Medical device recalls: June 2025
The FDA posted 156 medical device recall records in June 2025, from 84 recall events by 71 companies. 19 are Class I, the most serious category.
Each product in a recall gets its own record, so one recall event can cover several records. What recall classes mean.
Companies with the most recall events
- Beckman Coulter, Inc. · 4 events, 4 records
- Siemens Medical Solutions USA, Inc · 2 events, 6 records
- Dexcom, Inc. · 2 events, 5 records, 4 Class I
- Olympus Corporation of the Americas · 2 events, 5 records
- Integra LifeSciences Corp. (NeuroSciences) · 2 events, 4 records, 4 Class I
- MEDLINE INDUSTRIES, LP - Northfield · 2 events, 4 records
- Ortho-Clinical Diagnostics, Inc. · 2 events, 4 records
- Stryker GmbH · 2 events, 3 records
FDA root cause
- Under Investigation by firm · 36
- Process control · 28
- Software design · 15
- Process design · 12
- Nonconforming Material/Component · 10
- Component design/selection · 10
- Software design (manufacturing process) · 6
- Radiation Control for Health and Safety Act · 6
Class I recalls (19)
The most serious: a reasonable probability that using the device will cause serious health problems or death.
- Z-2013-2025 · Maquet Critical Care AB · posted Jun 27, 2025 · Open, Classified
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
- Z-2014-2025 · Maquet Critical Care AB · posted Jun 27, 2025 · Open, Classified
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
- Z-2015-2025 · Maquet Critical Care AB · posted Jun 27, 2025 · Open, Classified
Potential for inaccurate measurement of the patient circuit compliance during the patient circuit test due to combination of incorrect pre-use leakage test and use of an incorrect active humidifier.
- Z-1904-2025 · SunMed Holdings, LLC · posted Jun 20, 2025 · Open, Classified
The impacted tape was manufactured with incorrect information on the tape. Incorrect values are printed in the Red zone, Orange zone, and Grey zone. Using the incorrect values in the Red zone could lead to shocking a patient with an excessive dose of…
- Z-1958-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Jun 20, 2025 · Open, Classified
Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.
- Z-1959-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Jun 20, 2025 · Open, Classified
Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.
- Z-1960-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Jun 20, 2025 · Open, Classified
Devices contain potentially out-of-specification levels of endotoxins and lack of sterility assurance.
- Z-1961-2025 · Integra LifeSciences Corp. (NeuroSciences) · posted Jun 20, 2025 · Open, Classified
Lack of sterility assurance.
- Z-1902-2025 · Cook Incorporated · posted Jun 16, 2025 · Open, Classified
Cook Medical identified that catheters supplied in the affected device lots may experience tip separation.
- Dexcom G6 Glucose Receiver, SKUs (Part Number): STK-FM-001 (MT27408-1) & STK-FR-001 (MT27408-1)Class IZ-1943-2025 · Dexcom, Inc. · posted Jun 16, 2025 · Open, Classified
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead…
- Z-1944-2025 · Dexcom, Inc. · posted Jun 16, 2025 · Open, Classified
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead…
- Z-1945-2025 · Dexcom, Inc. · posted Jun 16, 2025 · Open, Classified
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead…
- Dexcom One Continuous Glucose Monitoring System, SKUs(Part Number): STK-DO-006 (MT27409-1), ↗Class IZ-1946-2025 · Dexcom, Inc. · posted Jun 16, 2025 · Open, Classified
Defective foam or an assembly error may cause the receiver speaker to lose contact with the printed circuit board, leading to missed audible alerts for low or high blood glucose values. A missed audible alert for low or high blood glucose values could lead…
- Z-1915-2025 · Covidien · posted Jun 9, 2025 · Open, Classified
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to…
- Z-1865-2025 · Zyno Medical LLC · posted Jun 6, 2025 · Open, Classified
Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1866-2025 · Zyno Medical LLC · posted Jun 6, 2025 · Open, Classified
Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1867-2025 · Zyno Medical LLC · posted Jun 6, 2025 · Open, Classified
Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1868-2025 · Zyno Medical LLC · posted Jun 6, 2025 · Open, Classified
Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1869-2025 · Fresenius Kabi USA, LLC · posted Jun 6, 2025 · Open, Classified
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
Class II recalls (135)
The device may cause temporary or medically reversible health problems, or the chance of serious harm is remote.
