Skip to content

Access Testosterone assay, Catalog Number 33560 recallZ-1901-2025 · Beckman Coulter, Inc. · initiated May 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

FDA root cause category: Equipment maintenance

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 4/23/25 was sent to customers. ACTION: Discontinue using and discard all remaining Access Testosterone reagent packs from the lot listed above. Beckman Coulter recommends performing a retrospective review on patient results generated with the Access Testosterone lot listed above and assessing the need for retesting. Contact your local Beckman Coulter representative for replacement product requests. Beckman Coulter is no longer distributing the affected Access Testosterone reagent pack lot number and has corrected the manufacturing issue that led to cross-contamination. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com If you need replacement product: o Complete the attached Replacement Order Form and email to [email protected] or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821

Recall details

Recall number
Z-1901-2025
Status
Open, Classified
Initiated by the company
May 5, 2025
Posted by the FDA
Jun 2, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
70 units
Product code
CDZ
FDA submissions
K223405
Distribution
US State: FL. UAE, Vietnam

Product description

Access Testosterone assay, Catalog Number 33560

Code information (models, lots, serial numbers)

UDI-DI: 15099590231347; Serial Number: 538081

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page