction notices were mailed to customers and distributors informing them of the following: I) The User/Operator Manual provides precautions and warnings on maintaining appropriate power or battery supply, which may reduce the risk of occurrence of an "MPM Application" software error message: -Warning (In Chapter 2, Getting started): Ensure the latches on both sides of the battery are fully engaged prior to using the Tempus - an incorrectly fitted battery could result in the tempus losing power during use. -Note: Before removing the battery, you must switch off the Tempus Pro by pressing the power button. Do not remove the battery when the Tempus is on (unless there is a power supply attached to the device). II) If you have confirmed that your Tempus Pro has the standard (Louvain) algorithm, continue to follow the User/Operator Manual, which warns users to not make evaluations based solely on interpretation statements from the algorithm. -Warning (In Chapter 6, Taking medical readings): ECG interpretation performed by software is not a substitute for review and evaluation of ECG recordings by a qualified clinician. Keep a copy of this notice with your device until you receive the correction. Complete and return the response form via email to https://philips.efmfeedback.com/se/705E3ED80DE11FAF In addition, distributors were asked to do the following: - Modify the Response Form to substitute your firm's email and fax information. - Send a copy of this Urgent Medical Device Correction (with modified response form) to each customer to whom you distributed the affected product. Philips has developed a software update to resolve these issues that is expected to be available Q3 2025 via FCO86700012. If you need any further information or support concerning this issue, please contact the firm at (800) 722-9377 or email [email protected]
Recall details
- Recall number
- Z-1729-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 15, 2025
- Posted by the FDA
- May 7, 2025
- Recalling firm
- Remote Diagnostic Technologies Ltd.
- Quantity
- 6961
- Product code
- MHX
- FDA submissions
- K201746
- Distribution
- Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.
Product description
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Code information (models, lots, serial numbers)
Software versions prior to vx.40. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls8 deaths1,217 reports
- Tempus Pro 00-1024-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls541 reports
- Tempus Pro 00-1026-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls1 death324 reports
- Tempus Pro 00-1026-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls233 reports
- Tempus Pro 00-1007-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)7 recalls114 reports
- Tempus Pro 00-1004-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)6 recalls36 reports
- Tempus Pro 00-1007-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)5 recalls64 reports
- Tempus Pro 00-1004-RRemote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls23 reports
- Tempus Pro 00-1007Remote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls2 reports
- Tempus Pro 00-1004Remote Diagnostic Technologies Limited · Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)4 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page