- Z-2043-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jun 30, 2025 · Open, Classified
Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.
- Z-2044-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jun 30, 2025 · Open, Classified
Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.
- Z-2045-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jun 30, 2025 · Open, Classified
Medline Industries, LP. is issuing a recall for specific Medline kits containing BD Insyte Autoguard Shielded IV Catheters 22 GA x 1.00 IN which was identified that during insertion of the Catheter, the needle is slow to retract or fails to retract.
- Z-2019-2025 · Siemens Medical Solutions USA, Inc · posted Jun 27, 2025 · Open, Classified
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2020-2025 · Siemens Medical Solutions USA, Inc · posted Jun 27, 2025 · Open, Classified
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2021-2025 · Siemens Medical Solutions USA, Inc · posted Jun 27, 2025 · Open, Classified
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Z-2022-2025 · Siemens Medical Solutions USA, Inc · posted Jun 27, 2025 · Open, Classified
In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.
- Brand Name: VITROSClass IIZ-2038-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jun 27, 2025 · Open, Classified
A software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading.…
- Brand Name: VITROSClass IIZ-2039-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jun 27, 2025 · Open, Classified
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading.…
- Brand Name: VITROSClass IIZ-2040-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jun 27, 2025 · Open, Classified
a software anomaly allows test results to be reported using Micro Tip and Micro Slide diluent packs that the system treats as expired even when the Use Expired Reagents setting is not enabled if a shelf expiration date is omitted during manual loading.…
- Z-2041-2025 · Hardy Diagnostics · posted Jun 27, 2025 · Open, Classified
Identified lots failed incoming quality control testing. Lots exhibited breakthrough growth of Klebsiella pneumoniae.
- Z-2042-2025 · Onkos Surgical, Inc. · posted Jun 27, 2025 · Open, Classified
Required inspections were not performed on finished product prior to release and distribution.
- Z-2029-2025 · GE Medical Systems China Co., Ltd. · posted Jun 26, 2025 · Open, Classified
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead…
- GE Healthcare Versana Ultrasound Systems, sold under the following names and Model/Catalog Numbers:Class IIZ-2030-2025 · GE Medical Systems China Co., Ltd. · posted Jun 26, 2025 · Open, Classified
GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previous obstetric patient data in the Whizz report. This could potentially lead…
- Z-2033-2025 · Jewel Precision Sheet Metal & Machining Co, Inc. · posted Jun 26, 2025 · Open, Classified
The 8-inch model of the Jewel Precision Reusable Rigid Sterilization Container System (i.e., model number: JP-24-8) was not cleared by the FDA. The letter to file justifying the change in size was insufficient.
- Z-2034-2025 · Philips North America Llc · posted Jun 26, 2025 · Open, Classified
Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional software issues that affect CT performance.
- Z-2023-2025 · Becton Dickinson & Co. · posted Jun 25, 2025 · Open, Classified
BD has received additional complaints of intermittent false resistance results for PZA during susceptibility testing of Mycobacterium tuberculosis isolates.
- Z-2024-2025 · Limacorporate S.p.A · posted Jun 25, 2025 · Open, Classified
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
- Z-2025-2025 · Limacorporate S.p.A · posted Jun 25, 2025 · Open, Classified
There is the potential that the eccentrical adaptor with screw, black laser marking, use to identify the position may be missing.
- Jewel Precision Reusable Rigid Sterilization Container System. Model/Catalog Number: JP-24-6.Class IIZ-2026-2025 · Jewel Precision Sheet Metal & Machining Co, Inc. · posted Jun 25, 2025 · Open, Classified
The Instructions for Use do not match the 510(k) FDA cleared indications for use or sterility shelf life.
- Z-2027-2025 · Siemens Healthcare Diagnostics, Inc. · posted Jun 25, 2025 · Open, Classified
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
- REMISOL AdvanceClass IIZ-2028-2025 · Beckman Coulter Inc. · posted Jun 25, 2025 · Open, Classified
Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a…
- Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP.Class IIZ-2016-2025 · Insulet Corporation · posted Jun 24, 2025 · Open, Classified
Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone…
- Product Name [REF]:Class IIZ-2017-2025 · CareFusion 303, Inc. · posted Jun 24, 2025 · Open, Classified
During automated dispensing cabinet upgrade/installation/reimaging, component manager was left in installed mode, and if a Microsoft patch is then installed this may lead to an error preventing user access to the dispensing software application, which may…
- CDI OneView System BPM Probe:Class IIZ-2018-2025 · Terumo Cardiovascular Systems Corporation · posted Jun 24, 2025 · Open, Classified
Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes…
- Z-1999-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- Z-2000-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- Z-2001-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- Z-2002-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- Z-2003-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- REF: 381811,Class IIZ-2004-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- REF: 382544Class IIZ-2005-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- REF: 381923Class IIZ-2006-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- REF: 381434Class IIZ-2007-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · posted Jun 23, 2025 · Open, Classified
Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.
