Skip to content

BIB recallZ-1754-2025 · Numed Inc · initiated Apr 21, 2025

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Numed Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Balloon extension lumens are not printed correctly. Specifically, the inner balloon inflation hub was labeled as the outer balloon, and the outer balloon inflation hub was labeled as the inner balloon.

FDA root cause category: Mixed-up of materials/components

What the company is doing

NuMED sent a "URGENT: MEDICAL DEVICE RECALL" letter dated 4/21/2025 via email on 4/21/2025 to their single customer. The customer was instructed to immediately return the affected units as soon as possible. If you have any questions or concerns, please contact Regulatory Affairs Manager at (315) 328-4491, or via email at [email protected].

Recall details

Recall number
Z-1754-2025
Status
Terminated
Initiated by the company
Apr 21, 2025
Posted by the FDA
May 13, 2025
Recalling firm
Numed Inc
Quantity
4 units
Product code
NVM
FDA submissions
K050857
Distribution
US distribution to Pennsylvania.

Product description

BIB Balloon in Balloon Dilatation Catheter. Model Number: 420

Code information (models, lots, serial numbers)

Model Number 420. UDI-DI: 04046964899768. Lot Numbers: BIB-16972 (Expiration date: 11/30/2028), BIB-16972-1 (Expiration date: 12/31/2028). Catheter Product Code: BB005.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

5 devices cleared under K050857 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page