5008X HDF Hemodialysis Blood Tubing Set, 24 units per case… recallZ-1992-2025 · Fresenius Medical Care Holdings, Inc. · initiated Apr 29, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Fresenius Medical Care Holdings, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
FDA root cause category: Process control
What the company is doing
Fresenius Medical Care notified its single consignee on 04/29/2025 via telephone call. The consignee was made aware of the issue and affected units and instructed to examine their inventory for those affected units. If they had any affected units, they were instructed to remove them from their inventory immediately and provide contact information to arrange for their return. On 05/14/2025, Fresenius Medical Care expanded the recall to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. On 06/05/2025, Fresenius Medical Care expanded the recall again to include additional lot numbers and notified consignees. The notification was via telephone call and included the same instructions as the initial letter but included the additional lots. A recall notification letter was sent to the consignee on about 06/23/2025. The letter provided the same information as was verbally communicated in notifications.
Recall details
- Recall number
- Z-1992-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 29, 2025
- Posted by the FDA
- Jun 18, 2025
- Recalling firm
- Fresenius Medical Care Holdings, Inc.
- Quantity
- Model Number, Cases (Eaches): 03-5100-7, 76 (1,824); 03-5100-7C, 5 (120); 03-5150-2, 8 (192)
- Product code
- KDI
- FDA submissions
- K231534
- Distribution
- US Nationwide distribution in the state of Massachusetts.
Product description
5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.
Code information (models, lots, serial numbers)
a. Model Number: 03-5100-7; UDI-DI: 00840861102365 (Bag), 10840861102362 (Case); Lot Numbers: 24LR01214, 24SR01010, 25ER01800. b. Model Number: 03-5100-7C; UDI-DI: 0084086110237 (Bag), 10840861102379 (Case); Lot Number: 24LR01215, 24SR01011, 25ER01804. c. Model Number: 03-5150-2; UDI-DI: 00840861102389 (Bag), 10840861102386 (Case); Lot Number: 24LR01216, 24SR01012, 25ER01805.
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- 5008X HDF Hemodialysis Blood Tubing Set 03-5100-7Fresenius Medical Care Renal Therapies Group, LLC · Accessories, Blood Circuit, Hemodialysis1 recall10 reports
- 5008X HDF Hemodialysis Blood Tubing Set with CLiC Blood Chamber 03-5100-7CFresenius Medical Care Renal Therapies Group, LLC · Accessories, Blood Circuit, Hemodialysis1 recall0 reports
- 5008X HDF Hemodialysis Blood Tubing Set with Access Flow Reversing Connector (Tw 03-5150-2Fresenius Medical Care Renal Therapies Group, LLC · Accessories, Blood Circuit, Hemodialysis1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page