- Brand Name: ALTIVATE REVERSEClass IIZ-2008-2025 · Encore Medical, LP · posted Jun 23, 2025 · Open, Classified
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
- Brand Name: ALTIVATE REVERSEClass IIZ-2009-2025 · Encore Medical, LP · posted Jun 23, 2025 · Open, Classified
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
- Brand Name: ALTIVATE REVERSEClass IIZ-2010-2025 · Encore Medical, LP · posted Jun 23, 2025 · Open, Classified
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
- Brand Name: ALTIVATE REVERSEClass IIZ-2011-2025 · Encore Medical, LP · posted Jun 23, 2025 · Open, Classified
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
- Z-2012-2025 · Recordati Rare Diseases Inc. · posted Jun 23, 2025 · Open, Classified
Out of specification stability test result for the Caphosol B solution from process validation lots
- Z-1994-2025 · Barco N.V. · posted Jun 20, 2025 · Open, Classified
Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might…
- Z-1995-2025 · MEDLINE INDUSTRIES, LP - Northfield · posted Jun 20, 2025 · Open, Classified
There is a possible packaging defect impacting the outer Tyvec pouch that protects the sterile product.
- Z-1996-2025 · Beckman Coulter, Inc. · posted Jun 20, 2025 · Open, Classified
Analyzer has calibration issues where curves switched from passed to failed due to system errors during aspiration. This anomaly causes erroneous results and delays if unnoticed. RLU values may be misinterpreted as valid measurements by LIS.
- Z-1997-2025 · Bigfoot Biomedical · posted Jun 20, 2025 · Open, Classified
Due to a software/firmware issue diabetes management system may provide a device error and cause the insulin pen to become unresponsive and unable to function. This error may pose a potential health risk for people living with diabetes as they may not…
- LUCAS 2, 3 and 3.1Class IIZ-1998-2025 · Jolife AB · posted Jun 20, 2025 · Open, Classified
Due to demonstration units of chest compression system being provided to a customer which may potentially be utilized for clinical use.
- Lago X ↗Class IIZ-1941-2025 · Spectral Instruments Inc · posted Jun 18, 2025 · Open, Classified
The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.
- Ami HTX. ↗Class IIZ-1942-2025 · Spectral Instruments Inc · posted Jun 18, 2025 · Open, Classified
The X-ray safety and manufacturing date labels affixed to the Lago X/ AMI HTX instrument do not correspond exactly to the labeling stipulated by the FDA / CDRH.
- Z-1987-2025 · Preat Corp · posted Jun 18, 2025 · Open, Classified
Due to manufacturing issues, abutment products were manufactured with screw seat located too high, which may cause the mating screw to have a shorter engagement length into the associated implant.
- Z-1988-2025 · Balt USA, LLC · posted Jun 18, 2025 · Open, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
- Z-1989-2025 · Balt USA, LLC · posted Jun 18, 2025 · Open, Classified
Due to radiopaque (RO) marker was not visible during angiography and it was determined that the RO marker was not manufactured to specification.
- Z-1990-2025 · ICU Medical, Inc. · posted Jun 18, 2025 · Open, Classified
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication…
- Z-1991-2025 · ICU Medical, Inc. · posted Jun 18, 2025 · Open, Classified
Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication…
- Z-1992-2025 · Fresenius Medical Care Holdings, Inc. · posted Jun 18, 2025 · Open, Classified
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
- Z-1993-2025 · Diagnostica Stago, Inc. · posted Jun 18, 2025 · Open, Classified
Potential for shorter clotting times in clotting tests which have a dilution, leading to possible overestimate of around 1%
- Z-1981-2025 · Siemens Medical Solutions USA, Inc · posted Jun 17, 2025 · Open, Classified
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
- Z-1982-2025 · Siemens Medical Solutions USA, Inc · posted Jun 17, 2025 · Open, Classified
A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.
- Z-1983-2025 · Chamberlain Technologies LLC · posted Jun 17, 2025 · Completed
Potential presence of residual adhesive material on the mesh surface.
- Z-1984-2025 · Chamberlain Technologies LLC · posted Jun 17, 2025 · Completed
Potential presence of residual adhesive material on the mesh surface.
- Z-1985-2025 · Chamberlain Technologies LLC · posted Jun 17, 2025 · Completed
Potential presence of residual adhesive material on the mesh surface.
- Z-1986-2025 · Chamberlain Technologies LLC · posted Jun 17, 2025 · Completed
Potential presence of residual adhesive material on the mesh surface.
- Z-1977-2025 · Orthofix Srl · posted Jun 13, 2025 · Open, Classified
Retrospective; Their is the potential that implantable intramedullary lengthening nails may bend at the distal tail during insertion resulting in having to be replaced.
- Brand Name: MiniDock ↗Class IIZ-1978-2025 · Guangzhou Pluslife Biotech Co., Ltd. · posted Jun 13, 2025 · Open, Classified
Devices were distributed in the U.S. market without obtaining the required FDA premarket authorization (e.g., 510(k) clearance or PMA).
- Z-1979-2025 · Guangzhou Pluslife Biotech Co., Ltd. · posted Jun 13, 2025 · Open, Classified
Devices were distributed in the U.S. market without obtaining the required FDA premarket authorization (e.g., 510(k) clearance or PMA).
- Z-1980-2025 · Ohio Medical Corporation · posted Jun 13, 2025 · Open, Classified
Erroneous calibration values, potentially resulting in: as the device ages, oxygen purity will gradually decrease to the point where the device alarms to indicate low oxygen purity will be incorrect.
- Microtek Patient Drape ECOLAB ↗Class IIZ-1963-2025 · Microtek Medical, Inc. · posted Jun 12, 2025 · Open, Classified
Due to non-sterile products being labeled as sterile
- Z-1964-2025 · Baxter Healthcare Corporation · posted Jun 12, 2025 · Open, Classified
Baxter has identified a design issue which may cause the pump to detect the presence of a syringe flange when no syringe is loaded. When this issue occurs, the pumps 'user interface becomes stuck on the "Remove Syringe" screen when an infusion is complete,…
- Z-1965-2025 · Mobius Mobility LLC · posted Jun 12, 2025 · Open, Classified
Software issue that could potentially lead to the device tipping over from Balance Mode.
- Z-1966-2025 · Belmont Instrument LLC · posted Jun 12, 2025 · Open, Classified
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature…
- Brand Name: Guardian 2 SystemClass IIZ-1967-2025 · Turncare, Inc · posted Jun 12, 2025 · Open, Classified
A software issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- Brand Name: Guardian 2 SystemClass IIZ-1968-2025 · Turncare, Inc · posted Jun 12, 2025 · Open, Classified
A software update corrected an issue where "low" and "terminal" battery alerts did not display on the screen to the user in specific circumstances when the device was not running patient therapy.
- Azurion R1.x and R2.x systemsClass IIZ-1969-2025 · Philips Medical Systems Nederland B.V. · posted Jun 12, 2025 · Open, Classified
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result…
- Z-1970-2025 · Bd Switzerland Sarl · posted Jun 12, 2025 · Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment…
- Z-1971-2025 · Bd Switzerland Sarl · posted Jun 12, 2025 · Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment…
- Z-1972-2025 · Bd Switzerland Sarl · posted Jun 12, 2025 · Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment…
- Z-1973-2025 · Bd Switzerland Sarl · posted Jun 12, 2025 · Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment…
- Z-1974-2025 · Bd Switzerland Sarl · posted Jun 12, 2025 · Open, Classified
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment…
- Z-1975-2025 · MicroAire Surgical Instruments, LLC · posted Jun 12, 2025 · Open, Classified
Potential that the blade can unintentionally cut tissue prior to the user deploying the blade.
- Z-1947-2025 · Remel, Inc · posted Jun 11, 2025 · Open, Classified
Product may have an off color affecting perfomance
- Z-1948-2025 · Ortho-Clinical Diagnostics, Inc. · posted Jun 11, 2025 · Open, Classified
It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
- Brand Name: Tecfen Medical ↗Class IIZ-1949-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1950-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1951-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1952-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1953-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1954-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Brand Name: Tecfen Medical ↗Class IIZ-1955-2025 · Tecfen Medical · posted Jun 11, 2025 · Open, Classified
Sterility assurance for Ophthalmic knives.
- Z-1956-2025 · Contamac Solutions, Inc. · posted Jun 11, 2025 · Open, Classified
Contact lens insertion solution may lack sterility.
- Z-1957-2025 · Microbiologics Inc · posted Jun 11, 2025 · Open, Classified
Affected products were packaged with the incorrect KWIK STIK assembly resulting in the product containing the wrong microorganism. 0621K was packaged 01065K swabs; 01065K was packaged with 0621K swabs. This may lead to delayed diagnosis.
- Z-1962-2025 · Baxter Healthcare Corporation · posted Jun 11, 2025 · Open, Classified
The replacement rechargeable batteries inserted into the Welch Allyn PocketScope handles have the potential to cause a short circuit in the handle, which can lead to exposure to excessive heat and user burns.
- Z-1938-2025 · GE Medical Systems Information Technologies Inc · posted Jun 10, 2025 · Open, Classified
The MUSE Administrator Account active directory or local credential could potentially be used from your local network to log into a GE HealthCare MUSE 5 system and potentially manipulate recorded patient data.
- Z-1939-2025 · Stryker GmbH · posted Jun 10, 2025 · Open, Classified
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
- Z-1940-2025 · Stryker GmbH · posted Jun 10, 2025 · Open, Classified
Investigation determined that one lot of the T2 Greater Trochanter Nail (GTN) was labeled as left version but contained a right nail and another lot was labeled as right version but contained a left nail.
- Z-1819-2025 · GE Medical Systems, LLC · posted Jun 9, 2025 · Open, Classified
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- OEC 3DClass IIZ-1820-2025 · GE Medical Systems, LLC · posted Jun 9, 2025 · Open, Classified
OEC Elite and OEC 3D Mobile C-Arms Exceeding 4% Source to Image Distance in X-ray Field.
- Z-1925-2025 · Natus Neurology DBA Excel Tech., Ltd. (XLTEK) · posted Jun 9, 2025 · Open, Classified
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
- EPIONE Model 30-0001Class IIZ-1927-2025 · Quantum Surgical Sas · posted Jun 9, 2025 · Open, Classified
Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user.
- Achieva XR ↗Class IIZ-1928-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Achieva IT Nova ↗Class IIZ-1929-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Achieva Nova ↗Class IIZ-1930-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Achieva Nova-Dual ↗Class IIZ-1931-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Explorer/Nova Dual ↗Class IIZ-1932-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Master/Nova ↗Class IIZ-1933-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera 1.5T Omni/Stellar ↗Class IIZ-1934-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Intera Achieva 1.5T Pulsar ↗Class IIZ-1935-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- SmartPath to dStream for 1.5T ↗Class IIZ-1936-2025 · Philips North America · posted Jun 9, 2025 · Open, Classified
identified an issue where a component failure in the specific Gradient Coil type of affected systems can potentially act as a heat source, with a potential to produce smoke and/or fire. If smoke or a fire occurs, the risk to patients or operators may…
- Z-1937-2025 · Copan Italia · posted Jun 9, 2025 · Open, Classified
Swabs for specimen collection may be prone to breakage during product usage.
- Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.Class IIZ-1914-2025 · Stryker GmbH · posted Jun 6, 2025 · Open, Classified
Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
- LOWER EXTREMITY, Model: NCEX20IClass IIZ-1916-2025 · American Contract Systems, Inc. · posted Jun 6, 2025 · Open, Classified
Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged…
- Z-1917-2025 · Cardinal Health 200, LLC · posted Jun 6, 2025 · Completed
Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
- Kendall SCD Comfort SleevesClass IIZ-1918-2025 · Cardinal Health 200, LLC · posted Jun 6, 2025 · Completed
Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
- Kendall SCD Blended ComfortClass IIZ-1919-2025 · Cardinal Health 200, LLC · posted Jun 6, 2025 · Completed
Incorrect software logic may cause frequent E1 errors (system high pressure). Issue may cause delay in treatment or therapy.
- Z-1920-2025 · Olympus Corporation of the Americas · posted Jun 6, 2025 · Open, Classified
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
- Z-1921-2025 · Olympus Corporation of the Americas · posted Jun 6, 2025 · Open, Classified
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
- Z-1922-2025 · Olympus Corporation of the Americas · posted Jun 6, 2025 · Open, Classified
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
- Z-1923-2025 · Olympus Corporation of the Americas · posted Jun 6, 2025 · Open, Classified
Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.
- Z-1924-2025 · Medtronic, Inc. · posted Jun 6, 2025 · Open, Classified
Identified devices may experience sudden battery power failure without alarm, power supply power-on-resets, and interruption of therapy. Potential patient harms include insufficient pacing support leading to cardiac arrest, syncope, or hospitalization; or…
- Z-1926-2025 · Beckman Coulter, Inc. · posted Jun 6, 2025 · Open, Classified
A new motor/encoder and cable was introduced into instruments manufactured starting with serial number 300141 incorrectly. This incorrect installation is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in…
- Z-1909-2025 · Healthmark Industries Co., Inc. · posted Jun 5, 2025 · Open, Classified
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and…
- Z-1910-2025 · Healthmark Industries Co., Inc. · posted Jun 5, 2025 · Open, Classified
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and…
- Z-1911-2025 · Olympus Corporation of the Americas · posted Jun 5, 2025 · Open, Classified
Recent post-market surveillance data suggests a possible association of higher microbial contamination levels from TJF duodenoscopes when manual cleaning was delayed beyond one hour and a presoak was performed, compared to those TJF duodenoscopes where…
- Z-1913-2025 · Beckman Coulter, Inc. · posted Jun 5, 2025 · Open, Classified
A new software update (v1.20) is introducing error codes for the Dxl 9000 which prevents the assay from being performed and a delay in results from being generated, and may require field service to resolve the delay.
- Z-1905-2025 · Virtual Incision Corporation · posted Jun 4, 2025 · Completed
The MIRA Surgical System Software Version 4.0.3 loaded on the Surgeon Control Console may have the potential to exhibit interrupted motion on the display screen, caused by a software latency between the movement of the hand controllers and the movement of…
- Z-1906-2025 · Dexcom, Inc. · posted Jun 4, 2025 · Open, Classified
A bug in the 2.7.0 version of the G7 iOS App can cause the app to enter a state where it stops updating the estimated glucose value (EGV) and continues to show this stale EGV. If the user is unaware that the EGVs are not being updated and that glucose…
- Z-1907-2025 · Beckman Coulter Mishima K.K. · posted Jun 4, 2025 · Completed
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that…
- Z-1908-2025 · Beckman Coulter Mishima K.K. · posted Jun 4, 2025 · Completed
Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at…
- Z-1878-2025 · Flux Technology Inc. · posted Jun 3, 2025 · Open, Classified
Lazervida 10W Diode Laser Engraver does not fully meet current Federal performance standards in certain areas. Specifically, the product does not incorporate the additional beam attenuator, remote interlock connector, key switch, specific labels and manual…
- Z-1903-2025 · Luminex Corporation · posted Jun 3, 2025 · Completed
It was determined that the consumable issue may result in No Call - INT CTL 1 and/or False Negative results on a small portion of the lot of test cartridges.
- Z-1893-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1894-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1895-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1896-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1897-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1898-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1899-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1900-2025 · Silex Medical, LLC · posted Jun 2, 2025 · Open, Classified
Insulation on the tube did not extend fully to the proximal end of the shaft on the instrument, leaving an exposed (not insulated) area of the shaft closer to the rotator wheel.
- Z-1901-2025 · Beckman Coulter, Inc. · posted Jun 2, 2025 · Open, Classified
Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23%…
Class III recalls (2)
Using the device is not likely to cause health problems, but it violates FDA rules.
- Z-2046-2025 · Lumithera Inc · posted Jun 30, 2025 · Open, Classified
U.S. customers were shipped devices that were configured for the European Union and were unable to plug the device in the electrical outlet.
- AcoSound. Model Number: LW12-BTE-MClass IIIZ-1976-2025 · Hangzhou Acosound Technology CO., Ltd. · posted Jun 13, 2025 · Open, Classified
Three issues 1. Label error: The manufacturer information in the labeling is not correct, the manufacturer's name is mislabeled as "Developed by AcoSound," Correct should be "Manufactured by Hangzhou AcoSound." And the labeling has no statement of the…
Source: FDA Medical Device Recalls and FDA enforcement reports (recall class), via openFDA. Records are grouped by the date the FDA posted them. Methodology · Not medical advice.